The Impact of Pump Settings on the Quality of Pulsatility

被引:20
作者
Rider, Alan R. [1 ]
Ressler, Noel M. [1 ]
Karkhanis, Tushar R. [1 ]
Kunselman, Allen R. [2 ]
Wang, Shigang [1 ,3 ]
Uendar, Akif [1 ,3 ,4 ]
机构
[1] Penn State Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Pediat Cardiac Res Labs,Dept Pediat, Hershey, PA USA
[2] Penn State Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Dept Hlth Evaluat Sci, Hershey, PA USA
[3] Penn State Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Dept Surg, Hershey, PA USA
[4] Penn State Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Dept Bioengn, Hershey, PA USA
关键词
CARDIOPULMONARY BYPASS PROCEDURES; NONPULSATILE PERFUSION; CPB MODEL; QUANTIFICATION; PRESSURE; BENEFITS; CANNULAE; TERMS;
D O I
10.1097/MAT.0b013e3181914e91
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The study objective was to evaluate the Jostra HL-20 roller pump under different baseflow and pump head settings with quantified energy values from pressure and flow waveforms, in a simulated pediatric bypass circuit. Pump flow rate was set at 800 mL/min for both pulsatile and nonpulsatile perfusion modes and the mean arterial pressure (MAP) of the pseudopatient was maintained at 40 mm Hg for each experiment. Pulsatile baseflow settings and pump head start points varied with each experiment. Pressure and flow waveforms were recorded at preoxygenator, precannula, and postcannula sites under each pump setting. A total of 91 experiments were performed (n = 7, nonpulsatile; n = 84, pulsatile). Increasing baseflow caused decreases in the mean circuit pressure and surplus hemodynamic energy (SHE) levels for all pump head start times. When increasing pump head start time within each baseflow, values for MAP and SHE increased significantly. Regardless of baseflow or pump head start time, values for mean circuit pressure and SHE were lower for nonpulsatile flow than for pulsatile flow. Total hemodynamic energy values were also significantly higher under pulsatile perfusion and increased pump start times while decreasing with increased baseflows in the circuit. This study concludes that decreased baseflows with increased pump head settings on the Jostra HL-20 roller pump could significantly increase quality of generated pulsatile energy. Further research is necessary to evaluate these various pump settings under microembolic loads and with different circuit components. ASAIO Journal 2009; 55:100-105.
引用
收藏
页码:100 / 105
页数:6
相关论文
共 13 条
[1]   Benefits of pulsatile perfusion on vital organ recovery during and after pediatric open heart surgery [J].
Alkan, Tijen ;
Akcevin, Atif ;
Uendar, Akif ;
Tuerkoglu, Halil ;
Paker, Tufan ;
Aytac, Aydin .
ASAIO JOURNAL, 2007, 53 (06) :651-654
[2]   An evaluation of the benefits of pulsatile versus nonpulsatile perfusion during cardiopulmonary bypass procedures in pediatric and adult cardiac patients [J].
Ji, Bingyang ;
Uendar, Akif .
ASAIO JOURNAL, 2006, 52 (04) :357-361
[3]  
Little R.C., 2006, SAS for Mixed Models, VSecond
[4]  
*MAQ CARD AG, 2000, JOSTR HL 20 HEART LU, P3
[5]   A performance evaluation of eight geometrically different 10 Fr pediatric arterial cannulae under pulsatile and nonpulsatile perfusion conditions in an infant cardiopulmonary bypass model [J].
Rider, Alan R. ;
Ji, Bingyang ;
Kunselman, Allen R. ;
Weiss, William J. ;
Myers, John L. ;
Uendar, Akif .
ASAIO JOURNAL, 2008, 54 (03) :306-315
[6]  
SHEPARD RB, 1966, ARCH SURG-CHICAGO, V93, P730
[7]   Comparison of four different pediatric 10F aortic cannulae during Pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass [J].
Uendar, Akif ;
Ji, Bingyang ;
Rider, Alan ;
Lukic, Branka ;
Kunselman, Allen R. ;
Weiss, William J. ;
Myers, John L. .
ASAIO JOURNAL, 2007, 53 (06) :778-784
[8]   Precise quantification of pulsatility is a necessity for direct comparisons of six different pediatric heart-lung machines in a neonatal CPB model [J].
Uendar, Akif ;
Eichstaedt, Harald C. ;
Masai, Takafumi ;
Bigley, Joyce E. ;
Kunselman, Allen R. .
ASAIO JOURNAL, 2005, 51 (05) :600-603
[9]   Comparison of hollow-fiber membrane oxygenators in terms of pressure drop of the membranes during normothermic and hypothermic cardiopulmonary bypass in neonates [J].
Ündar, A ;
Owens, WR ;
McGarry, MC ;
Surprise, DL ;
Kilpack, VD ;
Mueller, MW ;
McKenzie, ED ;
Fraser, CD .
PERFUSION-UK, 2005, 20 (03) :135-138
[10]   Comparison of six pediatric cardiopulmonary bypass pumps during pulsatile and nonpulsatile perfusion [J].
Ündar, A ;
Eichstaedt, HC ;
Masai, T ;
Yang, SQ ;
Bigley, JE ;
McGarry, MC ;
Mueller, M ;
Vaughn, WK ;
Fraser, CD .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 122 (04) :827-829