摘要
A real human skull was selected to be a mold for the construction of an anthropomorphic head phantom with a mixture of dolomite and polymethyl methacrylate (PMMA). Using linear attenuation coefficients, we show that it is possible to use dolomite as a bone simulator as long as the proportion of the mixture is 1:1. Acrylic tubes were placed in the phantom constructed to enable the insertion of the ionization chamber to estimate the effective dose. Values for a typical head computed tomography examination found in the literature vary from 0.9 to 4.0 mSv. Dosimetric studies showed that the effective dose for the anthropomorphic phantom was (2.70 +/- 0.03) mSv and for the geometric PMMA phantom (3.67 +/- 0.04) mSv, values which are in agreement with the intervals reported in the literature. The investment to produce the phantom was approximately US$160.00. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:203 / 208
页数:6
相关论文
共 50 条
  • [1] Automatic validation of the gantry tilt in a computed tomography scanner using a head polymethyl methacrylate phantom
    Noviliawati, Rakhma
    Anam, Choirul
    Sutanto, Heri
    Dougherty, Geoff
    Mak'ruf, Muhammad Ridha
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2021, 27 (01): : 57 - 62
  • [2] MAMMOGRAPHY DOSIMETRY USING AN IN-HOUSE DEVELOPED POLYMETHYL METHACRYLATE PHANTOM
    Sharma, Reena
    Sharma, Sunil Dutt
    Mayya, Y. S.
    Chourasiya, G.
    RADIATION PROTECTION DOSIMETRY, 2012, 151 (02) : 379 - 385
  • [3] Measurement of computed tomography modulation transfer function with a novel polymethyl methacrylate phantom
    Svenson, J.
    Irvine, M. A.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, : 1773 - 1780
  • [4] Comparison of computed tomography dose index in polymethyl methacrylate and nylon dosimetry phantoms
    Sookpeng, Supawitoo
    Cheebsumon, Patsuree
    Pengpan, Thanyawee
    Martin, Colin
    JOURNAL OF MEDICAL PHYSICS, 2016, 41 (01) : 45 - 51
  • [5] Radiation dose evaluation of dental cone beam computed tomography using an anthropomorphic adult head phantom
    Wu, Jay
    Shih, Cheng-Ting
    Ho, Chang-hung
    Liu, Yan-Lin
    Chang, Yuan-Jen
    Chao, Max Min
    Hsu, Jui-Ting
    RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 287 - 291
  • [6] Development, Construction, and Evaluation of an Alternative Dosimetry Phantom for Computed Tomography
    Lubis, Lukmanda Evan
    Najmah, Windi Dliya
    Muliyanti, Yuni
    Hariyati, Ika
    Ryangga, Dea
    Mart, Terry
    Bosmans, Hilde
    Soejoko, Djarwani Soeharso
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (04) : 402 - 408
  • [7] An investigation of internal scatter during computed tomography using an anthropomorphic phantom
    Van Wyk, Bronwin
    Ngobese, Zanele Eunice
    Lewis, Shantel
    HEALTH AND TECHNOLOGY, 2024, 14 (04) : 747 - 752
  • [8] Evaluation of electrical impedance and computed tomography fusion algorithms using an anthropomorphic phantom
    Chugh, Brige Paul
    Krishnan, Kalpagam
    Liu, Jeff
    Kohli, Kirpal
    MEDICAL PHYSICS, 2014, 41 (08) : 8 - 8
  • [9] Computed tomography phantom for radiochromic film dosimetry
    Gotanda, R.
    Katsuda, T.
    Gotanda, T.
    Eguchi, M.
    Takewa, S.
    Tabuchi, A.
    Yatake, H.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) : 194 - 199
  • [10] Computed tomography phantom for radiochromic film dosimetry
    R. Gotanda
    T. Katsuda
    T. Gotanda
    M. Eguchi
    S. Takewa
    A. Tabuchi
    H. Yatake
    Australasian Physics & Engineering Sciences in Medicine, 2007, 30 : 194 - 199