共 7 条
Third-generation computed tomography with energy information of X-rays using a CdTe flat panel detector
被引:1
作者:
Kanno, Ikuo
[1
]
Imamura, Ryo
[1
]
Minami, Yuko
[1
]
Ohtaka, Masahiko
[2
]
Hashimoto, Makoto
[2
]
Ara, Kuniaki
[2
]
Onabe, Hideaki
[3
]
机构:
[1] Kyoto Univ, Dept Nucl Engn, Sakyo Ku, Kyoto 6068501, Japan
[2] Japan Atom Energy Agcy, Oarai, Ibaraki 3111393, Japan
[3] Raytech Corp, Utsunomiya, Tochigi 3210904, Japan
来源:
关键词:
X-ray;
Energy-resolved CT;
Response function;
Unfolding;
Flat panel detector;
CONTRAST-MEDIA;
D O I:
10.1016/j.nima.2011.11.085
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
The present authors developed the transXend detector, which measures incident X-rays as current and also gives their energy distribution after an unfolding process with using previously obtained response functions. This detector enables X-ray energy-resolved computed tomography (CT). To incorporate the transXend detector into a third-generation CT scanner, a flat panel detector (FPD) made of CdTe pixels is employed with placing a variable number of aluminum absorbers in front of the FPD. Although changing the number of absorbers can be time consuming, the time necessary to generate a CT image is much less when compared with image acquisition by first-generation CT. (C) 2011 Elsevier B.V. All rights reserved.