Investigation of the Variability in the Assessment of Digital Chest X-ray Image Quality

被引:32
作者
Whaley, Jacquelyn S. [1 ]
Pressman, Barry D. [2 ]
Wilson, Jonathan R. [2 ]
Bravo, Lionel [2 ]
Sehnert, William J. [1 ]
Foos, David H. [1 ]
机构
[1] Carestream Hlth Inc, Clin Applicat Res, Rochester, NY 14615 USA
[2] Cedars Sinai Med Ctr, S Mark Taper Fdn Imaging Ctr, Dept Imaging, Los Angeles, CA 90048 USA
关键词
Image quality; quality assurance; quality control; digital radiography; computed radiography; algorithms; reject analysis; repeat analysis; image defect detection; REJECT ANALYSIS; RADIOGRAPHY; CHALLENGES; AGREEMENT;
D O I
10.1007/s10278-012-9515-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A large database of digital chest radiographs was developed over a 14-month period. Ten radiographic technologists and five radiologists independently evaluated a stratified subset of images from the database for quality deficiencies and decided whether each image should be rejected. The evaluation results showed that the radiographic technologists and radiologists agreed only moderately in their assessments. When compared against each other, radiologist and technologist reader groups were found to have even less agreement than the inter-reader agreement within each group. Radiologists were found to be more accepting of limited-quality studies than technologists. Evidence from the study suggests that the technologists weighted their reject decisions more heavily on objective technical attributes, while the radiologists weighted their decisions more heavily on diagnostic interpretability relative to the image indication. A suite of reject-detection algorithms was independently run on the images in the database. The algorithms detected 4 % of postero-anterior chest exams that were accepted by the technologist who originally captured the image but which would have been rejected by the technologist peer group. When algorithm results were made available to the technologists during the study, there was no improvement in inter-reader agreement in deciding whether to reject an image. The algorithm results do, however, provide new quality information that could be captured within a site-wide, reject-tracking database and leveraged as part of a site-wide QA program.
引用
收藏
页码:217 / 226
页数:10
相关论文
共 27 条
[1]  
Castelletti A., 2010, 201055 DIP EL INF PO, DOI DOI 10.1109/ITAB.2010.5687726
[2]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[3]  
Dayhoff JE, 2001, CANCER, V91, P1615, DOI 10.1002/1097-0142(20010415)91:8+<1615::AID-CNCR1175>3.0.CO
[4]  
2-L
[5]   Digital Radiography Reject Analysis: Data Collection Methodology, Results, and Recommendations from an In-depth Investigation at Two Hospitals [J].
Foos, David H. ;
Sehnert, W. James ;
Reiner, Bruce ;
Siegel, Eliot L. ;
Segal, Arthur ;
Waldman, David L. .
JOURNAL OF DIGITAL IMAGING, 2009, 22 (01) :89-98
[6]  
Gallet J., 2010, Carestream Technical Brief Series. CAT No. 120 7091
[7]  
Hamel L.H., 2011, Knowledge Discovery with Support Vector Machines
[8]   The Foundation of the Concept of Relevance [J].
Hjorland, Birger .
JOURNAL OF THE AMERICAN SOCIETY FOR INFORMATION SCIENCE AND TECHNOLOGY, 2010, 61 (02) :217-237
[9]  
IMV Medical Information Division, 2010, IMV MED INF DIV 2010
[10]   One Year's Results from a Server-Based System for Performing Reject Analysis and Exposure Analysis in Computed Radiography [J].
Jones, A. Kyle ;
Polman, Raimund ;
Willis, Charles E. ;
Shepard, S. Jeff .
JOURNAL OF DIGITAL IMAGING, 2011, 24 (02) :243-255