Tolerability of 24-Hour Intraocular Pressure Monitoring of a Pressure-sensitive Contact Lens

被引:43
作者
Lorenz, Katrin [1 ]
Korb, Christina [1 ]
Herzog, Nicola [1 ]
Vetter, Jan M. [1 ]
Elflein, Heike [1 ]
Keilani, Munir M. [1 ]
Pfeiffer, Norbert [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Ophthalmol, Univ Med Ctr, D-55131 Mainz, Germany
关键词
glaucoma; contact lens; intraocular pressure; 24-hour IOP profile; OPEN-ANGLE GLAUCOMA; INDEPENDENT RISK-FACTOR; NORMAL-TENSION GLAUCOMA; OCULAR HYPERTENSION; PROGRESSION; TONOMETER;
D O I
10.1097/IJG.0b013e318241b874
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate tolerability and safety of a new diagnostic device for 24-hour intraocular pressure monitoring in healthy subjects and age-matched glaucoma patients. Patients and Methods: Twenty healthy subjects (group 1) and 20 age-matched glaucoma patients (group 2) were included in this prospective, single-center, open, observational parallel group study. The SENSIMED Triggerfish Sensor is a soft disposable contact lens embedding a telemetry chip and strain gauge sensor for continuous intraocular pressure monitoring. The Sensor was placed in 1 eye for 24 hours. Tolerability was evaluated using a visual analog scale (range, 0 to 100; 0=no discomfort; 100=very severe discomfort). Safety parameters included best corrected visual acuity, pachymetry, epithelial defects, conjunctival erythema, and corneal topography. Results: Mean age was 61.7 years in group 1 and 65.0 years in group 2. Nineteen healthy subjects and 19 glaucoma patients (95%) completed the 24-hour wearing period. Early discontinuation resulted from pain (n=1) or inappropriate fitting of the sensor due to steep corneal radii (n=1). Mean tolerability was 21.8 in group 1 (range, 7 to 67) and 26.8 in group 2 (range, 0 to 71). Corneal epithelial staining (Modified Oxford scale, grade 0 to 4) changed from 0.4 (group 1) and 1.0 (group 2) at baseline to 1.8 (group 1) and 2.8 (group 2) after monitoring. No statistically significant differences could be detected between both groups. Conclusions: This new pressure-sensitive contact lens is tolerable and safe over a 24-hour wearing period in healthy subjects and glaucoma patients. Both normals and glaucoma patients had a similar safety and tolerability profile.
引用
收藏
页码:311 / 316
页数:6
相关论文
共 50 条
  • [41] Assessing the Functionality and Biocompatibility of a Wireless Contact Lens Sensor for 24 hour-intraocular Pressure Monitoring in Volunteers: Preliminary Results
    De Smedt, S. K., Sr.
    Mermoud, A.
    Roy, S.
    Schnyder, C.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [42] Twenty-four hour intraocular pressure monitoring with the SENSIMED Triggerfish contact lens: effect of body posture during sleep
    Beltran-Agullo, Laura
    Buys, Yvonne M.
    Jahan, Farzana
    Shapiro, Colin M.
    Flanagan, John G.
    Cheng, Jason
    Trope, Graham E.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2017, 101 (10) : 1323 - 1328
  • [43] High-sensitive microfluidic contact lens sensor for intraocular pressure visualized monitoring
    Yuan, Miao
    Liu, Zhiduo
    Wu, Xiaoting
    Gou, Haoyang
    Zhang, Yugui
    Ning, Xin
    Li, Weijun
    Yao, Zhaohui
    Wang, Yijun
    Pei, Weihua
    Chen, Hongda
    SENSORS AND ACTUATORS A-PHYSICAL, 2023, 354
  • [44] 24-Hour versus daytime intraocular pressure phasing in the management of patients with treated glaucoma
    Moodie, Jonathan
    Wilde, Craig
    Rotchford, Alan P.
    Vernon, Stephen A.
    King, Anthony J.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (08) : 999 - 1002
  • [45] Soft wearable contact lens sensor for continuous intraocular pressure monitoring
    Chen, Guo-Zhen
    Chan, Ion-Seng
    Leung, Leo K. K.
    Lam, David C. C.
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (09) : 1134 - 1139
  • [46] Non-invasive intraocular pressure monitoring with a contact lens engineered with a nanostructured polymeric sensing film
    Laukhin, Vladimir
    Sanchez, Irene
    Moya, Ana
    Laukhina, Elena
    Martin, Raul
    Ussa, Fernando
    Rovira, Concepcio
    Guimera, Anton
    Villa, Rosa
    Aguilo, Jordi
    Pastor, Jose-Carlos
    Veciana, Jaume
    SENSORS AND ACTUATORS A-PHYSICAL, 2011, 170 (1-2) : 36 - 43
  • [47] Capacitive contact lens sensor for continuous non-invasive intraocular pressure monitoring
    Chen, Guo-Zhen
    Chan, Ion-Seng
    Lam, David C. C.
    SENSORS AND ACTUATORS A-PHYSICAL, 2013, 203 : 112 - 118
  • [48] Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes
    Leonardi, Matteo
    Pitchon, Elie M.
    Bertsch, Arnaud
    Renaud, Philippe
    Mermoud, Andre
    ACTA OPHTHALMOLOGICA, 2009, 87 (04) : 433 - 437
  • [49] Efficacy of a Contact Lens Sensor for Monitoring 24-H Intraocular Pressure Related Patterns
    Mansouri, Kaweh
    Weinreb, Robert N.
    Liu, John H. K.
    PLOS ONE, 2015, 10 (05):
  • [50] Seasonal changes of 24-hour intraocular pressure rhythm in healthy Shanghai population
    Cheng, Jingyi
    Xiao, Ming
    Xu, Huan
    Fang, Shaobin
    Chen, Xu
    Kong, Xiangmei
    Sun, Xinghuai
    MEDICINE, 2016, 95 (31)