Three year experience of electronic portal imaging device based daily QA for photon radiation beams

被引:1
|
作者
Cai, Bin [1 ]
He, Yu [1 ]
Bollinger, Douglas [1 ]
Li, Hua [1 ]
Goddu, Sreekrishna M. [1 ]
Mutic, Sasa [1 ]
Sun, Baozhou [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
来源
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS | 2019年 / 5卷 / 01期
关键词
EPID; daily QA; photon radiotherapy; QUALITY-ASSURANCE;
D O I
10.1088/2057-1976/aae419
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Anin-house developed EPID based QA strategy was deployed clinically to test daily linac performance before treatment. Beam dosimetric parameters, including output, flatness, symmetry, uniformity, TPR20/10, and MLC repeatability, as well as image quality metrics, such as high and low contrast resolutions, were recorded based on MV and kV EPID imaging analysis. Over a period of three years, data was collected and analyzed. Trend analysis with data visualization, comparisons to ionization chamber (IC) results and QA failure studies were performed to evaluate the reliability and robustness of the EPID based daily QA system. The EPID output tests over the three-year period cross-correlated with the monthly QA IC system within 1.3% and TG51 results within 1%. A consistent positive output drift was detected for all energies during that time. A faster drifting speed was observed after the installation of a new beam monitoring chamber on the machine but slowed down over time. Symmetry, flatness, and uniformity were stable and the changes were directly related to beam steering or QA baseline resetting. TPR20/10 was not sensitive to dosimetry fluctuations and a good delivery repeatability of MLC fields was observed. The image quality metrics were stable and a passing rate greater than 96% was observed. For the most recent eight month of data, a one-time QA passing rate was calculated to be 87.6%. Beam interruptions, human errors and data transfer issues were the major reasons for QA failures. The data gathered over 3 years indicates that the EPID-based process is an efficient and reliable QA solution for daily linac performance checks. The long-term trend analysis and data visualization can help physicists understand the behavior of a linac, optimize the daily QA procedure, and has the potential to predict machine failures prior to incidents.
引用
收藏
页数:9
相关论文
共 36 条
  • [31] Linear array measurements of enhanced dynamic wedge and treatment planning system (TPS) calculation for 15 MV photon beam and comparison with electronic portal imaging device (EPID) measurements
    Petrovic, Borislava
    Grzadziel, Aleksandra
    Rutonjski, Laza
    Slosarek, Krzysztof
    RADIOLOGY AND ONCOLOGY, 2010, 44 (03) : 199 - 206
  • [32] Evaluation of set-up errors and estimation of set-up margin during external beam radiation therapy of prostate cancer using electronic portal imaging device (EPID)
    Khoramian, Daryoush
    Sistani, Soroush
    Farhood, Bagher
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 493 - 500
  • [33] Calculation of exit dose for conformal and dynamically-wedged fields, based on water-equivalent path length measured with an amorphous silicon electronic portal imaging device
    Kavuma, Awusi
    Glegg, Martin
    Metwaly, Mohamed
    Currie, Garry
    Elliott, Alex
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (03): : 44 - 60
  • [34] Deep learning-based 3D in vivo dose reconstruction with an electronic portal imaging device for magnetic resonance-linear accelerators: a proof of concept study
    Li, Yongbao
    Xiao, Fan
    Liu, Biaoshui
    Qi, Mengke
    Lu, Xingyu
    Cai, Jiajun
    Zhou, Linghong
    Song, Ting
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (23):
  • [35] Commissioning and dosimetric verification of volumetric modulated arc therapy for multiple modalities using electronic portal imaging device-based 3D dosimetry system: a novel approach
    Hajare, Raghavendra
    Sreelakshmi, K. K.
    Kumar, Anil
    Kalita, Rituraj
    Kaginelli, Shanmukhappa
    Mahantshetty, Umesh
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (02) : 412 - 424
  • [36] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: II. Evaluation of a dosimetric uncertainty budget and development of a new film-in-EPID absorbed dose calibration methodology
    Renaud, James
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (02) : 1238 - 1247