Evaluating the impact of extended field-of-view CT reconstructions on CT values and dosimetric accuracy for radiation therapy

被引:14
|
作者
Cheung, Joey P. [1 ,3 ]
Shugard, Erin [1 ,4 ]
Mistry, Nilesh [2 ]
Pouliot, Jean [1 ]
Chen, Josephine [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
[2] Siemens Healthcare USA, Malvern, PA 19355 USA
[3] Sutter Hlth Mills Peninsula Med Ctr, Dept Radiat Oncol, San Mateo, CA 94401 USA
[4] Kaiser Permanente Oakland, Dept Radiat Oncol, 3701 Broadway, Oakland, CA 94611 USA
关键词
CT simulation; extended field of view; radiation therapy; treatment planning; QUALITY-ASSURANCE; DOSE CALCULATIONS; RADIOTHERAPY; ALGORITHM; SCANNER; NUMBER;
D O I
10.1002/mp.13299
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWide bore CT scanners use extended field-of-view (eFOV) reconstruction algorithms to attempt to recreate tissue truncated due to large patient habitus. Radiation therapy planning systems rely on accurate CT numbers in order to correctly plan and calculate radiation dose. This study looks at the impact of eFOV reconstructions on CT numbers and radiation dose calculations in real patient geometries. MethodsA large modular phantom based on real patient geometries was created to surround a CIRS Model 062M phantom. The modular sections included a smooth patient surface, a skin fold in the patient surface, and the addition of arms for simulation of the patient in arms up or arms down position. This phantom was used to evaluate the accuracy of CT numbers for three extended FOV algorithms implemented on Siemens CT scanners: eFOV, HDFOV, and HDProFOV. Six different configurations of the phantoms were scanned and images were reconstructed for the three different extended FOV algorithms. The CIRS phantom inserts and overall phantom geometry were contoured in each image, and the Hounsfield units (HU) numbers were compared to an image of the phantom within the standard scan FOV (sFOV) without the modular sections. To evaluate the effect on dose calculations, six radiotherapy patients previously treated at our institution (three head and neck and three chest/pelvis) whose body circumferences extended past the 50cm sFOV in the treatment planning CT were used. Images acquired on a Siemens Sensation Open scanner were reconstructed using sFOV, eFOV and HDFOV algorithms. A physician and dosimetrist identified the radiation target, critical organs, and external patient contour. A benchmark CT was created for each patient, consisting of an average of the 3 CT reconstructions with a density override applied to regions containing truncation artifacts. The benchmark CT was used to create an optimal radiation treatment plan. The plan was copied onto each CT reconstruction without density override and dose was recalculated. ResultsTissue extending past the sFOV impacts the HU numbers for tissues inside and outside the sFOV when using extended FOV reconstructions. On average, the HU for all CIRS density inserts in the arms up (arms down) position varied by 43HU (67HU), 39HU (73HU), and 18HU (51HU) for the eFOV, HDFOV, and HDProFOV scans, respectively. In the patient dose calculations, patients with a smooth patient contour had the least deviation from the benchmark in the HDFOV (0.1-0.5%) compared to eFOV (0.4-1.8%) reconstructions. In cases with large amounts of tissue and irregular skin folds, the eFOV deviated the least from the benchmark (range 0.2-0.6% dose difference) compared to HDFOV (range 1.3-1.8% dose difference). ConclusionsAll reconstruction algorithms demonstrated good CT number accuracy in the center of the image. Larger artifacts are seen near and extending outside the scan FOV, however, dose calculations performed using typical beam arrangements using the extended FOV reconstructions were still mostly within 2.5% of best estimated reference values.
引用
收藏
页码:892 / 901
页数:10
相关论文
共 50 条
  • [41] -Paediatric Dose Optimization for Long Axial Field-of-View FDG PET/CT
    van der Kaap, D.
    Deuten, M.
    van Snick, P. J. H.
    Noordzij, W.
    Slart, R. H. J. A.
    Brouwers, A. H.
    Glaudemans, A. W. J. M.
    Koopmans, K. P.
    Ivashchenko, V.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S310 - S311
  • [42] Long Axial Field-of-View PET/CT: New Opportunities for Pediatric Imaging
    Dias, Andre Henrique
    Andersen, Kim Francis
    Fosbol, Marie obro
    Gormsen, Lars Christian
    Andersen, Flemming Littrup
    Munk, Ole Lajord
    SEMINARS IN NUCLEAR MEDICINE, 2025, 55 (01) : 76 - 85
  • [43] Limited Field-Of-View Cone-Beam CT Reconstruction for Adaptive Radiotherapy
    Lu, W.
    Yan, H.
    Zhou, L.
    Cervino, L.
    Jiang, S.
    Jia, X.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [44] Impact of radiation dose level on CT number accuracy in photon counting CT
    Zhang, Ran
    Cruz-Bastida, Juan P.
    Gomez-Cardona, Daniel
    Hayes, John
    Li, Ke
    Chen, Guang-Hong
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [45] Brain-Only Imaging on Long Axial Field-of-View PET/CT
    Khan, J.
    Willson, T.
    Davis, N.
    Holman, B. F.
    Ferreira, B.
    McCool, D.
    Edwards, J.
    Szyszko, T.
    Wagner, T.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S702 - S703
  • [46] Impact of view reduction in CT on radiation dose for patients
    Parcero, E.
    Flores, L.
    Sanchez, M. G.
    Vidal, V.
    Verdu, G.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 137 : 173 - 175
  • [47] Quantifying the Dosimetric Impact of a Novel Synthetic CT Algorithm for MR Simulation of Prostate Cancer Radiation Therapy
    Garbarino, K. S.
    Kim, J. P.
    Wen, N.
    Schultz, L.
    Levin, K. J.
    Movsas, B.
    Siddiqui, M. S. U.
    Chetty, I. J.
    Glide-Hurst, C. K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S157 - S158
  • [48] Ultra-low foetal radiation exposure in 18F-FDG PET/CT imaging with a long axial field-of-view PET/CT system
    Smith, Charlotte L. C.
    Yaqub, Maqsood
    Wellenberg, Ruud H. H.
    Knip, Jelijn J.
    Boellaard, Ronald
    Zwezerijnen, Gerben J. C.
    EJNMMI PHYSICS, 2024, 11 (01):
  • [49] Positronium lifetime measurements in patients with a commercial long axial field-of-view PET/CT
    Cardoso, A.
    Mercolli, L.
    Steinberger, W.
    Afshar-Oromieh, A.
    Caobelli, F.
    Conti, M.
    Mingels, C.
    Pyka, T.
    Rominger, A.
    Sari, H.
    Seifert, R.
    Shi, K.
    Viscione, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S216 - S216
  • [50] Long-axial field-of-view PET/CT improves radiomics feature reliability
    Alberts, Ian L.
    Xue, Song
    Sari, Hasan
    Cavinato, Lara
    Prenosil, George
    Afshar-Oromieh, Ali
    Mingels, Clemens
    Shi, Kuangyu
    Caobelli, Federico
    Rahmim, Arman
    Pyka, Thomas
    Rominger, Axel
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2025, 52 (03) : 1004 - 1016