Design and evaluation of a single-pivot supported centrifugal blood pump

被引:11
|
作者
Yoshino, M
Uemura, T
Takahashi, T
Watanabe, T
Hoshi, H
Ohuchi, K
Nakamura, M
Fujita, H
Sakamoto, T
Takatani, S
机构
[1] Tokyo Med & Dent Univ, Dept Artificial Organs, Inst Biomat & Bioengn, Chiyoda Ku, Tokyo, Japan
[2] Nihon Univ, Coll Sci & Technol, Dept Mech Engn, Tokyo 102, Japan
[3] Shibaura Inst Technol, Dept Engn Mech, Tokyo 108, Japan
[4] Tokyo Denki Univ, Fac Sci & Engn, Dept Appl Elect Engn, Tokyo 101, Japan
[5] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Dept Thorac Organ Replacement, Tokyo, Japan
关键词
single-pivot bearing system; centrifugal blood pump; ventricular assist device; sealless continuous flow pump;
D O I
10.1046/j.1525-1594.2001.06856.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to develop a centrifugal blood pump that meets the requirements of a long-term, implantable circulatory support device, in this study a sing e-pivot bearing supported centrifugal blood pump was designed to evaluate its basic performance. The single-pivot structure consisted of a ceramic ball male pivot mounted on the bottom surface of the impeller and a polyethylene female pivot incorporated in the bottom pump casing. The follower magnet mounted inside the impeller was magnetically coupled to the driver magnet mounted on the shaft of the direct current brushless motor. As the motor rotated, the impeller rotated supported entirely by a single-pivot bearing system. The static pump performance obtained in the mock circulatory loop revealed an acceptable performance as a left ventricular assist device in terms of flow and head pressure. The pump flow of 5 L/min against the head pressure of 100 mm Hg was obtained at rotational speeds of 2,000 to 2,200 rpm. The maximum pump flow was 9 L/min with 2,200 rpm. The maximum electrical-to-hydraulic power conversion efficiency was around 14% at pump flows of 4 to 5 L/min. The stability of the impeller was demonstrated at the pump rpm higher than 1,400 with a single-pivot bearing without an additional support at its top. The single-pivot supported centrifugal pump can provide adequate flow and pressure as a ventricular assist device, but its mechanical stability and hemolytic as well as thrombotic performances must be tested prior to clinical use.
引用
收藏
页码:683 / 687
页数:5
相关论文
共 50 条
  • [31] A STRAIGHT PATH CENTRIFUGAL BLOOD PUMP CONCEPT IN THE CAPIOX CENTRIFUGAL PUMP
    KIJIMA, T
    OSHIYAMA, H
    HORIUCHI, K
    NOGAWA, A
    HAMASAKI, H
    AMANO, N
    NOJIRI, C
    FUKASAWA, H
    AKUTSU, T
    ARTIFICIAL ORGANS, 1993, 17 (07) : 593 - 598
  • [32] Design of a magnetic bearing-supported prototype centrifugal artificial heart Pump
    Allaire, PE
    Maslen, EH
    Kim, HC
    Bearnson, GB
    Olsen, DB
    TRIBOLOGY TRANSACTIONS, 1996, 39 (03): : 663 - 669
  • [33] Design of a magnetic bearing-supported prototype centrifugal artificial heart pump
    University of Virginia, Mechanical,, Aerospace, and Nuclear Engineering, Department, Charlottesville, Virginia 22901
    Tribol Trans, 3 (663-669):
  • [34] Centrifugal blood pump scaling
    Lorenz, M.
    Smith, W.
    ASAIO Journal, 2000, 46 (02)
  • [36] Real-time studies of the pivot bearings in the NEDO gyro PI-710 centrifugal blood pump
    Asai, T
    Watanabe, K
    Ito, S
    Tsujimura, S
    Motomura, T
    Shinohara, T
    Glueck, JA
    Nosé, Y
    ARTIFICIAL ORGANS, 2004, 28 (10) : 899 - 903
  • [37] Disposable magnetically levitated centrifugal blood pump: Design and in vitro performance
    Hoshi, H
    Asama, J
    Shinshi, T
    Ohuchi, K
    Nakamura, M
    Mizuno, T
    Arai, H
    Shimokohbe, A
    Takatani, S
    ARTIFICIAL ORGANS, 2005, 29 (07) : 520 - 526
  • [38] Hydrodynamic design and performance analysis of a centrifugal blood pump for cardiopulmonary circulation
    Oh, HW
    Yoon, ES
    Park, MR
    Sun, K
    Hwang, CM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART A-JOURNAL OF POWER AND ENERGY, 2005, 219 (A7) : 525 - 532
  • [39] Design of a self-bearing slice motor for a centrifugal blood pump
    Ueno, S
    Chen, C
    Ohishi, T
    Matsuda, K
    Okada, Y
    Taenaka, Y
    Masuzawa, T
    PROCEEDINGS OF THE SIXTH INTERNATIONAL SYMPOSIUM ON MAGNETIC BEARINGS, 1998, : 143 - 151
  • [40] A COMPACT CENTRIFUGAL BLOOD PUMP FOR EXTRACORPOREAL-CIRCULATION - DESIGN AND PERFORMANCE
    TANAKA, S
    YAMAMOTO, S
    YAMAKOSHI, K
    KAMIYA, A
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1987, 109 (03): : 272 - 278