Prosthetic radial head components and proximal radial morphology: A mismatch

被引:61
作者
Beredjiklian, PK [1 ]
Nalbantoglu, U [1 ]
Potter, HG [1 ]
Hotchkiss, RN [1 ]
机构
[1] Hosp Special Surg, Div Hand Surg & Radiol, New York, NY 10021 USA
关键词
D O I
10.1016/S1058-2746(99)90079-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A morphometric study of the proximal radius was performed with magnetic resonance imaging scans to measure the anatomic dimensions of the radial head and neck. These dimensions were then compared with the manufacturer's size specifications of commercially available titanium prosthetic radial head components to determine whether these designs adequately match the morphologic characteristics of the proximal radius. A cadaveric correlation was performed to validate the accuracy and reliability of measurements obtained from the magnetic resonance scans. The narrow intramedullary dimensions of the radial neck negated fitting of even the smallest available metallic prosthetic component stem in 18 (39%) of 46 scans reviewed (confidence interval 26% to 53%). Of the 31 male patients who underwent scanning, 4 (13%) would not be able to be fitted with a prosthetic component according to the manufacturer's technique guide (confidence interval 0% to 29%). Of the 15 female patients who underwent scanning, 14 (93%) would not able to be fitted with a prosthetic stem (confidence interval 70% to 99%). In those patients in whom the radial neck could accommodate a prosthetic stem (n = 26), there was ineffective restoration of proximal radial head length in all cases (100%, confidence interval 87% to 100%). The average shortening was 4 mm of proximal radial length (range 1 to 7 mm). Our findings reveal that the commercially available metallic radial head design may overestimate the dimensions of the radial neck, inadequate sizing of radial head prostheses may lead to an inadvertent change in proximal radial length, with potentially adverse effects on elbow, forearm, and wrist mechanics. Newer designs taking into account these anatomic dimensions may lead to an improvement in function after reconstruction.
引用
收藏
页码:471 / 475
页数:5
相关论文
共 13 条
[1]   FRACTURE OF A SILASTIC RADIAL-HEAD PROSTHESIS - DIAGNOSIS AND LOCALIZATION OF FRAGMENTS BY XEROGRAPHY - A CASE-REPORT [J].
BOHL, WR ;
BRIGHTMAN, E .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (09) :1482-1483
[2]   THE NEED FOR CONFIDENCE-INTERVALS IN THE PRESENTATION OF ORTHOPEDIC DATA [J].
DOREY, F ;
NASSER, S ;
AMSTUTZ, H .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (12) :1844-1852
[3]   Biomechanical and computer analysis of radial head prostheses [J].
Gupta, GG ;
Lucas, G ;
Hahn, DL .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 1997, 6 (01) :37-48
[4]  
Gupta GG., 1992, ORTHOP T, V16, P342
[5]   VALGUS STABILITY OF THE ELBOW [J].
HOTCHKISS, RN ;
WEILAND, AJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (03) :372-377
[6]   AN ANATOMIC AND MECHANICAL STUDY OF THE INTEROSSEOUS MEMBRANE OF THE FOREARM - PATHOMECHANICS OF PROXIMAL MIGRATION OF THE RADIUS [J].
HOTCHKISS, RN ;
AN, KN ;
SOWA, DT ;
BASTA, S ;
WEILAND, AJ .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1989, 14 (02) :256-261
[7]   Floating prosthesis for radial-head fractures [J].
Judet, T ;
deLoubresse, CG ;
Piriou, P ;
Charnley, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (02) :244-249
[8]   PRIMARY REPLACEMENT OF THE FRACTURED RADIAL HEAD WITH A METAL PROSTHESIS [J].
KNIGHT, DJ ;
RYMASZEWSKI, LA ;
AMIS, AA ;
MILLER, JH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (04) :572-576
[9]   FRACTURE OF SILASTIC RADIAL-HEAD PROSTHESIS - CASE-REPORT [J].
MAYHALL, WST ;
TILEY, FT ;
PALUSKA, DJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (03) :459-460
[10]  
MEADOW SL, 1992, J FORENSIC SCI, V31, P147