Suppression of current source noise with an atomic magnetometer

被引:16
|
作者
Shen, Liang [1 ,2 ,3 ]
Zhang, Rui [2 ,3 ,4 ,5 ]
Wu, Teng [2 ,3 ]
Peng, Xiang [2 ,3 ]
Yu, Song [1 ]
Chen, Jingbiao [2 ,3 ]
Guo, Hong [2 ,3 ]
机构
[1] Beijing Univ Posts & Telecommun, Inst Informat Photon & Opt Commun, Beijing 100876, Peoples R China
[2] Peking Univ, State Key Lab Adv Opt Commun Syst & Networks, Dept Elect, Beijing 100871, Peoples R China
[3] Peking Univ, Ctr Quantum Informat Technol, Beijing 100871, Peoples R China
[4] Natl Univ Def Technol, Coll Liberal Arts & Sci, Changsha 410073, Peoples R China
[5] Natl Univ Def Technol, Interdisciplinary Ctr Quantum Informat, Changsha 410073, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2020年 / 91卷 / 08期
基金
中国国家自然科学基金;
关键词
SELF-OSCILLATING MAGNETOMETER; RELAXATION;
D O I
10.1063/5.0009987
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper introduces a method for using an atomic magnetometer to suppress current source noise. In this method, the current noise is (i) suppressed with a lamp-pumped cesium atomic magnetometer and (ii) measured with a higher-sensitivity laser-pumped cesium atomic magnetometer. The current noise is compared before and after noise suppression at an output current of 50 mA, and the noise-suppression ratio is similar to 27 dB. The noise-suppression performance can be enhanced by improving the sensitivity of the atomic magnetometer and the ability to restrain the ambient magnetic noise. This method has practical applications in metrology and fundamental physics research.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Measurement of noise of current source by pump-probe atomic magnetometer
    Chen Da-Yong
    Miao Pei-Xian
    Shi Yan-Chao
    Cui Jing-Zhong
    Liu Zhi-Dong
    Chen Jiang
    Wang Kuan
    ACTA PHYSICA SINICA, 2022, 71 (02)
  • [2] Suppression of thermal coupling noise in the SERF atomic co-magnetometer
    Ruan, Jiasen
    Duan, Lihong
    Fu, Yang
    Yang, Lv
    Fan, Wenfeng
    Zhou, Xinxiu
    Quan, Wei
    RESULTS IN PHYSICS, 2023, 53
  • [3] Frequency noise suppression of the high-power pumping laser in the SERF atomic magnetometer
    Wang, Guiqi
    Duan, Lihong
    Ma, Ning
    Liu, Qingbo
    Hu, Zeyi
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2025, 36 (01)
  • [4] Characterizing current noise of commercial constant-current sources by using an optically pumped rubidium atomic magnetometer
    Zhao, Ni
    Zhang, Lulu
    Yang, Yongbiao
    He, Jun
    Wang, Yanhua
    Li, Tingyu
    Wang, Junmin
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2023, 94 (09):
  • [5] SUPPRESSION OF SIGNAL NOISE IN NUCLEAR MAGNETIC RESONANCE MAGNETOMETER
    ANUFRIEV, GS
    MAMYRIN, BA
    AFONIN, OF
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1964, (01) : 125 - &
  • [6] Analysis and suppression of magnetic field interference for SERF atomic magnetometer
    Jia, Le
    Li, Jianli
    Song, Xinda
    Qi, Shengjie
    Suo, Yuchen
    Long, Tengyue
    Dai, Yan
    Ning, Xiaolin
    MEASUREMENT, 2024, 227
  • [7] The suppression of the probe laser pumping effect in SERF atomic magnetometer
    Cao, Qian
    Zhai, Yueyang
    Huang, Binyue
    Li, Renjie
    Li, Jiajie
    FRONTIERS IN PHYSICS, 2023, 11
  • [8] Magnetic field interference suppression for minimized SERF atomic magnetometer
    Jia, Le
    Song, Xinda
    Suo, Yuchen
    Li, Jianli
    Long, Tengyue
    Ning, Xiaolin
    SENSORS AND ACTUATORS A-PHYSICAL, 2023, 351
  • [9] A Calibration Method of Three-Axis Magnetometer With Noise Suppression
    Song Zhongguo
    Zhang Jinsheng
    Zhang Xuehui
    Xi Xiaoli
    IEEE TRANSACTIONS ON MAGNETICS, 2014, 50 (11)
  • [10] Current sensor based on an atomic magnetometer for DC application
    李国祝
    辛青
    耿旭兴
    梁植
    梁尚清
    黄光明
    李高翔
    杨国卿
    ChineseOpticsLetters, 2020, 18 (03) : 35 - 38