New generation continuous cardiac output monitoring from carbon dioxide elimination

被引:10
作者
Peyton, Philip J. [1 ,2 ]
Wallin, Mats [3 ,4 ]
Hallback, Magnus [3 ]
机构
[1] Univ Melbourne, Melbourne Med Sch, Anaesthesia Perioperat & Pain Med Unit, Heidelberg, Vic 3084, Australia
[2] Austin Hlth, Dept Anaesthesia, Heidelberg, Vic 3084, Australia
[3] Maquet Crit Care AB, Rontgenvagen 2, S-17154 Solna, Sweden
[4] Karolinska Inst, Dept Physiol & Pharmacol, Sect Anesthesiol & Intens Care, Stockholm, Sweden
关键词
Monitoring; Carbon dioxide; Cardiac output; Perioperative; CONTINUOUS CAPNODYNAMIC METHOD; PULMONARY BLOOD-FLOW; MAJOR SURGERY; HEMODYNAMIC OPTIMIZATION; PRECISION STATISTICS; FLUID MANAGEMENT; METAANALYSIS; THERMODILUTION; PERFORMANCE; ACCURACY;
D O I
10.1186/s12871-019-0699-5
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: There is continuing interest among clinicians in the potential for advanced hemodynamic monitoring and "goal directed" intravenous fluid administration guided by minimally-invasive cardiac output measurement to reduce complication rates in high risk patients undergoing major surgery. However, the adoption of the available technologies has been limited, due to cost, complexity and reliability of measurements provided. We review progress in the development of new generation methods for continuous non-invasive monitoring of cardiac output from measurement of carbon dioxide elimination in ventilated patients using the Differential Fick method. Main text: The history and underlying theoretical basis are described, and its recent further development and implementation using modern generation anesthesia monitoring and delivery systems by two separate but parallel methods, termed "Capnotracking" and "Capnodynamics". Both methods generate breath-by-breath hands-free cardiac output monitoring from changes in carbon dioxide elimination produced by automatic computerized modulation of respiratory rate delivered by an electronic ventilator. Extensive preclinical validation in animal models of hemodynamic instability, with implanted ultrasonic flow probes for gold standard reference measurements, shows this approach delivers reliable, continuous cardiac output measurement in real time. The accuracy and precision of measurement by the Capnodynamic method were maintained under a wide range of both hemodynamic and respiratory conditions, including inotropic stimulation, vasodilatation, hemorrhage, caval compression, alveolar lavage, changes in tidal volume and positive end-expiratory pressure, and hypercapnia, with only brief derangement observed in a model of lower body ischemia involving release of prolonged aortic occlusion by an intra-aortic balloon. Phase 2 testing of a Capnotracking system in patients undergoing cardiac surgery and liver transplantation has achieved a percentage error of agreement with thermodilution of +/- 38.7% across a wide range of hemodynamic states. Conclusions: Progress in development of these technologies suggest that a robust, automated and reliable method of non-invasive cardiac output monitoring from capnography is close at hand for use in major surgery and critical care. The great advantage of this approach is that it can be fully integrated into the anesthesia machine and ventilator, using components that are already standard in modern anesthesia and intensive care workstations, and should be virtually hands-free and automatic.
引用
收藏
页数:10
相关论文
共 35 条
[1]   Comparison of invasive and less-invasive techniques of cardiac output measurement under different haemodynamic conditions in a pig model [J].
Bajorat, J ;
Hofmockel, R ;
Vagts, DA ;
Janda, M ;
Pohl, B ;
Beck, C ;
Noeldge-Schomburg, G .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2006, 23 (01) :23-30
[2]   Wet, dry or something else? [J].
Bellamy, M. C. .
BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (06) :755-757
[3]   Measurement of cardiac output before and after cardiopulmonary bypass:: Comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing [J].
Botero, M ;
Kirby, D ;
Lobato, EB ;
Staples, ED ;
Gravenstein, N .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2004, 18 (05) :563-572
[4]   HOW DO CHANGES IN EXHALED CO2 MEASURE CHANGES IN CARDIAC OUTPUT? A NUMERICAL ANALYSIS MODEL [J].
Breen, Peter H. .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2010, 24 (06) :413-419
[5]  
Capek JM, 1988, ENCY MED DEVICES INS, P1309
[6]   A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM) [J].
Chong, S. W. ;
Peyton, P. J. .
ANAESTHESIA, 2012, 67 (11) :1266-1271
[7]   Perioperative Fluid Management Strategies in Major Surgery: A Stratified Meta-Analysis [J].
Corcoran, Tomas ;
Rhodes, Julia Emma Joy ;
Clarke, Sarah ;
Myles, Paul S. ;
Ho, Kwok M. .
ANESTHESIA AND ANALGESIA, 2012, 114 (03) :640-651
[8]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[9]  
Critchley LAH, 2011, ANESTHESIOLOGY, V114, P1245, DOI 10.1097/ALN.0b013e318215e1e5
[10]   DETERMINATION OF PVBARCO2 FROM THE EXPONENTIAL CO2 RISE DURING REBREATHING [J].
DEFARES, JG .
JOURNAL OF APPLIED PHYSIOLOGY, 1958, 13 (02) :159-164