Respiratory motion guided four dimensional cone beam computed tomography: encompassing irregular breathing

被引:21
作者
O'Brien, Ricky T. [1 ]
Cooper, Benjamin J. [1 ,2 ]
Kipritidis, John [1 ]
Shieh, Chun-Chien [1 ,3 ]
Keall, Paul J. [1 ]
机构
[1] Univ Sydney, Sydney Med Sch, Radiat Phys Lab, Sydney, NSW 2006, Australia
[2] Canberra Hosp, Dept Med Phys, Canberra, ACT 2605, Australia
[3] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
基金
英国医学研究理事会; 澳大利亚国家健康与医学研究理事会;
关键词
respiratory motion guided 4DCBCT; RMG-4DCBCT; 4DCBCT; LUNG-TUMORS; CT;
D O I
10.1088/0031-9155/59/3/579
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Four dimensional cone beam computed tomography (4DCBCT) images suffer from angular under sampling and bunching of projections due to a lack of feedback between the respiratory signal and the acquisition system. To address this problem, respiratory motion guided 4DCBCT (RMG-4DCBCT) regulates the gantry velocity and projection time interval, in response to the patient's respiratory signal, with the aim of acquiring evenly spaced projections in a number of phase or displacement bins during the respiratory cycle. Our previous study of RMG-4DCBCT was limited to sinusoidal breathing traces. Here we expand on that work to provide a practical algorithm for the case of real patient breathing data. We give a complete description of RMG-4DCBCT including full details on how to implement the algorithms to determine when to move the gantry and when to acquire projections in response to the patient's respiratory signal. We simulate a realistic working RMG-4DCBCT system using 112 breathing traces from 24 lung cancer patients. Acquisition used phase-based binning and parameter settings typically used on commercial 4DCBCT systems (4 min acquisition time, 1200 projections across 10 respiratory bins), with the acceleration and velocity constraints of current generation linear accelerators. We quantified streaking artefacts and image noise for conventional and RMG-4DCBCT methods by reconstructing projection data selected from an oversampled set of Catphan phantom projections. RMG-4DCBCT allows us to optimally trade-off image quality, acquisition time and image dose. For example, for the same image quality and acquisition time as conventional 4DCBCT approximately half the imaging dose is needed. Alternatively, for the same imaging dose, the image quality as measured by the signal to noise ratio, is improved by 63% on average. C-arm cone beam computed tomography systems, with an acceleration up to 200 degrees /s(2), a velocity up to 100 degrees /s and the acquisition of 80 projections per second, allow the image acquisition time to be reduced to below 60 s. We have made considerable progress towards realizing a system to reduce projection clustering in conventional 4DCBCT imaging and hence reduce the imaging dose to the patient.
引用
收藏
页码:579 / 595
页数:17
相关论文
共 17 条
[1]   Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers [J].
Berbeco, RI ;
Mostafavi, H ;
Sharp, GC ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) :4481-4490
[2]   The use of a realistic VMAT delivery emulator to optimize dynamic machine parameters for improved treatment efficiency [J].
Boylan, C. J. ;
Rowbottom, C. G. ;
Mackay, R. I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (13) :4119-4133
[3]   Respiratory triggered 4D cone-beam computed tomography: A novel method to reduce imaging dose [J].
Cooper, Benjamin J. ;
O'Brien, Ricky T. ;
Balik, Salim ;
Hugo, Geoffrey D. ;
Keall, Paul J. .
MEDICAL PHYSICS, 2013, 40 (04)
[4]   Actively triggered 4d cone-beam CT acquisition [J].
Fast, Martin F. ;
Wisotzky, Eric ;
Oelfke, Uwe ;
Nill, Simeon .
MEDICAL PHYSICS, 2013, 40 (09)
[5]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[6]   Audio-visual biofeedback for respiratory-gated radiotherapy: Impact of audio instruction and audio-visual biofeedback on respiratory-gated radiotherapy [J].
George, Rohini ;
Chung, Theodore D. ;
Vedam, Sastry S. ;
Ramakrishnan, Viswanathan ;
Mohan, Radhe ;
Weiss, Elisabeth ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (03) :924-933
[7]   Obtaining breathing patterns from any sequential thoracic x-ray image set [J].
Kavanagh, Anthony ;
Evans, Philip M. ;
Hansen, Vibeke N. ;
Webb, Steve .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (16) :4879-4888
[8]   The comparative performance of four respiratory motion predictors for real-time tumour tracking [J].
Krauss, A. ;
Nill, S. ;
Oelfke, U. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) :5303-5317
[9]   Streaking artifacts reduction in four-dimensional cone-beam computed tomography [J].
Leng, Shuai ;
Zambelli, Joseph ;
Tolakanahalli, Ranjini ;
Nett, Brian ;
Munro, Peter ;
Star-Lack, Joshua ;
Paliwal, Bhudatt ;
Chena, Guang-Hong .
MEDICAL PHYSICS, 2008, 35 (10) :4649-4659
[10]   Optimizing 4D cone beam computed tomography acquisition by varying the gantry velocity and projection time interval [J].
O'Brien, Ricky T. ;
Cooper, Benjamin J. ;
Keall, Paul J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (06) :1705-1723