Microfabricated needle-type sensors for PO2, PCO2, and pH

被引:12
作者
Wang, XW [1 ]
Suzuki, H
Hayashi, K
Kaneko, T
Sunagawa, K
机构
[1] Univ Tsukuba, Inst Mat Sci, Tsukuba, Ibaraki 3058573, Japan
[2] NTT Microsyst Integrat Labs, Ubiquitous Interface Lab, Biosensing Res Grp, Kanagawa 2430198, Japan
[3] Terumo Co Ltd, Prod Technol Ctr, Shizuoka 4180004, Japan
[4] Kyushu Univ, Grad Sch Med Sci, Dept Cardiovasc Med, Fukuoka 8128582, Japan
关键词
blood gas; carbon dioxide; oxygen; pH; reference electrode; sterilization;
D O I
10.1109/JSEN.2005.853594
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In order to conduct in vivo continuous monitoring of blood gas levels, very thin disposable pO(2), pCO(2), and pH sensors, which could be incorporated into a syringe needle or a catheter, were microfabricated. Electrode patterns were formed on a polyimide substrate and stacked with intervening polyimide insulating layers. An electrolyte layer was formed, which was covered with a silicone rubber gas-permeable membrane or a passivating layer. The most critical part of the sensor was a Ag-AgCl reference electrode with a pinhole structure. Because of the structure and novel operational mode of the reference electrode, its potential stabilized immediately, and reproducible responses were realized for the respective sensors. Performance characterization revealed clear responses and linear calibration plots in the ranges of partial pressures or pH of physiological levels. No significant damages or changes in the performance of the sensors were observed after ethylene oxide sterilization.
引用
收藏
页码:11 / 18
页数:8
相关论文
共 24 条
  • [1] ARQUINT P, 1994, CLIN CHEM, V40, P1805
  • [2] INTEGRATED BLOOD-GAS SENSOR FOR PO2, PCO2 AND PH
    ARQUINT, P
    VANDENBERG, A
    VANDERSCHOOT, BH
    DEROOIJ, NF
    BUHLER, H
    MORF, WE
    DURSELEN, LFJ
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1993, 13 (1-3) : 340 - 344
  • [3] CHEMICAL SENSORS FOR BEDSIDE MONITORING OF CRITICALLY ILL PATIENTS
    COLLISON, ME
    MEYERHOFF, ME
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (07) : A425 - &
  • [4] REFERENCE ELECTRODE WITH FREE-FLOWING FREE-DIFFUSION LIQUID JUNCTION
    DOHNER, RE
    WEGMANN, D
    MORF, WE
    SIMON, W
    [J]. ANALYTICAL CHEMISTRY, 1986, 58 (12) : 2585 - 2589
  • [5] ERICKSON KA, 1993, CLIN CHEM, V39, P283
  • [6] ULTRAMICRO OXYGEN ELECTRODE
    FATT, I
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1964, 19 (02) : 326 - &
  • [7] Simultaneous monitoring of pH, CO2 and O-2 using an optical imaging fiber
    Ferguson, JA
    Healey, BG
    Bronk, KS
    Barnard, SM
    Walt, DR
    [J]. ANALYTICA CHIMICA ACTA, 1997, 340 (1-3) : 123 - 131
  • [8] GUMBRECHT W, 1994, SENSOR ACTUAT B-CHEM, V18, P704
  • [9] LATNER AL, 1975, CLIN BIOCH, pCH12
  • [10] Catheter-Type sensor for potentiometric monitoring of oxygen, pH and carbon dioxide
    Meruva, RK
    Meyerhoff, ME
    [J]. BIOSENSORS & BIOELECTRONICS, 1998, 13 (02) : 201 - 212