Incidence and causes of inappropriate detection and therapy by implantable defibrillators of cardioversion in patients with ventricular tachyarrhythmia

被引:4
作者
Chen, RH
Chen, KP
Wang, FZ
Hua, W
Chen, X
Zhang, S [1 ]
机构
[1] Fuwai Hosp, Chinese Acad Med Sci, Ctr Arrhythmia Diag & Treatment, Beijing 100037, Peoples R China
[2] Peking Union Med Coll, Beijing 100037, Peoples R China
关键词
cardioversion; inappropriate detection; inappropriate therapy; implantable cardioverter defibrillator; ventricular tachyarrhythmia;
D O I
10.1097/00029330-200604010-00006
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Implantable cardioverter defibrillator (ICD) is the only effective therapy in patients with life threatening ventricular arrhythmias. Inappropriate detection and therapy by ICDs are the most common causes of side effects that affect the quality of life in ICD recipients. This study evaluated the incidence and causes of inappropriate detection and therapy by ICDs in patients in our hospital. Methods From January 2000 to December 2005, fifty patients who received ICD implantation for ventricular arrhythmias for prevention of sudden cardiac death were evaluated in this study. Each ICD was programmed using clinical arrhythmic and cardiac data of the patient before discharge. Patients were followed up by standard schedule after implantation and all data retrieved from each device were collected and saved for further analysis. Results No arrhythmic event was detected in 12/50 (24%) patients during the period of follow-up. Among the remaining patients, 11 (22%) experienced inappropriate detections and therapies during follow-up in this study. ICD detected 383 ventricular tachyarrhythmia (VT) and 108 ventricular fibrillation (VF) episodes and delivered 678 therapies. In VT group, ICD delivered 413 antitachycardiac pacings (ATPs) and 118 shocks, among which 78 ATPs and 9 shocks were initiated by 55/383 (14.3%) inappropriate detections. In VF group ICD delivered 147 shocks, among which 56 shocks were initiated by 28/108 (26.9%) inappropriate detections. Overall, more than 50% of these episodes were caused by atrial fibrillation (AF) with rapid ventricular response, followed by electromagnetic or myopotential interference. In addition, most inappropriate therapies occurred within one year after ICD implantation. Conclusions About one fifth of patients experienced ICD inappropriate detection and therapy after implantation. The main cause was AF with rapid ventricular response, followed by electromagnetic or myopotential interference.
引用
收藏
页码:557 / 563
页数:7
相关论文
共 21 条
[1]   A randomized study of the prevention of sudden death in patients with coronary artery disease [J].
Buxton, AE ;
Lee, KL ;
Fisher, JD ;
Josephson, ME ;
Prystowsky, EN ;
Hafley, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (25) :1882-1890
[2]   Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study [J].
Deisenhofer, I ;
Kolb, C ;
Ndrepepa, G ;
Schreieck, J ;
Karch, M ;
Schmieder, S ;
Zrenner, B ;
Schmitt, C .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (02) :134-142
[3]   Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing [J].
Green, UB ;
Garg, A ;
Al-Kandari, F ;
Ungab, G ;
Tone, L ;
Feld, GK .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2003, 8 (01) :71-75
[4]  
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[5]   IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR THERAPY IN THE ABSENCE OF SIGNIFICANT SYMPTOMS - RHYTHM DIAGNOSIS AND MANAGEMENT AIDED BY STORED ELECTROGRAM ANALYSIS [J].
HOOK, BG ;
CALLANS, DJ ;
KLEIMAN, RB ;
FLORES, BT ;
MARCHLINSKI, FE .
CIRCULATION, 1993, 87 (06) :1897-1906
[6]   New ICD morphology criterion for differentiation of supraventricular and ventricular tachycardia [J].
Korte, T ;
Trappe, HJ ;
Grönefeld, G ;
Schulte, B ;
Wolpert, C ;
Meesmann, M ;
Böcker, D ;
Meininghaus, DG ;
Vogt, J ;
Stellbrink, C .
ZEITSCHRIFT FUR KARDIOLOGIE, 2000, 89 (11) :1019-1025
[7]   INCIDENCE OF ICD LEAD RELATED COMPLICATIONS DURING LONG-TERM FOLLOW-UP - COMPARISON OF EPICARDIAL AND ENDOCARDIAL ELECTRODE SYSTEMS [J].
KORTE, T ;
JUNG, W ;
SPEHL, S ;
WOLPERT, C ;
MOOSDORF, R ;
MANZ, M ;
LUDERITZ, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (11) :2053-2061
[8]   Prospective evaluation of catheter ablation in patients with implantable cardioverter defibrillators and multiple inappropriate ICD therapies due to atrial fibrillation and type I atrial flutter [J].
Korte, T ;
Niehaus, M ;
Meyer, O ;
Tebbenjohanns, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (07) :1061-1066
[9]   Hospital readmission after transvenous cardioverter/defibrillator implantation -: A single centre study [J].
Korte, T ;
Jung, W ;
Ostermann, G ;
Wolpert, C ;
Spehl, S ;
Esmailzadeh, B ;
Lüderitz, B .
EUROPEAN HEART JOURNAL, 2000, 21 (14) :1186-1191
[10]   Clinical experience with the new detection algorithms for atrial fibrillation of a defibrillator with dual chamber sensing and pacing [J].
Kühlkamp, V ;
Dörnberger, V ;
Mewis, C ;
Suchalla, R ;
Bosch, RF ;
Seipel, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (07) :905-915