Factors influencing taper failure of modular revision hip stems

被引:21
|
作者
Krull, A. [1 ]
Morlock, M. M. [1 ]
Bishop, N. E. [2 ]
机构
[1] TUHH Hamburg Univ Technol, Inst Biomech, Denickestr 15, D-21073 Hamburg, Germany
[2] HAW Hamburg Univ Appl Sci, Dept Biomed Engn, Fak Life Sci, Ulmenliet 20, D-21033 Hamburg, Germany
关键词
Hip revision; Taper junction; Fracture; Corrosion; Fatigue; Strength; Titanium; Finite element model; Surface damage; HEAD-NECK TAPER; FRETTING CORROSION; CONTACT FORCES; FEMORAL STEMS; ARTHROPLASTY; FRACTURE; REPLACEMENT; PROSTHESIS; JUNCTION; JOINT;
D O I
10.1016/j.medengphy.2018.02.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stem modularity of revision hip implant systems offers the advantage of the restoration of individual patient geometry but introduces additional interfaces, which are subjected to repetitive bending loading and have a propensity for fretting corrosion. The male stem taper is the weakest part of the modular junction due to its reduced cross section compared to the outside diameter of the stem. Taper fractures can be the consequence of overloading in combination with corrosion. The purpose of this study was to assess the influence of implant design factors, patient factors, and surgical factors on the risk of taper failure of the modular junction of revision stems. An analytical bending model was used to estimate the strength of the taper connection for pristine, fatigued and corroded conditions. Additionally, a finite element contact model of the taper connection was developed to assess the relative motion and potential for surface damage at the taper interface under physiological loading for varyied assembly and design parameters. Increasing the male taper diameter was shown to be the most effective means for increasing taper strength but would require a concurrent increase in the outer implant diameter to limit a greater risk of total surface damage for a thinner female taper wall. Increasing the assembly force decreases the total surface damage but not local magnitudes, which are probably responsible for crack initiation. It is suggested that in unfavourable loading conditions a monobloc implant system will reduce the risk of failure. (C) 2018 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 50 条
  • [1] Is taper corrosion in modular revision hip stem junctions associated with patient or implant specific factors? A retrieval analysis
    Bormann, Therese
    Kretzer, J. Philippe
    Jaeger, Sebastian
    Lohmann, Christoph H.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 150
  • [2] Failure mechanisms in CoCrMo modular femoral stems for revision total hip arthroplasty
    Wang, Qiong
    Parry, Michael
    Masri, Bassam A.
    Duncan, Clive
    Wang, Rizhi
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (06) : 1525 - 1535
  • [3] Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems Failure Analysis and Computed Tomography Diagnostics in Two Cases
    Norman, Pontus
    Iyengar, Srinivasan
    Svensson, Ingrid
    Flivik, Gunnar
    JOURNAL OF ARTHROPLASTY, 2014, 29 (04): : 850 - 855
  • [4] The different failure modes of the connecting elements of the modular hip arthroplasty revision stem Revitan
    Ransohoff, Christopher Butler
    Wanner, Ronald
    Solinger, Theo
    Gautier, Emanuel
    Eijer, Henk
    Wahl, Peter
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 123
  • [5] Failure analysis and mechanical testing of Ti-6Al-4V modular revision hip stems
    Nayar, P. L.
    Bohm, E. R.
    Petrak, M. J.
    Caley, W. F.
    Cahoon, J. R.
    CANADIAN METALLURGICAL QUARTERLY, 2016, 55 (04) : 420 - 428
  • [6] Nonmodular Stems Are a Viable Alternative to Modular Stems in Revision Total Hip Arthroplasty
    Clair, Andrew J.
    Cizmic, Zlatan
    Vigdorchik, Jonathan M.
    Poultsides, Lazaros A.
    Schwarzkopf, Ran
    Rathod, Parthiv A.
    Deshmukh, Ajit J.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (07): : S292 - S296
  • [7] Fixation pattern of conical and cylindrical modular revision hip stems in different size bone defects
    Kinkel, Stefan
    Nadorf, Jan
    Thomsen, Marc N.
    Heisel, Christian
    Jahnke, Alexander
    Kretzer, Jan P.
    Jakubowitz, Eike
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (09) : 1819 - 1825
  • [8] Failure analysis of three uncemented titanium-alloy modular total hip stems
    Paliwal, Manish
    Allan, D. Gordon
    Filip, Peter
    ENGINEERING FAILURE ANALYSIS, 2010, 17 (05) : 1230 - 1238
  • [9] Modular Taper Junction Corrosion and Failure: How to Approach a Recalled Total Hip Arthroplasty Implant
    Pivec, Robert
    Meneghini, R. Michael
    Hozack, William J.
    Westrich, Geoffrey H.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (01): : 1 - 6
  • [10] The survivorship of revision total hip replacement with severe proximal bone deficiency using a modular taper fluted prosthesis
    Govilkar, Siddharth
    Gandhi, Maulik J.
    Bhachu, Davinder S.
    Whittaker, John-Paul
    Evans, Christopher R.
    Spencer-Jones, Richard
    ACTA ORTHOPAEDICA BELGICA, 2022, 88 (02): : 303 - 309