Post-traumatic zygomatic osteotomy and orbital floor reconstruction

被引:9
作者
Peel, Sean [1 ]
Eggbeer, Dominic [1 ]
Sugar, Adrian [2 ]
Evans, Peter Llewelyn [2 ]
机构
[1] Cardiff Metropolitan Univ, PDR Int Ctr Design & Res, Cardiff, S Glam, Wales
[2] Abertawe Bro Morgannwg Univ, Maxillofacial Unit, Morriston Hosp, Hlth Board, Swansea, W Glam, Wales
关键词
Medical; Computer-aided design; Additive manufacturing; Surgery; Implants; Surgical guides; FABRICATION; GUIDES;
D O I
10.1108/RPJ-03-2015-0037
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Purpose - Post-traumatic zygomatic osteotomy, fracture reduction, and orbital floor reconstruction pose many challenges for achieving a predictable, accurate, safe, and aesthetically pleasing result. This paper aims to describe the successful application of computer-aided design (CAD) and additive manufacturing (AM) to every stage of the process - from planning to surgery. Design/methodology/approach - A multi-disciplinary team was used - comprising surgeons, prosthetists, technicians, and designers. The patient's computed tomography scan data were segmented for bone and exported to a CAD software package. Medical models were fabricated using AM; for diagnosis, patient communication, and device verification. The surgical approach was modelled in the virtual environment and a custom surgical cutting guide, custom bone-repositioning guide, custom zygomatic implant, and custom orbital floor implant were each designed, prototyped, iterated, and validated using polymer AM prior to final fabrication using metal AM. Findings - Post-operative clinical outcomes were as planned. The patient's facial symmetry was improved, and their inability to fully close their eyelid was corrected. The length of the operation was reduced relative to the surgical team's previous experiences. Post-operative scan analysis indicated a maximum deviation from the planned location for the largest piece of mobilised bone of 3.65 mm. As a result, the orbital floor implant which was fixed to this bone demonstrated a maximum deviation of 4.44 mm from the plan. Originality/value - This represents the first application of CAD and AM to every stage of the process for this procedure - from diagnosis, to planning, and to surgery.
引用
收藏
页码:878 / 886
页数:9
相关论文
共 15 条
  • [11] Imperfect symmetry transform for orbital prosthesis modelling
    Sun, Jin
    Xiong, Yaoyang
    Chen, Xiaobo
    Xi, Juntong
    [J]. RAPID PROTOTYPING JOURNAL, 2013, 19 (03) : 180 - 188
  • [12] Template-based framework for nasal prosthesis fabrication
    Sun, Jin
    Chen, Xiaobo
    Liao, Haihong
    Xi, Juntong
    [J]. RAPID PROTOTYPING JOURNAL, 2013, 19 (02) : 68 - 76
  • [13] A CAD/CAM system for fabrication of facial prostheses
    Sun, Jin
    Xi, Juntong
    Chen, Xiaobo
    Xiong, Yaoyang
    [J]. RAPID PROTOTYPING JOURNAL, 2011, 17 (04) : 253 - 261
  • [14] Image-based planning and clinical validation of zygoma and pterygoid implant placement in patients with severe bone atrophy using customized drill guides. Preliminary results from a prospective clinical follow-up study
    Vrielinck, L
    Politis, C
    Schepers, S
    Pauwels, M
    Naert, U
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2003, 32 (01) : 7 - 14
  • [15] Medical rapid prototyping and 3D CT in the manufacture of custom made cranial titanium plates
    Winder, J
    Cooke, RS
    Gray, J
    Fannin, T
    Fegan, T
    [J]. JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1999, 23 (01) : 26 - 28