Precision and reproducibility of measurements of human corneal thickness with rapid optical low-coherence reflectometry (OLCR)

被引:28
作者
Böhnke, M [1 ]
Masters, BR
Wälti, R
Ballif, JJ
Chavanne, P
Gianotti, R
Salathé, RP
机构
[1] Univ Bern, Dept Ophthalmol, CH-3010 Bern, Switzerland
[2] Swiss Fed Inst Technol, Appl Opt Lab, CH-1015 Lausanne, Switzerland
关键词
optical low-coherence reflectometry; human corneal thickness; precision and reproducibility of corneal pachometry;
D O I
10.1117/1.429901
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study investigates the precision and intraindividual variability of a Clinical optical pachometer based on low-coherence reflectometry, which was used to measure the central thickness of a human cornea in vivo. The instrument, attached to a slit lamp, is a single mode fiber optic based Michelson interferometer with a high repetition rate as previously described. The same operator performed ten sets of measurements on the same subject, each consisting of 20 consecutive scans, on each day for three consecutive days. By computing the means from every scan series, the thickness of the central cornea with optical pachometry was found to be 519.6 +/- 1.2 (range 518-521) mu m on day 1, 519.9 +/- 0.9 (range 519-521) mu m on day 2, and 523.8 +/- 0.65 (range 523-525) mu m on day 3. The thickness values on day 3, where the subject suffered from a cold without clinical ocular involvement, were different from the two precious days (p < 0.001, one way analysis of variance). Optical low-coherence reflectometry measurements of corneal thickness can be performed with high precision of about 1 mu m and a high intra- and intersession reproducibility. (C) 1999 Society of Photo-Optical Instrumentation Engineers. [S1083-3668(99)01301-5].
引用
收藏
页码:152 / 156
页数:5
相关论文
共 25 条
  • [1] Anderson D.R., 1997, OPTICAL TIME DOMAIN
  • [2] AZEN SP, 1979, INVEST OPHTH VIS SCI, V18, P535
  • [3] Rapid and scalable scans at 21 m/s in optical low-coherence reflectometry
    Ballif, J
    Gianotti, R
    Chavanne, P
    Walti, R
    Salathe, RP
    [J]. OPTICS LETTERS, 1997, 22 (11) : 757 - 759
  • [4] Bohnke M, 1997, GER J OPHTHALMOL, V5, P338
  • [5] Bohnke M, 1998, J REFRACT SURG, V14, P140
  • [6] CHAVANNE P, 1994, P APPL OPT OPT C YOR, P399
  • [7] Drexler W, 1997, INVEST OPHTH VIS SCI, V38, P1304
  • [8] Fercher A F, 1996, J Biomed Opt, V1, P157, DOI 10.1117/12.231361
  • [9] VARIABILITY OF ULTRASONIC PACHOMETRY
    GORDON, A
    BOGGESS, EA
    MOLINARI, JF
    [J]. OPTOMETRY AND VISION SCIENCE, 1990, 67 (03) : 162 - 165
  • [10] INTERFEROMETRIC MEASUREMENT OF CORNEAL THICKNESS WITH MICROMETER PRECISION
    HITZENBERGER, CK
    BAUMGARTNER, A
    DREXLER, W
    FERCHER, AF
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1994, 118 (04) : 468 - 476