Review of Penile Prosthetic Reservoir: Complications and Presentation of a Modified Reservoir Placement Technique

被引:75
作者
Levine, Laurence A. [1 ]
Hoeh, Michael P. [1 ]
机构
[1] Rush Univ, Med Ctr, Dept Urol, Chicago, IL 60612 USA
关键词
Penile Prosthesis Complications; Penile Prosthesis Reservoir; Reservoir Complications; Three-Piece Inflatable Prosthesis; SMALL BOWEL OBSTRUCTION; DEEP VENOUS THROMBOSIS; GLYCOSYLATED HEMOGLOBIN; RADICAL PROSTATECTOMY; PATIENT SATISFACTION; RETAINED RESERVOIR; EXPERIENCE; INFECTION; EROSION; BLADDER;
D O I
10.1111/j.1743-6109.2012.02807.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction. Multiple modifications have been made to the inflatable penile prosthesis (IPP) since its inception in the 1970s. These modifications have made reservoir-related mechanical malfunctions highly unlikely in current IPP models. Although these complications are rare, it would be incumbent upon the implanting surgeon to be aware of these potential complications, how they present, how they are best treated, and how to prevent them from occurring. Aims. The aim of this article was to present our experience with complications associated with penile prosthesis reservoirs, perform a review of the literature regarding reservoir-related complications, and present our modified technique to place the reservoir into the space of Retzius. Main Outcome Measures. Reservoir-related complications including inguinal herniation, erosion into bladder or bowel, intraperitoneal reservoir placement with subsequent visceral injury, vascular injury, autoinflation, and infection. Methods. We retrospectively reviewed our experience with penile prosthesis reservoir complications or procedures requiring an alternative implantation approach at our center over the past 10 years where over 400 devices were implanted. We also review reservoir-related complications published in the English literature since the 1980s. Results. While exceedingly rare, reservoir complications do occur. Six cases from our institution are presented including one reservoir herniation, one postoperative direct inguinal hernia, one bladder laceration during revision surgery, one ectopic reservoir placement due to morbid obesity, one iliac vein compression syndrome, and one vascular laceration during reservoir revision. Reported reservoir complications include inguinal herniation, erosion into the bladder or bowel, intraperitoneal reservoir placement with subsequent injury to the ureter or bowel, vascular injury, autoinflation, and infection. Conclusion. Penile prosthesis reservoirs rarely fail mechanically but are associated with a variety of complications or may require alternate implantation technique. In our experience, the Jorgensen scissors technique allows safe entry into the space of Retzius with diminished risk of hernia as well as vascular, bladder, or bowel injury. Levine LA and Hoeh MP. Review of penile prosthetic reservoir: Complications and presentation of a modified reservoir Placement technique. J Sex Med 2012;9:2759-2769.
引用
收藏
页码:2759 / 2769
页数:11
相关论文
共 48 条
[1]   Laparascopic Capsulotomy to Treat Autoinflation of Inflatable Penile Prostheses [J].
Abbosh, Phillip H. ;
Thom, Matthew R. ;
Bullock, Arnold .
JOURNAL OF SEXUAL MEDICINE, 2012, 9 (04) :1212-1215
[2]   USE OF GLYCOSYLATED HEMOGLOBIN TO IDENTIFY DIABETICS AT HIGH-RISK FOR PENILE PERIPROSTHETIC INFECTIONS [J].
BISHOP, JR ;
MOUL, JW ;
SIHELNIK, SA ;
PEPPAS, DS ;
GORMLEY, TS ;
MCLEOD, DG .
JOURNAL OF UROLOGY, 1992, 147 (02) :386-388
[3]   Reservoir repositioning and successful thrombectomy for deep venous thrombosis secondary to compression of pelvic veins by an inflatable penile prosthesis reservoir [J].
Brison, Daniel ;
Ilbeigi, Pedram ;
Sadeghi-Nejad, Hossein .
JOURNAL OF SEXUAL MEDICINE, 2007, 4 (04) :1185-1187
[4]   Retrieval of inflatable penile prosthesis reservoir from the bladder [J].
Brusky, J ;
Dikranin, A ;
Aboseif, S .
INTERNATIONAL JOURNAL OF IMPOTENCE RESEARCH, 2005, 17 (03) :302-303
[5]   Efficacy, safety and patient satisfaction outcomes of the AMS 700CX inflatable penile prosthesis: Results of a long-term multicenter study [J].
Carson, CC ;
Mulcahy, JJ ;
Govier, FE .
JOURNAL OF UROLOGY, 2000, 164 (02) :376-380
[6]   Deep venous thrombosis due to compression of external iliac vein by the penile prosthesis reservoir [J].
daJusta, DG ;
Bianco, FJ ;
Ogle, A ;
Dhabuwala, CB .
UROLOGY, 2003, 61 (02) :462
[7]   Severe Vascular Complication after Implantation of a Three-Piece Inflatable Penile Prosthesis [J].
Deho', Federico ;
Henry, Gerard D. ;
Marone, Enrico M. ;
Sacca', Antonino ;
Chiesa, Roberto ;
Rigatti, Patrizio ;
Montorsi, Francesco .
JOURNAL OF SEXUAL MEDICINE, 2008, 5 (12) :2956-2959
[8]  
Dhar NB, 2006, J UROL 1, V176, P601
[9]   Long-term mechanical reliability of AMS 700CX™/CXM inflatable penile prosthesis [J].
Dhar, Nivedita Bhatta ;
Angermeier, Kenneth W. ;
Montague, Drogo K. .
JOURNAL OF UROLOGY, 2006, 176 (06) :2599-2601
[10]   EROSION OF AN INFLATABLE PENILE PROSTHESIS RESERVOIR INTO THE BLADDER, PRESENTING AS BLADDER CALCULI [J].
DUPONT, MC ;
HOCHMAN, HI .
JOURNAL OF UROLOGY, 1988, 139 (02) :367-368