Evaluation of 39 medical implants at 7.0T

被引:22
作者
Feng, David X. [1 ]
Mccauley, Joseph P. [1 ]
Morgan-Curtis, Fea K. [1 ]
Salam, Redoan A. [1 ]
Pennell, David R. [2 ]
Loveless, Mary E. [1 ]
Dula, Adrienne N. [2 ,3 ]
机构
[1] Vanderbilt Univ, Sch Sci & Math, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Inst Imaging Sci, Nashville, TN 37235 USA
[3] Vanderbilt Univ, Med Ctr, Dept Radiol & Radiol Sci, Nashville, TN 37232 USA
基金
美国国家卫生研究院;
关键词
7; T; MAGNETIC-FIELD; TESLA MRI; DEVICES; DISEASE; TERMINOLOGY; INFORMATION; EPILEPSY; SAFETY;
D O I
10.1259/bjr.20150633
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: With increased signal to noise ratios, 7.0-T MRI has the potential to contribute unique information regarding anatomy and pathophysiology of a disease. However, concerns for the safety of subjects with metallic medical implants have hindered advancement in this field. The purpose of the present research was to evaluate the MRI safety for 39 commonly used medical implants at 7.0T. Methods: Selected metallic implants were tested for magnetic field interactions, radiofrequency-induced heating and artefacts using standardized testing techniques. Results: 5 of the 39 implants tested may be unsafe for subjects undergoing MRI at 7.0T. Conclusion: Implants were deemed either "MR Conditional" or "MR Unsafe" for the 7.0-T MRI environment. Further research is needed to expand the existing database categorizing implants that are acceptable for patients referred for MRI examinations at 7.0T. Advances in knowledge: Lack of MRI testing for common metallic medical implants limits the translational potential of 7.0-T MRI. For safety reasons, patients with metallic implants are not allowed to undergo a 7.0-T MRI scan, precluding part of the population that can benefit from the detailed resolution of ultra-high-field MRIs. This investigation provides necessary MRI testing of common medical implants at 7.0T.
引用
收藏
页数:10
相关论文
共 30 条
  • [1] American Society for Testing and Materials, 2013, STAND PRACT MARK MED
  • [2] American Society for Testing Materials, 2011, STAND TEST METH MEAS
  • [3] ASTM, STAND TEST METH EV M
  • [4] ASTM, 2012, STAND TEST METH MEAS
  • [5] ASTM, 2006, Standard test method for measuring pavement macrotexture depth using a
  • [6] 7 Tesla MRI in cerebral small vessel disease
    Benjamin, Philip
    Viessmann, Olivia
    MacKinnon, Andrew D.
    Jezzard, Peter
    Markus, Hugh S.
    [J]. INTERNATIONAL JOURNAL OF STROKE, 2015, 10 (05) : 659 - 664
  • [7] Ultra-High-Field MR Imaging in Polymicrogyria and Epilepsy
    De Ciantis, A.
    Barkovich, A. J.
    Cosottini, M.
    Barba, C.
    Montanaro, D.
    Costagli, M.
    Tosetti, M.
    Biagi, L.
    Dobyns, W. B.
    Guerrini, R.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (02) : 309 - 316
  • [8] Feasibility of Using Ultra-High Field (7 T) MRI for Clinical Surgical Targeting
    Duchin, Yuval
    Abosch, Aviva
    Yacoub, Essa
    Sapiro, Guillermo
    Harel, Noam
    [J]. PLOS ONE, 2012, 7 (05):
  • [9] Assessment of MRI Issues at 7 T for 28 Implants and Other Objects
    Dula, Adrienne N.
    Virostko, John
    Shellock, Frank G.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (02) : 401 - 405
  • [10] Application of Chemical Exchange Saturation Transfer (CEST) MRI for Endogenous Contrast at 7 Tesla
    Dula, Adrienne N.
    Smith, Seth A.
    Gore, John C.
    [J]. JOURNAL OF NEUROIMAGING, 2013, 23 (04) : 526 - 532