Ultra-low dose chest computed tomography: Effect of iterative reconstruction levels on image quality

被引:30
作者
Afadzi, Mercy [1 ]
Lysvik, Elisabeth Kirkeby [1 ]
Andersen, Hilde Kjernlie [1 ]
Martinsen, Anne Catrine T. [1 ,2 ]
机构
[1] Oslo Univ Hosp, Dept Diagnost Phys, Oslo, Norway
[2] Univ Oslo, Dept Phys, Oslo, Norway
关键词
Ultra-low dose; ASIR-V; Image quality; Chest CT; Iterative reconstruction; FILTERED BACK-PROJECTION; NOISE POWER SPECTRA; CT; REDUCTION; TEXTURE;
D O I
10.1016/j.ejrad.2019.02.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To optimize image quality and radiation dose of chest CT with respect to various iterative reconstruction levels, detector collimations and body sizes. Method: A Kyoto Kagaku Lungman with and without extensions was scanned using fixed ultra-low doses of 0.25, 0.49 and 0.74 mGy CTDIvol, and collimations of 40 and 80 mm. Images were reconstructed with the lung kernel, filtered back projection (FBP) and different ASIR-V levels (10-100%). Contrast-to-noise ratios (CNR) were calculated for 12 mm simulated lesions of different densities in the lung. Image noise, signal-to-noise ratios (SNR), variations in Hounsfield units (HU), noise power spectrum (NPS) and noise texture deviations (NTD) were evaluated for all reconstructions. NTD was calculated as percentage of pixels outside 3 standard deviations to evaluate IR-specific artefacts. Results: Compared to the FBP, image noise reduced (5-55%) with ASIR-V levels irrespective of dose or collimation. SNR correlated positively (r >= 0.925, p <= 0.001) with ASIR-V levels at all doses, collimations, and phantom sizes. ASIR-V enhanced the CNR of the lesion with the lowest contrast from 12.7-42.1 (0-100% ASIRV) at 0.74 mGy with 40 mm collimation. As expected, higher SNR and CNR were measured in the smaller phantom than the bigger phantom. Uniform HU were observed between FBP and ASIR-V levels at all doses, collimations, and phantom sizes. NPS curves left-shifted towards lower frequencies at increasing levels of ASIR-V irrespective of collimation. A positive correlation (r >= 0.946, p >= 0.001) was observed between NTD and ASIR-V levels. NTD of the FBP was not significantly (p <= 0.087) different from NTD of ASIR-V <= 20%. The data from the NPS and NTD indicates a blotchier and coarser noise texture at higher levels of ASIR-V, especially at 100% ASIR-V. Conclusion: In comparison with the FBP technique, ASIR-V enhanced quantitative image quality parameters at all ultra-low doses tested. Moreover, the use of ASIR-V showed consistency with body size and collimation. Hence, ASIR-V may be useful for improving image quality of chest CT at ultra-low doses.
引用
收藏
页码:62 / 68
页数:7
相关论文
共 39 条
[1]   Image quality with iterative reconstruction techniques in CT of the lungs-A phantom study [J].
Andersen, Hilde Kjernlie ;
Volgyes, David ;
Martinsen, Anne Latrine Trgde .
EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2018, 5 :35-40
[2]  
[Anonymous], 2012, J ICRU
[3]  
Boas F.E., 2012, Imaging Med, V4, P229
[4]   Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta [J].
Boedeker, K. L. ;
Cooper, V. N. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4027-4046
[5]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[6]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[7]   New adaptive statistical iterative reconstruction ASiR-V: Assessment of noise performance in comparison to ASiR [J].
De Marco, Paolo ;
Origgi, Daniela .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02) :275-286
[8]   Practical considerations for noise power spectra estimation for clinical CT scanners [J].
Dolly, Steven ;
Chen, Hsin-Chen ;
Anastasio, Mark ;
Mutic, Sasa ;
Li, Hua .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (03) :392-407
[9]  
Fan J., 2014, BENEFITS ASIR 5 RECO
[10]   A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom [J].
Friedman, Saul N. ;
Fung, George S. K. ;
Siewerdsen, Jeffrey H. ;
Tsui, Benjamin M. W. .
MEDICAL PHYSICS, 2013, 40 (05)