Advancement in the routine identification of anaerobic bacteria by MALDI-TOF mass spectrometry

被引:44
|
作者
Coltella, L. [1 ]
Mancinelli, L. [1 ]
Onori, M. [1 ]
Lucignano, B. [2 ]
Menichella, D. [3 ]
Sorge, R. [4 ]
Raponi, M. [3 ]
Mancini, R. [5 ]
Russo, C. [1 ]
机构
[1] IRCCS, Bambino Gesu Childrens Hosp, Virol Unit, Dept Lab Med, I-00165 Rome, Italy
[2] IRCCS, Bambino Gesu Childrens Hosp, Bacteriol Unit, Dept Lab Med, I-00165 Rome, Italy
[3] IRCCS, Bambino Gesu Childrens Hosp, Med Direct, I-00165 Rome, Italy
[4] Univ Tor Vergata, Lab Biometry, Rome, Italy
[5] Univ Roma La Sapienza, Fac Med & Odontoiatry, Rome, Italy
关键词
DESORPTION IONIZATION-TIME; SPECIES IDENTIFICATION; MICROBIOLOGY; INFECTIONS; NOV; MS;
D O I
10.1007/s10096-013-1865-1
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
We evaluated matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) Biotyper as a tool for the identification of anaerobic bacteria compared with 500 base-pair (bp) 16S ribosomal ribonucleic acid (rRNA) gene sequencing analysis, which is considered to be the "gold standard" method. A total of 484 anaerobic bacteria were retrieved from the clinical specimens of 318 pediatric patients. Molecular identification resulted in 18 genera and 51 species. The most prevalent genus was Clostridium (76.85 %), with 70 % C. difficile isolates. The concordance and sensitivity determined by MALDI-TOF MS for C. difficile, the most prevalent species isolated, was 94.08 %, whereas the specificity was 100 %. For the other anaerobes, the sensitivity and specificity were 94.07 % and 81.82 %, respectively, with a concordance of 93.15 %. Low performance was observed for Propionibacterium acnes and Fusobacterium nucleatum, for which a dedicated pretreatment procedure should likely be set up. MALDI-TOF MS was shown to be a valid alternative for the fast and reliable identification of the most clinically relevant anaerobic bacteria; moreover, it is less time-consuming, the cost for reagents is minimized, and it does not require dedicated personnel.
引用
收藏
页码:1183 / 1192
页数:10
相关论文
共 50 条
  • [1] Advancement in the routine identification of anaerobic bacteria by MALDI-TOF mass spectrometry
    L. Coltella
    L. Mancinelli
    M. Onori
    B. Lucignano
    D. Menichella
    R. Sorge
    M. Raponi
    R. Mancini
    C. Russo
    European Journal of Clinical Microbiology & Infectious Diseases, 2013, 32 : 1183 - 1192
  • [2] Evaluation of MALDI-TOF mass spectrometry for identification of anaerobic bacteria
    Barba, M. J.
    Fernandez, A.
    Oviano, M.
    Fernandez, B.
    Velasco, D.
    Bou, G.
    ANAEROBE, 2014, 30 : 126 - 128
  • [3] Reliability of MALDI-TOF mass spectrometry in the identification of anaerobic bacteria
    Vega-Castano, Silvia
    Ferreira, Laura
    Gonzalez-Avila, Magdalena
    Sanchez-Juanes, Fernando
    Inmaculada Garcia-Garcia, M.
    Elias Garcia-Sanchez, Jose
    Manuel Gonzalez-Buitrago, Jose
    Luis Munoz-Bellido, Juan
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2012, 30 (10): : 597 - 601
  • [4] MALDI-TOF mass spectrometry of bacteria
    Lay, JO
    MASS SPECTROMETRY REVIEWS, 2001, 20 (04) : 172 - 194
  • [5] Routine identification of Nocardia species by MALDI-TOF mass spectrometry
    Girard, Victoria
    Mailler, Sandrine
    Polsinelli, Sophie
    Jacob, Delphine
    Saccomani, Marie Christine
    Celliere, Beatrice
    Monnin, Valerie
    van Belkum, Alex
    Hagen, Ferry
    Meis, Jacques F.
    Durand, Geraldine
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2017, 87 (01) : 7 - 10
  • [6] MALDI-TOF Mass Spectrometry Applied to Classification and Identification of Bacteria
    Schumann, Peter
    Maier, Thomas
    NEW APPROACHES TO PROKARYOTIC SYSTEMATICS, 2014, 41 : 275 - 306
  • [7] Rapid identification of bacteria in clinical microbiology routine diagnostics using MALDI-TOF mass spectrometry
    Reich, M.
    Stark, M.
    Huesgen, G.
    Beyser, K.
    Borgmann, S.
    INTERNATIONAL JOURNAL OF MEDICAL MICROBIOLOGY, 2009, 299 : 3 - 3
  • [8] Chemical extraction versus direct smear for MALDI-TOF mass spectrometry identification of anaerobic bacteria
    Fournier, Remi
    Wallet, Frederic
    Grandbastien, Bruno
    Dubreuil, Luc
    Courcol, Rene
    Neut, Christel
    Dessein, Rodrigue
    ANAEROBE, 2012, 18 (03) : 294 - 297
  • [9] Evaluation of MALDI-TOF mass spectrometry for the identification of bacteria growing as biofilms
    Gaudreau, A. M.
    Labrie, J.
    Goetz, C.
    Dufour, S.
    Jacques, M.
    JOURNAL OF MICROBIOLOGICAL METHODS, 2018, 145 : 79 - 81
  • [10] Use of MALDI-TOF mass spectrometry for identification of bacteria that are difficult to culture
    Biswas, Silpak
    Rolain, Jean-Marc
    JOURNAL OF MICROBIOLOGICAL METHODS, 2013, 92 (01) : 14 - 24