Traumatic fracture of the polyethylene tibial post and cone in a posterior-stabilized total knee arthroplasty: A case report

被引:2
|
作者
Alkheraiji, Abdulaziz [1 ]
Borai, Sultana [2 ,4 ]
Alfadhil, Rheema [2 ]
AlJassir, Fawzi [3 ]
机构
[1] Al Majmaah Univ, Coll Med, Dept Orthoped Surg, Al Majmaah, Saudi Arabia
[2] King Saud Univ Med City, Orthoped Surg Resident, Riyadh, Saudi Arabia
[3] King Saud Univ Med City, Riyadh, Saudi Arabia
[4] King Saud Univ Med City, Orthoped Dept, Postal code 12372, Riyadh, Saudi Arabia
来源
INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS | 2022年 / 97卷
关键词
Total knee arthroplasty; Revision total knee arthroplasty; Polyethylene fracture; INSERT EXCHANGE; INSTABILITY; FAILURE;
D O I
10.1016/j.ijscr.2022.107437
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction and importance: Very few reports addressed polyethylene tibial post and cone fractures requiring revision surgery. Up to our knowledge, this particular pattern of wear, involving 2 fracture areas simultaneously, has not been previously described.Case presentation: A 70 year-old female developed a fracture involving both the tibial insert post and cone after a minor trauma after undergoing posterior-stabilized total knee replacement 4 years prior.Clinical discussion: This unique pattern of wear may occur even after minor trauma.Conclusion: We recommend having a low threshold to investigate minor trauma in this patient population, even years after undergoing primary knee arthroplasty.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Fracture of a Polyethylene Tibial Post in a Scorpio Posterior-Stabilized Knee Prosthesis
    Lim, Hong Chul
    Bae, Ji Hoon
    Hwang, Jin Ho
    Kim, Seung Joo
    Yoon, Ji Yeol
    CLINICS IN ORTHOPEDIC SURGERY, 2009, 1 (02) : 118 - 121
  • [2] Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis
    Chiu, YS
    Chen, WM
    Huang, CK
    Chiang, CC
    Chen, TH
    JOURNAL OF ARTHROPLASTY, 2004, 19 (08) : 1045 - 1049
  • [3] Fracture of the polyethylene tibial post in posterior stabilized (Insall Burstein II) total knee arthroplasty
    Mestha, P
    Shenava, Y
    D'Arcy, JC
    JOURNAL OF ARTHROPLASTY, 2000, 15 (06) : 814 - 815
  • [4] Fracture of the polyethylene tibial post in the posterior-stabilized total knee prosthesis: arthrographic and CT arthrographic diagnosis
    Yung Hsu
    Chien-Hung Lin
    Ginger H. F. Shu
    Tsyh-Jyi Hsieh
    Clement Kuen-Huang Chen
    Skeletal Radiology, 2019, 48 : 1145 - 1148
  • [5] Fracture of the polyethylene tibial post in the posterior-stabilized total knee prosthesis: arthrographic and CT arthrographic diagnosis
    Hsu, Yung
    Lin, Chien-Hung
    Shu, Ginger H. F.
    Hsieh, Tsyh-Jyi
    Chen, Clement Kuen-Huang
    SKELETAL RADIOLOGY, 2019, 48 (07) : 1145 - 1148
  • [6] Tibial Post Refracture in Posterior-Stabilized Total Knee Arthroplasty Following Isolated Tibial Insert Exchange
    Baek, Ji-Hoon
    Lee, Su Chan
    Ryu, Suengryol
    Nam, Chang Hyun
    ARTHROPLASTY TODAY, 2023, 19
  • [7] Highly Crosslinked Polyethylene in Posterior-Stabilized Total Knee Arthroplasty In Vitro Performance Evaluation of Wear, Delamination, and Tibial Post Durability
    Stoller, Alex P.
    Johnson, Todd S.
    Popoola, Oludele O.
    Humphrey, Steven M.
    Blanchard, Cheryl R.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (03) : 483 - 491
  • [8] The Cam-Post Dislocation in Posterior-Stabilized Total Knee Arthroplasty A Case Series
    Chen, Jeffrey S.
    Bi, Andrew S.
    Slover, James D.
    Marwin, Scott E.
    Deshmukh, Ajit J.
    BULLETIN OF THE HOSPITAL FOR JOINT DISEASES, 2022, 80 (04): : 263 - 268
  • [9] Recurrent dissociation of the tibial insert after mini-subvastus posterior-stabilized total knee arthroplasty: A case report
    In, Yong
    Sur, Yoo-Joon
    Won, Ho-Yeon
    Moon, Young-Seok
    KNEE, 2011, 18 (06) : 461 - 463
  • [10] Polyethylene post failure in posterior stabilized total knee arthroplasty
    Clarke, HD
    Math, KR
    Scuderi, GR
    JOURNAL OF ARTHROPLASTY, 2004, 19 (05) : 652 - 657