Determination of serum creatinine by Jaffe method and how to calibrate to eliminate matrix interference problems

被引:63
作者
Chromy, Vratislav [1 ]
Rozkosna, Katerina [2 ]
Sedlak, Pavel [3 ]
机构
[1] Masaryk Univ, Fac Sci, Inst Chem, Brno, Czech Republic
[2] Pliva Lachema Diagnostika, Diagnost Res Dept, Brno, Czech Republic
[3] Municipal Hosp, Dept Clin Biochem, Caslav, Czech Republic
关键词
artificial serum matrix for calibration of Jaffe method; compensation of serum interferents in Jaffe method; Jaffe serum creatinine;
D O I
10.1515/CCLM.2008.224
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The calibration of Jaffe method for serum creatinine using one serum-based standard complemented with artificial matrix compensating standard's Jaffe-interfering substances allows two-point calibration with results well comparable with enzymatic methods. Method: Spectrophotometry was used. Results: Jaffe procedures compensating serum/plasma intereferents by subtracting a constant amount of creatinine poorly overcompensate creatinine in children. Two-point calibration with a pair of primary serum standards certified by the reference measurement procedure ( isotope-dilution, mass spectrometry) or with a pair of secondary standards linked to primary materials could provide results well agreeable with enzymatic determination. Such a calibration comprises an absorbance offset corresponding to other-than-creatinine Jaffe-interfering chromogens present in standards in the calibration line, while a two-point calibration combining one standard with physiological saline/water always grossly distorts calibration line. We calculated/prepared artificial serum matrices capable of compensating Jaffe-interfering chromogens in serum standards. The combination of even one standard with its artificial matrix also enables two-point calibration with practically the same results as with a pair of primary standards. Conclusions: A two-point calibration of Jaffe method for serum creatinine combining only one serum standard with creatinine solution matching standard's allow matrix-present interferents present results well comparable with enzymatic determination, providing the standard is attested/linked to the reference measurement procedure.
引用
收藏
页码:1127 / 1133
页数:7
相关论文
共 15 条
[1]  
Carobene A, 1997, CLIN CHEM, V43, P1342
[2]  
CHROMY V, 2007, KLIN BIOCH METAB S, V15, pS7
[3]  
*CREA, CREAT JAFF METH RAT
[4]  
Edmund L, 2006, TIETZ TXB CLIN CHEM, P797
[5]  
FISCHER J, 1982, Patent No. 1982228836
[6]   Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffe method [J].
Junge, W ;
Wilke, B ;
Halabi, A ;
Klein, G .
CLINICA CHIMICA ACTA, 2004, 344 (1-2) :137-148
[7]   DEVELOPMENT OF AN AUTOMATED KINETIC JAFFE METHOD DESIGNED TO MINIMIZE BILIRUBIN INTERFERENCE IN PLASMA CREATININE ASSAYS [J].
KNAPP, ML ;
MAYNE, PD .
CLINICA CHIMICA ACTA, 1987, 168 (02) :239-246
[8]   Creatinine assays: time for action? [J].
Lawson, N ;
Lang, T ;
Broughton, A ;
Prinsloo, P ;
Turner, C ;
Marenah, C .
ANNALS OF CLINICAL BIOCHEMISTRY, 2002, 39 :599-602
[9]   Recommendations for improving serum creatinine measurement: A report from the laboratory working group of the National Kidney Disease Education Program [J].
Myers, GL ;
Miller, WG ;
Coresh, J ;
Fleming, J ;
Greenberg, N ;
Greene, T ;
Hostetter, T ;
Levey, AS ;
Panteghini, M ;
Welch, M ;
Eckfeldt, JH .
CLINICAL CHEMISTRY, 2006, 52 (01) :5-18
[10]  
*NAT U STAND TECHN, 2007, STAND REF MAT 967