Three year experience of electronic portal imaging device based daily QA for photon radiation beams

被引:1
|
作者
Cai, Bin [1 ]
He, Yu [1 ]
Bollinger, Douglas [1 ]
Li, Hua [1 ]
Goddu, Sreekrishna M. [1 ]
Mutic, Sasa [1 ]
Sun, Baozhou [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
来源
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS | 2019年 / 5卷 / 01期
关键词
EPID; daily QA; photon radiotherapy; QUALITY-ASSURANCE;
D O I
10.1088/2057-1976/aae419
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Anin-house developed EPID based QA strategy was deployed clinically to test daily linac performance before treatment. Beam dosimetric parameters, including output, flatness, symmetry, uniformity, TPR20/10, and MLC repeatability, as well as image quality metrics, such as high and low contrast resolutions, were recorded based on MV and kV EPID imaging analysis. Over a period of three years, data was collected and analyzed. Trend analysis with data visualization, comparisons to ionization chamber (IC) results and QA failure studies were performed to evaluate the reliability and robustness of the EPID based daily QA system. The EPID output tests over the three-year period cross-correlated with the monthly QA IC system within 1.3% and TG51 results within 1%. A consistent positive output drift was detected for all energies during that time. A faster drifting speed was observed after the installation of a new beam monitoring chamber on the machine but slowed down over time. Symmetry, flatness, and uniformity were stable and the changes were directly related to beam steering or QA baseline resetting. TPR20/10 was not sensitive to dosimetry fluctuations and a good delivery repeatability of MLC fields was observed. The image quality metrics were stable and a passing rate greater than 96% was observed. For the most recent eight month of data, a one-time QA passing rate was calculated to be 87.6%. Beam interruptions, human errors and data transfer issues were the major reasons for QA failures. The data gathered over 3 years indicates that the EPID-based process is an efficient and reliable QA solution for daily linac performance checks. The long-term trend analysis and data visualization can help physicists understand the behavior of a linac, optimize the daily QA procedure, and has the potential to predict machine failures prior to incidents.
引用
收藏
页数:9
相关论文
共 36 条
  • [21] The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device
    Tuncel, Nina
    Mahdavi, Seied Rabi
    Haghparast, Mohammad
    Karakus, Ismail
    Haghparast, Abbas
    Nazari, Vahideh
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 : S815 - S820
  • [22] Sensitivity of Electronic Portal Imaging Device (EPID) Based Transit Dosimetry to Detect Inter-fraction Patient Variations
    Bawazeer, Omemh
    Sarasanandarajah, Sivananthan
    Herath, Sisira
    Kron, Tomas
    Deb, Pradip
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 477 - 480
  • [23] Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine
    Jinhan Zhu
    Lixin Chen
    Along Chen
    Guangwen Luo
    Xiaowu Deng
    Xiaowei Liu
    Radiation Oncology, 10
  • [24] Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation
    Hirashima, Hideaki
    Miyabe, Yuki
    Nakamura, Mitsuhiro
    Mukumoto, Nobutaka
    Mizowaki, Takashi
    Hiraoka, Masahiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (03): : 234 - 242
  • [25] Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine
    Zhu, Jinhan
    Chen, Lixin
    Chen, Along
    Luo, Guangwen
    Deng, Xiaowu
    Liu, Xiaowei
    RADIATION ONCOLOGY, 2015, 10
  • [26] Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device
    Zeidan, OA
    Li, JG
    Ranade, M
    Stell, AM
    Dempsey, JF
    MEDICAL PHYSICS, 2004, 31 (03) : 463 - 476
  • [27] Dose response characteristics of a novel CCD camera-based electronic portal imaging device comparison with OCTAVIUS detector
    Anvari, Akbar
    Aghamiri, Seyed Mahmoud Reza
    Mahdavi, Seyed Rabie
    Alaei, Parham
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2015, 11 (04) : 765 - 769
  • [28] Intensity modulated radiation therapy (IMRT) verification using the frame averaging feature of Varian Mark II electronic portal imaging device (EPID)
    Chang, J
    Mageras, G
    Ling, CC
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2059 - 2062
  • [29] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: I. Quantities influencing long-term stability
    Renaud, James
    Muir, Bryan
    MEDICAL PHYSICS, 2022, 49 (02) : 1231 - 1237
  • [30] DAILY MONITORING AND CORRECTION OF RADIATION-FIELD PLACEMENT USING A VIDEO-BASED PORTAL IMAGING-SYSTEM - A PILOT-STUDY
    EZZ, A
    MUNRO, P
    PORTER, AT
    BATTISTA, J
    JAFFRAY, DA
    FENSTER, A
    OSBORNE, S
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (01): : 159 - 165