Point-of-care testing of neonatal coagulation

被引:8
|
作者
Tan, K
Booth, D
Newell, SJ
Dear, PRF
Hughes, C
Richards, M
机构
[1] St Jamess Univ Hosp, Neonatol Unit, Leeds, W Yorkshire, England
[2] St Jamess Univ Hosp, Dept Paediat Haematol & Oncol, Leeds, W Yorkshire, England
来源
CLINICAL AND LABORATORY HAEMATOLOGY | 2006年 / 28卷 / 02期
关键词
coagulation; point-of-care testing; neonatal; validation study;
D O I
10.1111/j.1365-2257.2006.00765.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this validation study was to compare prothrombin time (PT) and activated partial thromboplastin time (APTT) results from a point-of-care testing (POCT) device (Rapidpoint Coag) with those from standard laboratory tests. The subjects were newborn infants needing coagulation screen for any clinical indications within a regional neonatal intensive care unit. The level of agreement between POCT and laboratory measurements of PT and APTT was determined. For PT: the bias was from -7.6 to 12.4 s and precision was 5.0 s. For the detection of prolonged PT at a level of 16 s, the sensitivity was 0.70, specificity was 0.57 and the positive predictive value (PPV) was 0.62. For APTT: the bias was from -39.1 to 23.7 s, and precision was 15.7 s. For the detection of prolonged APTT at a level of 55 s, the sensitivity was 0.80, specificity was 0.95 and the PPV was 0.80. The POCT device tested has limited utility as a cot-side device for screening for a prolongation of the APTT in the newborn but is not sensitive for screening for prolongation of the PT.
引用
收藏
页码:117 / 121
页数:5
相关论文
共 50 条
  • [31] Capillary blood for point-of-care testing
    Tang, Ruihua
    Yang, Hui
    Choi, Jane Ru
    Gong, Yan
    You, MinLi
    Wen, Ting
    Li, Ang
    Li, XiuJun
    Xu, Bo
    Zhang, Sufeng
    Mei, Qibing
    Xu, Feng
    CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2017, 54 (05) : 294 - 308
  • [32] Point-of-Care diagnostics of coagulation in the management of bleeding and transfusion in trauma patients
    Francis, Roland C. E.
    Theurl, Igor
    Maegele, Marc
    Graw, Jan A.
    CURRENT OPINION IN ANESTHESIOLOGY, 2020, 33 (02) : 246 - 252
  • [33] Educational Case: Point-of-Care Testing
    McCool, Isaac E.
    Muir, Jeannie M.
    Knollmann-Ritschel, Barbara E. C.
    ACADEMIC PATHOLOGY, 2020, 7
  • [34] Practical Challenges of Point-of-Care Testing
    Ortiz, Daniel A.
    Loeffelholz, Michael J.
    CLINICS IN LABORATORY MEDICINE, 2023, 43 (02) : 155 - 165
  • [35] Salivary Sensors in Point-of-Care Testing
    Yamaguchi, Masaki
    SENSORS AND MATERIALS, 2010, 22 (04) : 143 - 153
  • [36] Point-of-Care Testing for Pharyngitis in the Pharmacy
    Essack, Sabiha
    Bell, John
    Burgoyne, Douglas
    Tongrod, Wirat
    Duerden, Martin
    Sessa, Aurelio
    Altiner, Attila
    Shephard, Adrian
    ANTIBIOTICS-BASEL, 2020, 9 (11): : 1 - 16
  • [37] Point-of-Care Testing: General Aspects
    Ferreira, Carlos E. S.
    Guerra, Joao C. C.
    Slhessarenko, Natasha
    Scartezini, Marileia
    Franca, Carolina N.
    Colombini, Marjorie P.
    Berlitz, Fernando
    Machado, Antonia M. O.
    Campana, Gustavo A.
    Faulhaber, Adriana C. L.
    Galoro, Cesar A.
    Dias, Claudia M.
    Shcolnik, Wilson
    Martino, Marines D. V.
    Cesar, Katia R.
    Sumita, Nairo M.
    Mendes, Maria E.
    Faulhaber, Marcelo H. W.
    Pinho, Joao R. R.
    Barbosa, Ismar V.
    Batista, Marcelo C.
    Khawali, Cristina
    Pariz, Vitor M.
    Andriolo, Adagmar
    CLINICAL LABORATORY, 2018, 64 (1-2) : 1 - 9
  • [38] Improving the quality of point-of-care testing
    Price, Christopher P.
    Smith, Ian
    Van den Bruel, Ann
    FAMILY PRACTICE, 2018, 35 (04) : 358 - 364
  • [39] QUALITY IMPROVEMENT AND POINT-OF-CARE TESTING
    KIECHLE, FL
    INGRAMMAIN, R
    JOURNAL OF CLINICAL LIGAND ASSAY, 1995, 18 (01): : 14 - 20
  • [40] Cost analysis of a neonatal point-of-care monitor
    Alves-Dunkerson, JA
    Hilsenrath, PE
    Cress, GA
    Widness, JA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2002, 117 (05) : 809 - 818