A 0.83-μW QRS Detection Processor Using Quadratic Spline Wavelet Transform for Wireless ECG Acquisition in 0.35-μm CMOS

被引:95
作者
Ieong, Chio-In [1 ]
Mak, Pui-In [1 ]
Lam, Chi-Pang [1 ]
Dong, Cheng [1 ]
Vai, Mang-I [1 ]
Mak, Peng-Un [1 ]
Pun, Sio-Hang [1 ]
Wan, Feng [1 ]
Martins, Rui P. [1 ]
机构
[1] Univ Macau, Fac Sci & Technol, Dept Elect & Comp Engn, Taipa 999078, Peoples R China
关键词
QRS detection; quadratic spline wavelet transform; wavelet transform; wearable electrocardiograph (ECG) device; wireless ECG monitoring; NOTCH FILTER;
D O I
10.1109/TBCAS.2012.2188798
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Healthcare electronics count on the effectiveness of the on-patient signal preprocessing unit to moderate the wireless data transfer for better power efficiency. In order to reduce the system power in long-time ECG acquisition, this work describes an on-patient QRS detection processor for arrhythmia monitoring. It extracts the concerned ECG part, i.e., the RR-interval between the QRS complex for evaluating the heart rate variability. The processor is structured by a scale-3 quadratic spline wavelet transform followed by a maxima modulus recognition stage. The former is implemented via a symmetric FIR filter, whereas the latter includes a number of feature extraction steps: zero-crossing detection, peak (zero-derivative) detection, threshold adjustment and two finite state machines for executing the decision rules. Fabricated in 0.35-mu m CMOS the 300-Hz processor draws only 0.83 mu W, which is favorably comparable with the prior arts. In the system tests, the input data is placed via an on-chip 10-bit SAR analog-to-digital converter, while the output data is emitted via an off-the-shelf wireless transmitter (TI CC2500) that is configurable by the processor for different data transmission modes: 1) QRS detection result, 2) raw ECG data or 3) both. Validated with all recordings from the MIT-BIH arrhythmia database, 99.31% sensitivity and 99.70% predictivity are achieved. Mode 1 with solely the result of QRS detection exhibits 6 x reduction of system power over modes 2 and 3.
引用
收藏
页码:586 / 595
页数:10
相关论文
共 30 条
[1]   ECG beat detection using filter banks [J].
Afonso, VX ;
Tompkins, WJ ;
Nguyen, TQ ;
Luo, S .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (02) :192-202
[2]  
[Anonymous], MSP430 WORLDS LOW PO
[3]  
[Anonymous], MSP430F20X1 MSP430F2
[4]  
Ashouei M., 2011, 2011 IEEE International Solid-State Circuits Conference (ISSCC 2011), P332, DOI 10.1109/ISSCC.2011.5746341
[5]   Discrete Wavelet Transform-Based Time Series Analysis and Mining [J].
Chaovalit, Pimwadee ;
Gangopadhyay, Aryya ;
Karabatis, George ;
Chen, Zhiyuan .
ACM COMPUTING SURVEYS, 2011, 43 (02)
[6]   A Wireless Capsule Endoscope System With Low-Power Controlling and Processing ASIC [J].
Chen, Xinkai ;
Zhang, Xiaoyu ;
Zhang, Linwei ;
Li, Xiaowen ;
Qi, Nan ;
Jiang, Hanjun ;
Wang, Zhihua .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2009, 3 (01) :11-22
[7]  
Gari F. A., 2006, ADV METHODS TOOLS EC
[8]  
Haddad SAP, 2003, PROCEEDINGS OF THE 2003 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOL I, P121
[9]   QUANTITATIVE INVESTIGATION OF QRS DETECTION RULES USING THE MIT/BIH ARRHYTHMIA DATABASE [J].
HAMILTON, PS ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (12) :1157-1165
[10]   Heart rate variability: Measurement and clinical utility [J].
Kleiger, RE ;
Stein, PK ;
Bigger, JT .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2005, 10 (01) :88-101