Survivorship Between 2 Different Ceramic-on-Ceramic Total Hip Arthroplasty With or Without a Metal-Backed Titanium Sleeve Bearing: A 5-to 14-Year Follow-Up Study

被引:9
作者
Higuchi, Yoshitoshi [1 ]
Hasegawa, Yukiharu [2 ]
Komatsu, Daigo [1 ]
Seki, Taisuke [1 ]
Ishiguro, Naoki [1 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Orthopaed Surg, Nagoya, Aichi, Japan
[2] Nagoya Univ, Grad Sch Med, Dept Hip & Knee Reconstruct Surg, Nagoya, Aichi, Japan
基金
日本学术振兴会;
关键词
ceramic-on-ceramic bearings; total hip arthroplasty; wear; long-term follow-up; metal-backed titanium sleeve; CROSS-LINKED POLYETHYLENE; REPLACEMENT ARTHROPLASTIES; LINER; WEAR; THA; SQUEAKING; SURFACES; FRACTURE; TRIAL;
D O I
10.1016/j.arth.2016.06.014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aim of the study was to compare the clinical and radiographic results of consecutive ceramic-on-ceramic bearings with and without a metal-backed titanium sleeve in patients undergoing total hip arthroplasty. Methods: Eighty-five patients (64 women and 21 men; average age 55.2 years) were included in the A group without sleeve while 147 patients (116 women and 31 men; average age 54.2 years) were included in the B group with sleeve. Clinical and radiologic measurements at follow-up (range, 5-14 years; average, 8.1 years) were analyzed. Results: The mean latest postoperative Harris Hip Score was 89.1 for patients from both groups. One joint (1.2%) in the A group displayed ceramic liner fracture, while no incidences of liner fracture occurred in the B group. Audible squeaking was observed in 1 joint (1.2%) in the A group and 1 (0.7%) in the B group. The mean annual liner rate of wear was 0.0049 and 0.0046 mm/y for the A group and B group, respectively. Three joints in the A group (3.5%) required revision total hip arthroplasty because of individual episodes of aseptic cup loosening, ceramic liner fracture, and infection. One joint in the B group (0.7%) required revision because of progressive osteolysis of the proximal femur. Ten-year Kaplan-Meier survivorship, based on an end point of component loosening and bearing failure, was 97.6% for the A group and 99.3% for the B group. Conclusion: There were no clinical, radiographic, or survivorship differences between groups. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:155 / 160
页数:6
相关论文
共 38 条
[1]   Ceramic Liner Fractures Presenting as Squeaking After Primary Total Hip Arthroplasty [J].
Abdel, Matthew P. ;
Heyse, Thomas J. ;
Elpers, Marcella E. ;
Mayman, David J. ;
Su, Edwin P. ;
Pellicci, Paul M. ;
Wright, Timothy M. ;
Padgett, Douglas E. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (01) :27-31
[2]   Wear of 36-mm BIOLOX® delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions [J].
Al-Hajjar, Mazen ;
Fisher, John ;
Tipper, Joanne L. ;
Williams, Sophie ;
Jennings, Louise M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (H5) :535-542
[3]  
Archibeck MJ, 2000, CLIN ORTHOP RELAT R, P12
[4]   Radiographic wear measurements in a cementless metal-backed modular cobalt-chromium acetabular component [J].
Barrack, RL ;
Lavernia, C ;
Szuszczewicz, ES ;
Sawhney, J .
JOURNAL OF ARTHROPLASTY, 2001, 16 (07) :820-828
[5]   Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study [J].
Bascarevic, Zoran ;
Vukasinovic, Zoran ;
Slavkovic, Nemanja ;
Dulic, Borislav ;
Trajkovic, Goran ;
Bascarevic, Violeta ;
Timotijevic, Sladjan .
INTERNATIONAL ORTHOPAEDICS, 2010, 34 (08) :1129-1135
[6]   THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[7]   ECTOPIC OSSIFICATION FOLLOWING TOTAL HIP-REPLACEMENT - INCIDENCE AND A METHOD OF CLASSIFICATION [J].
BROOKER, AF ;
BOWERMAN, JW ;
ROBINSON, RA ;
RILEY, LH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08) :1629-1632
[8]   Wear of highly crosslinked polyethylene acetabular components [J].
Callary, Stuart A. ;
Solomon, Lucian B. ;
Holubowycz, Oksana T. ;
Campbell, David G. ;
Munn, Zachary ;
Howie, Donald W. .
ACTA ORTHOPAEDICA, 2015, 86 (02) :159-168
[9]   Ceramic-on-Ceramic Total Hip Arthroplasty: Update [J].
Capello, William N. ;
D'Antonio, James A. ;
Feinberg, Judy R. ;
Manley, Michael T. ;
Naughton, Marybetb .
JOURNAL OF ARTHROPLASTY, 2008, 23 (07) :39-43
[10]   Fracture of a Titanium Sleeve-encased Third-generation Ceramic Liner in a Modern THA [J].
Chotai, Pranit N. ;
Su, Edwin P. .
ORTHOPEDICS, 2011, 34 (10) :E682-E684