Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities

被引:71
作者
Thomson, Elizabeth A. [1 ]
Nuss, Kayla [1 ]
Comstock, Ashley [1 ]
Reinwald, Steven [1 ]
Blake, Sophie [1 ]
Pimentel, Richard E. [1 ]
Tracy, Brian L. [1 ]
Li, Kaigang [1 ]
机构
[1] Colorado State Univ, Dept Hlth & Exercise Sci, Ft Collins, CO 80523 USA
关键词
Heart rate measurement; exercise intensity; wearable devices; Apple Watch; Fitbit Charge HR 2; ACTIVITY MONITORS; CONSUMER-LEVEL; WORN; EQUIVALENCE; VALIDATION; ACCURACY; VALIDITY; ADULTS; TESTS;
D O I
10.1080/02640414.2018.1560644
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
This study compared heart rate (HR) measurements for the Fitbit Charge HR 2 (Fitbit) and the Apple Watch devices with HR measurements for electrocardiogram (ECG). Thirty young adults (15/15 females/males, age 23.5 +/- 3.0 years) completed the Bruce Protocol. HR measurements were recorded from the ECG and both devices every minute. Average HR for each participant was calculated for very light, light, moderate, vigorous and very vigorous intensities based on ECG-measured HR. A concordance correlation coefficient (CCC) was calculated to examine the strength of the relationship between ECG measured HR and HR measured by each device. Relative error rates (RER) were also calculated to indicate the difference between each device and ECG. An equivalence test was conducted to examine the equivalence of HRs measured by devices and ECG. The Apple Watch showed lower RER (2.4-5.1%) compared with the Fitbit (3.9-13.5%) for all exercise intensities. For both devices, the strongest relationship with ECG-measured HR was found for very light PA with very high CCC (>.90) and equivalence. The strength of the relationship declined as exercise intensity increased for both devices. These findings indicate that the accuracy of real-time HR monitoring by the Apple Watch and Fitbit Charge HR2 is reduced as exercise intensity increases.
引用
收藏
页码:1411 / 1419
页数:9
相关论文
共 34 条
[1]  
American College of Sports Medicine, 2013, ACSMS GUIDELINES EXE
[2]  
[Anonymous], 2016, J FIT RES
[3]  
Apple Support, 2017, APPL WATCH MEAS YOUR
[4]   MAXIMAL OXYGEN INTAKE AND NOMOGRAPHIC ASSESSMENT OF FUNCTIONAL AEROBIC IMPAIRMENT IN CARDIOVASCULAR DISEASE [J].
BRUCE, RA ;
KUSUMI, F ;
HOSMER, D .
AMERICAN HEART JOURNAL, 1973, 85 (04) :546-562
[5]   The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers [J].
Cadmus-Bertram, Lisa ;
Gangnon, Ronald ;
Wirkus, Emily J. ;
Thraen-Borowski, Keith M. ;
Gorzelitz-Liebhauser, Jessica .
ANNALS OF INTERNAL MEDICINE, 2017, 166 (08) :610-+
[6]  
Desilets P. A., 2016, INT J EXERCISE SCI, V9, P2
[7]  
Dooley Erin E, 2017, JMIR Mhealth Uhealth, V5, pe34, DOI 10.2196/mhealth.7043
[8]   The Validity of Consumer-Level Activity Monitors in Healthy Older Adults in Free-Living Conditions [J].
Farina, Nicolas ;
Lowry, Ruth G. .
JOURNAL OF AGING AND PHYSICAL ACTIVITY, 2018, 26 (01) :128-135
[9]   The validity of consumer-level, activity monitors in healthy adults worn in free-living conditions: a cross-sectional study [J].
Ferguson, Ty ;
Rowlands, Alex V. ;
Olds, Tim ;
Maher, Carol .
INTERNATIONAL JOURNAL OF BEHAVIORAL NUTRITION AND PHYSICAL ACTIVITY, 2015, 12
[10]  
Fitbit, 2017, KNOW YOUR HEART PURE