Oxygen flux fluorescence lifetime imaging

被引:54
作者
Hartmann, P [1 ]
Ziegler, W [1 ]
Holst, G [1 ]
Lubbers, DW [1 ]
机构
[1] MAX PLANCK INST MOL PHYSIOL,D-44139 DORTMUND,GERMANY
关键词
fluorescent lifetime imaging (FLIM); oxygen; luminescent sensors; skin; phase-resolved spectroscopy;
D O I
10.1016/S0925-4005(97)80179-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe a device capable of imaging distributions of surface PO2 as well as oxygen flux (e.g., into the human skin), Oxygen is monitored by a phase-angle-sensitive imaging technique using the effect of luminescence lifetime quenching of ruthenium complexes incorporated into a polymer layer system. The dyes are excited by modulated radiation of blue light-emitting diodes. The phase-shifted luminescence is detected by an intensified CCD camera. Several modulation techniques have been comparatively investigated. The obtained images correspond pixel-by-pixel to the local luminescence lifetimes, which are a direct measure for the distribution of surface PO2, or alternatively for the oxygen flux through a permeable diffusion barrier. For an illuminated sensor area of 10 cm(2) and a spatial resolution of I mm at the sensor layer, the oxygen resolution is better than Delta PO2=0.3 torr in the absence of oxygen, and better than Delta PO2=4 torr at PO2= 160 torr.
引用
收藏
页码:110 / 115
页数:6
相关论文
共 23 条
[1]  
CLEGG RM, 1992, P SOC PHOTO-OPT INS, V1640, P448, DOI 10.1117/12.58237
[2]   Lifetime imaging of luminescent oxygen sensors based on all-solid-state technology [J].
Hartmann, P ;
Ziegler, W .
ANALYTICAL CHEMISTRY, 1996, 68 (24) :4512-4514
[3]   LUMINESCENCE QUENCHING BEHAVIOR OF AN OXYGEN SENSOR-BASED ON A RU(II) COMPLEX DISSOLVED IN POLYSTYRENE [J].
HARTMANN, P ;
LEINER, MJP ;
LIPPITSCH, ME .
ANALYTICAL CHEMISTRY, 1995, 67 (01) :88-93
[4]   RESPONSE CHARACTERISTICS OF LUMINESCENT OXYGEN SENSORS [J].
HARTMANN, P ;
LEINER, MJP ;
LIPPITSCH, ME .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) :251-257
[5]   Effects of polymer matrices on calibration functions of luminescent oxygen sensors based on porphyrin ketone complexes [J].
Hartmann, P ;
Trettnak, W .
ANALYTICAL CHEMISTRY, 1996, 68 (15) :2615-2620
[6]   FLOX - AN OXYGEN-FLUX-MEASURING SYSTEM USING A PHASE-MODULATION METHOD TO EVALUATE THE OXYGEN-DEPENDENT FLUORESCENCE LIFETIME [J].
HOLST, GA ;
KOSTER, T ;
VOGES, E ;
LUBBERS, DW .
SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) :231-239
[7]  
Huch A, 1979, CONTINUOUS TRANSCUTA
[8]   IN-VIVO VISUALIZATION OF OXYGEN-TRANSPORT IN MICROVASCULAR NETWORK [J].
ITOH, T ;
YAEGASHI, K ;
KOSAKA, T ;
KINOSHITA, T ;
MORIMOTO, T .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1994, 267 (05) :H2068-H2078
[9]   LUMINESCENT BAROMETRY IN WIND TUNNELS [J].
KAVANDI, J ;
CALLIS, J ;
GOUTERMAN, M ;
KHALIL, G ;
WRIGHT, D ;
GREEN, E ;
BURNS, D ;
MCLACHLAN, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (11) :3340-3347
[10]  
KHALIL GE, 1991, Patent No. 5043286