Point-of-care testing of neonatal coagulation

被引:8
|
作者
Tan, K
Booth, D
Newell, SJ
Dear, PRF
Hughes, C
Richards, M
机构
[1] St Jamess Univ Hosp, Neonatol Unit, Leeds, W Yorkshire, England
[2] St Jamess Univ Hosp, Dept Paediat Haematol & Oncol, Leeds, W Yorkshire, England
来源
CLINICAL AND LABORATORY HAEMATOLOGY | 2006年 / 28卷 / 02期
关键词
coagulation; point-of-care testing; neonatal; validation study;
D O I
10.1111/j.1365-2257.2006.00765.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this validation study was to compare prothrombin time (PT) and activated partial thromboplastin time (APTT) results from a point-of-care testing (POCT) device (Rapidpoint Coag) with those from standard laboratory tests. The subjects were newborn infants needing coagulation screen for any clinical indications within a regional neonatal intensive care unit. The level of agreement between POCT and laboratory measurements of PT and APTT was determined. For PT: the bias was from -7.6 to 12.4 s and precision was 5.0 s. For the detection of prolonged PT at a level of 16 s, the sensitivity was 0.70, specificity was 0.57 and the positive predictive value (PPV) was 0.62. For APTT: the bias was from -39.1 to 23.7 s, and precision was 15.7 s. For the detection of prolonged APTT at a level of 55 s, the sensitivity was 0.80, specificity was 0.95 and the PPV was 0.80. The POCT device tested has limited utility as a cot-side device for screening for a prolongation of the APTT in the newborn but is not sensitive for screening for prolongation of the PT.
引用
收藏
页码:117 / 121
页数:5
相关论文
共 50 条
  • [21] Point-of-care testing for emergency assessment of coagulation in patients treated with direct oral anticoagulants
    Ebner, Matthias
    Birschmann, Ingvild
    Peter, Andreas
    Spencer, Charlotte
    Haertig, Florian
    Kuhn, Joachim
    Blumenstock, Gunnar
    Zuern, Christine S.
    Ziemann, Ulf
    Poli, Sven
    CRITICAL CARE, 2017, 21
  • [22] Guidelines for point-of-care testing: haematology
    Briggs, Carol
    Guthrie, David
    Hyde, Keith
    Mackie, Ian
    Parker, Norman
    Popek, Mary
    Porter, Neil
    Stephens, Clare
    BRITISH JOURNAL OF HAEMATOLOGY, 2008, 142 (06) : 904 - 915
  • [23] Point-of-care testing in diabetes mellitus
    Price, CP
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (09) : 1213 - 1219
  • [24] Artificial Intelligence in Point-of-Care Testing
    Khan, Adil I.
    Khan, Mazeeya
    Khan, Raheeb
    ANNALS OF LABORATORY MEDICINE, 2023, 43 (05) : 401 - 407
  • [25] Point-of-care testing: A critical review
    Mor, M
    Waisman, Y
    PEDIATRIC EMERGENCY CARE, 2000, 16 (01) : 45 - 48
  • [26] The point-of-care testing in the emergency department
    Carraro, Paolo
    EMERGENCY CARE JOURNAL, 2019, 15 (02) : 46 - 50
  • [27] Management of a Point-of-Care Testing Program
    Gregory, Kim
    Lewandrowski, Kent
    CLINICS IN LABORATORY MEDICINE, 2009, 29 (03) : 433 - +
  • [28] Infectious Mononucleosis Testing at the Point-of-Care
    Russell, Barbara L.
    Steele, John C. H., Jr.
    POINT OF CARE, 2009, 8 (01): : 29 - 31
  • [29] Management Challenges in Point-of-Care Testing
    Hill, Halcyon St.
    POINT OF CARE, 2013, 12 (02): : 118 - 120
  • [30] Management Challenges With Point-of-Care Testing
    Clark, Karen S.
    POINT OF CARE, 2013, 12 (02): : 97 - 98