Optimal Pulse Durations for the Treatment of Leg Telangiectasias with an Alexandrite Laser

被引:12
作者
Ross, E. V. [2 ]
Meehan, K. J. [2 ]
Gilbert, S. [1 ]
Domankevitz, Y. [1 ]
机构
[1] Candela Corp, Road Wayland, MA USA
[2] Naval Med Ctr, San Diego, CA USA
关键词
laser; vein; spider; light; leg; telangiectasia; alexandrite; ND-YAG LASER; CUTANEOUS BLOOD-VESSELS; RETICULAR VEINS; SCLEROTHERAPY; HAIR; VENULECTASIAS; IRRADIATION; MECHANISMS; THERAPY; FLUENCE;
D O I
10.1002/lsm.20737
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background and Objectives: Determine optimal settings using a long pulse 755 nm alexandrite laser in the treatment of superficial leg veins. Study Design\Materials and Methods: Fifteen patients with Fitzpatrick skin types I-III with telangiectasia. ranging from 0.2 to 1.0 mm were enrolled. Spot size varied from 3 to 6 mm. Pulse durations ranged from 3 to 100 milliseconds. For each pulse duration, test sites were performed to determine threshold radiant exposures using persistent bluing and/or immediate stenosis (closure) as the clinical endpoint. Test sites were re-evaluated 21 days later. Optimal settings, those that resulted in the greatest clearance with minimal side effects (pain, purpura, epidermal damage, pigment changes), were used to treat a larger area of like-sized vessels. Follow-up evaluations were conducted 12 weeks after a single treatment using the optimal setting. Polarized digital photographs were obtained at each visit. Improvement was determined by blinded evaluation of pre/post-treatment photographs. Results: Fourteen patients completed the study. Radiant exposure thresholds for immediate vessel changes depended on vessel diameter, with larger radiant exposures required for smaller spot sizes and smaller vessels. The average threshold radiant exposure for closure was 89 J/cm(2). The optimal pulse duration was 60 milliseconds for most of the patients. With this pulse width, clearance approached 65% 12 weeks after a single treatment. Transient hyperpigmentation occurred in four patients. Increasing the pulse duration improved epidermal tolerance and decreased the likelihood of purpura. Conclusions: By lengthening the pulsewidth beyond 3 milliseconds, a long pulse alexandrite laser achieves satisfactory clearance with an improved side effect profile. Lasers Surg. Med. 41:104-109, 2009. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:104 / 109
页数:6
相关论文
共 30 条
[1]   Photothermal coagulation of blood vessels: a comparison of high-speed optical coherence tomography and numerical modelling [J].
Barton, JK ;
Rollins, A ;
Yazdanfar, S ;
Pfefer, TJ ;
Westphal, V ;
Izatt, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (06) :1665-1678
[2]  
Barton JK, 1999, LASER SURG MED, V24, P236, DOI 10.1002/(SICI)1096-9101(1999)24:3<236::AID-LSM9>3.0.CO
[3]  
2-I
[4]  
Barton JK, 1999, PHOTOCHEM PHOTOBIOL, V70, P916, DOI 10.1562/0031-8655(1999)070<0916:LFFPDO>2.3.CO
[5]  
2
[6]  
Barton JK, 2001, PHOTOCHEM PHOTOBIOL, V73, P642, DOI 10.1562/0031-8655(2001)073<0642:CPITPT>2.0.CO
[7]  
2
[8]   Evaluation of the long pulsed high fluence alexandrite laser therapy of leg telangiectasia [J].
Brunnberg, S ;
Lorenz, S ;
Landthaler, M ;
Hohenleutner, U .
LASERS IN SURGERY AND MEDICINE, 2002, 31 (05) :359-362
[9]   Comparative pilot study evaluating the treatment of leg veins with a long pulse Nd:YAG laser and sclerotherapy [J].
Coles, CM ;
Werner, RS ;
Zelickson, BD .
LASERS IN SURGERY AND MEDICINE, 2002, 30 (02) :154-159
[10]   A side-by-side comparative study of 1064 nm Nd:YAG, 810 nm diode and 755 nm alexandrite lasers for treatment of 0.3-3 mm leg veins [J].
Eremia, S ;
Li, C ;
Umar, SH .
DERMATOLOGIC SURGERY, 2002, 28 (03) :224-230