Design and First Results of an Innovative and Dedicated Breast PET

被引:1
作者
Gonzalez Martinez, Antonio J. [1 ,3 ]
Aukema, Tjeerd S. [2 ]
Ballester, Julio B. [1 ]
Benlloch Baviera, Jose M. [3 ]
Salvador, Carlos C. [1 ]
Palomares, Abel O. [3 ]
Asensi, Antonio S. [3 ]
Vogel, Wouter [2 ]
Barreiros, Carlor V. [1 ]
Vidal San Sebastian, Luis F. [3 ]
Olmos, Renato A. Valdes [2 ]
机构
[1] ONCOVIS GEM Imaging SA, Valencia, Spain
[2] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Nucl Med, Amsterdam, Netherlands
[3] Ctr Mixto UPV CSIC CIEMAT, Inst Instrumentac Imagen Mol I3M, Valencia 46022, Spain
关键词
Positron emission mammography; breast cancer; PET; clinical trial; POSITRON-EMISSION-TOMOGRAPHY; CANCER; CRYSTALS;
D O I
10.2174/157340512800672199
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work the design and preliminary clinical results obtained with an innovative and dedicated breast PET is presented. A geometrical study clearly pointed out the advantages of the prone position and the technique of hanging breast in comparison to the supine, conventionally used for other imaging modalities. Occult, multifocal and also lesions close to the pectoral muscle were depicted by means of the prone method. The measured diameters of the hanging breasts allowed us to define an optimum transaxial FOV of up to 170 mm. This was achieved by means of 12 detector heads mounted on ring geometry, each using a monolithic LYSO crystal a position sensitive photomultiplier, among other components. The detector showed up an extrinsic spatial resolution as high as 1.4 mm in the center of the FOV with a physical sensitivity of 1.2 %. A clinical trial is undergoing at the Netherlands Cancer Institute in terms of a clinical validation of the dedicated breast PET. An adequate image matching between this device and the whole-body PET was observed, even for deep chest lesions. A preliminary quantitative comparison also shows a good agreement between both systems. However, a tiny deviation to higher Standard Uptake Values is observed for the prototype.
引用
收藏
页码:144 / 150
页数:7
相关论文
共 13 条
[1]   Accuracy of whole-body dual-modality fluorine-18-2-fluoro-2-deoxy-D-glucose positron emission tomography and computed tomography (FDG-PET/CT) for tumor staging in solid tumors: Comparison with CT and PET [J].
Antoch, G ;
Saoudi, N ;
Kuehl, H ;
Dahmen, G ;
Mueller, SP ;
Beyer, T ;
Bockisch, A ;
Debatin, JF ;
Freudenberg, LS .
JOURNAL OF CLINICAL ONCOLOGY, 2004, 22 (21) :4357-4368
[2]  
Aukema T, EUR J NUCL MED UNPUB
[3]   The role of FDG PET/CT in patients with locoregional breast cancer recurrence: A comparison to conventional imaging techniques [J].
Aukema, T. S. ;
Rutgers, E. J. Th. ;
Vogel, W. V. ;
Teertstra, H. J. ;
Oldenburg, H. S. ;
Peeters, M. T. F. D. Vrancken ;
Wesseling, J. ;
Russell, N. S. ;
Olmos, R. A. Valdes .
EJSO, 2010, 36 (04) :387-392
[4]  
Ay MR, 2007, IRAN J NUCL MED, V15, P1
[5]   Initial performance evaluation of a high resolution Albira small animal positron emission tomography scanner with monolithic crystals and depth-of-interaction encoding from a user's perspective [J].
Balcerzyk, Marcin ;
Kontaxakis, George ;
Delgado, Mercedes ;
Garcia-Garcia, Luis ;
Correcher, Carlos ;
Gonzalez, Antonio J. ;
Gonzalez, Aurora ;
Rubio, Jose L. ;
Benlloch, Jose M. ;
Pozo, Miguel A. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (10)
[6]   Scanner calibration of a small animal PET camera based on continuous LSO crystals and flat panel PSPMTs [J].
Benlloch, J. M. ;
Carrilero, V. ;
Gonzalez, A. J. ;
Catret, J. ;
Lerche, Ch. W. ;
Abellan, D. ;
de Quiros, F. Garcia ;
Gimenez, M. ;
Modia, J. ;
Sanchez, F. ;
Pavon, N. ;
Ros, A. ;
Martinez, J. ;
Sebastia, A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :26-29
[7]   F-18 fluorodeoxyglucose positron-emission tomography in the diagnosis of tumor recurrence and metastases in the follow-up of patients with breast carcinoma - A comparison to conventional imaging [J].
Gallowitsch, HJ ;
Kresnik, E ;
Gasser, J ;
Kumnig, G ;
Igerc, I ;
Mikosch, P ;
Lind, P .
INVESTIGATIVE RADIOLOGY, 2003, 38 (05) :250-256
[8]   [18F]FDG in recurrent breast cancer:: diagnostic performances, clinical impact and relevance of induced changes in management [J].
Grahek, D ;
Montravers, F ;
Kerrou, K ;
Aide, N ;
Lotz, JP ;
Talbot, JN .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (02) :179-188
[9]   DOI measurement with monolithic scintillation crystals:: a primary performance evaluation [J].
Lerche, Christoph W. ;
Ros, Ana ;
Gadea, Rafael ;
Colom, Ricardo J. ;
Toledo, Francisco J. ;
Herrero, Vicente ;
Monzo, Jose M. ;
Sebastia, Angel ;
Abellan, Dori ;
Sanchez, Filomeno ;
Correcher, Carlos ;
Gonzalez, Antonio J. ;
Munar, Antoni ;
Benlloch, Jose M. .
2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, :2594-+
[10]   Clinical Imaging Characteristics of the Positron Emission Mammography Camera: PEM Flex Solo II [J].
MacDonald, Lawrence ;
Edwards, John ;
Lewellen, Thomas ;
Haseley, David ;
Rogers, James ;
Kinahan, Paul .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (10) :1666-1675