Projection imaging of photon beams by the Cerenkov effect

被引:89
作者
Glaser, Adam K. [1 ]
Davis, Scott C. [1 ]
McClatchy, David M. [1 ]
Zhang, Rongxiao [2 ]
Pogue, Brian W. [1 ,2 ]
Gladstone, David J. [3 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Dartmouth Coll, Dept Phys & Astron, Hanover, NH 03755 USA
[3] Dartmouth Hitchcock Med Ctr, Norris Cotton Canc Ctr, Lebanon, NH 03766 USA
基金
美国国家卫生研究院;
关键词
cerenkov; dosimetry; quality assurance; photon beam; PLASTIC SCINTILLATION DETECTORS; CHERENKOV EMISSION-SPECTROSCOPY; RADIATION-DOSE DISTRIBUTIONS; DOSIMETRY; TISSUE; ELECTRON; THERAPY; ARRAY; RAYS;
D O I
10.1118/1.4770286
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A novel technique for beam profiling of megavoltage photon beams was investigated for the first time by capturing images of the induced Cerenkov emission in water, as a potential surrogate for the imparted dose in irradiated media. Methods: A high-sensitivity, intensified CCD camera (ICCD) was configured to acquire 2D projection images of Cerenkov emission from a 4 x 4 cm(2) 6 MV linear accelerator (LINAC) x-ray photon beam operating at a dose rate of 400 MU/min incident on a water tank with transparent walls. The ICCD acquisition was gated to the LINAC sync pulse to reduce background light artifacts, and the measurement quality was investigated by evaluating the signal to noise ratio and measurement repeatability as a function of delivered dose. Monte Carlo simulations were used to derive a calibration factor for differences between the optical images and deposited dose arising from the anisotropic angular dependence of Cerenkov emission. Finally, Cerenkov-based beam profiles were compared to a percent depth dose (PDD) and lateral dose profile at a depth of d(max) from a reference dose distribution generated from the clinical Varian ECLIPSE treatment planning system (TPS). Results: The signal to noise ratio was found to be 20 at a delivered dose of 66.6 cGy, and proportional to the square root of the delivered dose as expected from Poisson photon counting statistics. A 2.1% mean standard deviation and 5.6% maximum variation in successive measurements were observed, and the Monte Carlo derived calibration factor resulted in Cerenkov emission images which were directly correlated to deposited dose, with some spatial issues. The dose difference between the TPS and PDD predicted by. Cerenkov measurements was within 20% in the buildup region with a distance to agreement (DTA) of 1.5-2 mm and +/- 3% at depths beyond d(max). In the lateral profile, the dose difference at the beam penumbra was within +/- 13% with a DTA of 0-2 mm, +/- 5% in the central beam region, and 2%-3% in the beam umbra. Conclusions: The results from this initial study demonstrate the first documented use of Cerenkov emission imaging to profile x-ray photon LINAC beams in water. The proposed modality has several potential advantages over alternative methods, and upon future refinement may prove to be a robust and novel dosimetry method. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4770286]
引用
收藏
页数:14
相关论文
共 42 条
[21]   Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms [J].
Glaser, Adam K. ;
Zhang, Rongxiao ;
Davis, Scott C. ;
Gladstone, David J. ;
Pogue, Brian W. .
OPTICS LETTERS, 2012, 37 (07) :1193-1195
[22]   Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels .1. Development of an optical scanner [J].
Gore, JC ;
Ranade, M ;
Maryanski, MJ ;
Schulz, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12) :2695-2704
[23]   A new water-equivalent 2D plastic scintillation detectors array for the dosimetry of megavoltage energy photon beams in radiation therapy [J].
Guillot, Mathieu ;
Beaulieu, Luc ;
Archambault, Louis ;
Beddar, Sam ;
Gingras, Luc .
MEDICAL PHYSICS, 2011, 38 (12) :6763-6774
[24]   OPTICAL-CONSTANTS OF WATER IN 200-NM TO 200-MUM WAVELENGTH REGION [J].
HALE, GM ;
QUERRY, MR .
APPLIED OPTICS, 1973, 12 (03) :555-563
[25]   Fiber-optic Cerenkov radiation sensor for proton therapy dosimetry [J].
Jang, Kyoung Won ;
Yoo, Wook Jae ;
Shin, Sang Hun ;
Shin, Dongho ;
Lee, Bongsoo .
OPTICS EXPRESS, 2012, 20 (13) :13907-13914
[26]  
Jelley J. V., 1958, CERENKOV RAD ITS APP
[27]   Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide [J].
Lambert, J. ;
Yin, Y. ;
McKenzie, D. R. ;
Law, S. ;
Suchowerska, N. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11) :3071-3080
[28]   Molecular Optical Imaging with Radioactive Probes [J].
Liu, Hongguang ;
Ren, Gang ;
Miao, Zheng ;
Zhang, Xiaofen ;
Tang, Xiaodong ;
Han, Peizhen ;
Gambhir, Sanjiv S. ;
Cheng, Zhen .
PLOS ONE, 2010, 5 (03)
[29]   MAGNETIC-RESONANCE-IMAGING OF RADIATION-DOSE DISTRIBUTIONS USING A POLYMER-GEL DOSIMETER [J].
MARYANSKI, MJ ;
SCHULZ, RJ ;
IBBOTT, GS ;
GATENBY, JC ;
XIE, J ;
HORTON, D ;
GORE, JC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (09) :1437-1455
[30]   In vivo Cerenkov luminescence imaging: a new tool for molecular imaging [J].
Mitchell, Gregory S. ;
Gill, Ruby K. ;
Boucher, David L. ;
Li, Changqing ;
Cherry, Simon R. .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2011, 369 (1955) :4605-4619