Bayesian Speckle Tracking. Part I: An Implementable Perturbation to the Likelihood Function for Ultrasound Displacement Estimation

被引:24
作者
Byram, Brett [1 ]
Trahey, Gregg E. [1 ,2 ]
Palmeri, Mark [1 ]
机构
[1] Duke Univ, Dept Biomed Engn, Durham, NC 27706 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
基金
美国国家卫生研究院;
关键词
STRAIN ESTIMATION; ELASTOGRAPHY; PERFORMANCE;
D O I
10.1109/TUFFC.2013.2545
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Accurate and precise displacement estimation has been a hallmark of clinical ultrasound. Displacement estimation accuracy has largely been considered to be limited by the Cramer-Rao lower bound (CRLB). However, the CRLB only describes the minimum variance obtainable from unbiased estimators. Unbiased estimators are generally implemented using Bayes' theorem, which requires a likelihood function. The classic likelihood function for the displacement estimation problem is not discriminative and is difficult to implement for clinically relevant ultrasound with diffuse scattering. Because the classic likelihood function is not effective, a perturbation is proposed. The proposed likelihood function was evaluated and compared against the classic likelihood function by converting both to posterior probability density functions (PDFs) using a non-informative prior. Example results are reported for bulk motion simulations using a 6 lambda tracking kernel and 30 dB SNR for 1000 data realizations. The canonical likelihood function assigned the true displacement a mean probability of only 0.070 +/- 0.020, whereas the new likelihood function assigned the true displacement a much higher probability of 0.22 +/- 0.16. The new likelihood function shows improvements at least for bulk motion, acoustic radiation force induced motion, and compressive motion, and at least for SNRs greater than 10 dB and kernel lengths between 1.5 and 12 lambda.
引用
收藏
页码:132 / 143
页数:12
相关论文
共 33 条
[11]   Speckle decorrelation due to two-dimensional flow gradients [J].
Friemel, BH ;
Bohs, LN ;
Nightingale, KR ;
Trahey, GE .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (02) :317-327
[12]  
Jaynes E, 2003, PROBABILITY THEORY L, V1
[13]   CALCULATION OF PRESSURE FIELDS FROM ARBITRARILY SHAPED, APODIZED, AND EXCITED ULTRASOUND TRANSDUCERS [J].
JENSEN, JA ;
SVENDSEN, NB .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (02) :262-267
[14]   A parallelizable real-time motion tracking algorithm with applications to ultrasonic strain imaging [J].
Jiang, J. ;
Hall, T. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13) :3773-3790
[15]   A GENERALIZED SPECKLE TRACKING ALGORITHM FOR ULTRASONIC STRAIN IMAGING USING DYNAMIC PROGRAMMING [J].
Jiang, Jingfeng ;
Hall, Timothy J. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (11) :1863-1879
[16]   SPECKLE MOTION ARTIFACT UNDER TISSUE ROTATION [J].
KALLEL, F ;
BERTRAND, M ;
MEUNIER, J .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1994, 41 (01) :105-122
[17]   REAL-TIME TWO-DIMENSIONAL BLOOD-FLOW IMAGING USING AN AUTO-CORRELATION TECHNIQUE [J].
KASAI, C ;
NAMEKAWA, K ;
KOYANO, A ;
OMOTO, R .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (03) :458-464
[18]  
Kay S., 1993, Fundamentals of statistical processing, volume I: estimation theory, VI
[19]   A new elastographic method for estimation and imaging of lateral displacements, lateral strains, corrected axial strains and Poisson's ratios in tissues [J].
Konofagou, E ;
Ophir, J .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (08) :1183-1199
[20]   AN AXIAL VELOCITY ESTIMATOR FOR ULTRASOUND BLOOD-FLOW IMAGING, BASED ON A FULL EVALUATION OF THE DOPPLER EQUATION BY MEANS OF A 2-DIMENSIONAL AUTOCORRELATION APPROACH [J].
LOUPAS, T ;
POWERS, JT ;
GILL, RW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1995, 42 (04) :672-688