Multicollector ICP-MS analysis of Pb isotopes using a 207pb-204pb double spike demonstrates up to 400 ppm/amu systematic errors in Tl-normalization

被引:283
作者
Thirlwall, MF [1 ]
机构
[1] Univ London, Royal Holloway & Bedford New Coll, Dept Geol, Egham TW20 0EX, Surrey, England
关键词
lead; mass bias; ICP-MS; isotopes; errors; TIMS;
D O I
10.1016/S0009-2541(01)00365-5
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
Pb isotope analyses using multicollector magnetic sector inductively coupled plasma mass spectrometers (ICP-MS) may potentially be corrected for mass bias, a major source of uncertainty, by assuming that the mass bias for TI is similar to that of Pb. Around 7 ng of Pb has been analysed in this study using a Micromass IsoProbe MC-ICP-MS. With detailed attention to backgrounds arising from instrument memory and tailing from adjacent peaks, external precision of 55 -180 ppm 2sd has been achieved over a year on Pb isotope ratios normalized to constant Pb-208/Pb-206. This involves use of a wide range of SRM981 solutions and two different nebulizers. Mean ratios are very close to multidynamic TIMS data normalized in the same way, provided that anomalous TIMS behaviour of (207)pb [Chem. Geol. Isot. Geosci. Sect. 163 (2000) 2991 is corrected to the least fractionated ratios measured. Tl-normalized Pb isotope ratios vary substantially between different SRM981 solutions and between nebulizers, with external precision over the year being up to 541 ppm (2sd) on Pb-208/Pb-204. However, a (207)pb-(204)pb double spike (DS) can accurately correct for mass bias on the IsoProbe, and external precision of at worst 245 ppm (2sd) on Pb-208/Pb-204 has been obtained over the year using a wide range of SRM981 solutions and two different nebulizers. Mean ratios on DS-corrected SRM981 and silicate samples are within error of DS-TIMS results, even using an independent calibration for the DS on the IsoProbe. A new Pb-208/Pb-206 value for SRM981 of 2.1677+/-2 is proposed, normalized to SRM982 Pb-208/Pb-206=1.00016. Use of SRM981 to optimize the Tl-205/Tl-203 normalizing ratio would give Tl-normalized sample ratios similar to400 ppm/amu higher than those calculated using the DS correction. It is shown that these problems with TI normalization cannot result from isobaric interferences, but they appear to be due to unusual solution chemistry of TI in the nebulizer. (C) 2002 Published by Elsevier Science B.V.
引用
收藏
页码:255 / 279
页数:25
相关论文
共 21 条
[11]   Investigation of matrix effects for Pb isotope ratio measurements by multiple collector ICP-MS:: verification and application of optimized analytical protocols [J].
Rehkämper, M ;
Mezger, K .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (11) :1451-1460
[12]   U-Pb dating of a speleothem of Quaternary age [J].
Richards, DA ;
Bottrell, SH ;
Cliff, RA ;
Ströhle, K ;
Rowe, PJ .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1998, 62 (23-24) :3683-3688
[13]   CA ISOTOPE FRACTIONATION ON EARTH AND OTHER SOLAR-SYSTEM MATERIALS [J].
RUSSELL, WA ;
PAPANASTASSIOU, DA ;
TOMBRELLO, TA .
GEOCHIMICA ET COSMOCHIMICA ACTA, 1978, 42 (08) :1075-1090
[14]   Inappropriate tail corrections can cause large inaccuracy in isotope ratio determination by MC-ICP-MS [J].
Thirlwall, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (10) :1121-1125
[15]   Inter-laboratory and other errors in Ph isotope analyses investigated using a 207Pb-204Pb double spike [J].
Thirlwall, MF .
CHEMICAL GEOLOGY, 2000, 163 (1-4) :299-322
[16]  
THIRLWALL MF, 1998, MINERAL MAG A, V62, P1507
[17]  
Todt W., 1996, GEOPHYS MONOGR SER, DOI [DOI 10.1029/GM095, 10.1029/GM095p0429, DOI 10.1029/GM095P0429]
[18]  
VANCE D, UNPUB GEOCHIM COSMOC
[19]  
VANCE D, 1999, LPI CONTRIBUTION, V971, P307
[20]   High-precision analysis of Pb isotope ratios by multi-collector ICP-MS [J].
White, WM ;
Albarède, F ;
Télouk, P .
CHEMICAL GEOLOGY, 2000, 167 (3-4) :257-270