共 12 条
Eigen-Vector Based Leads for Reconstruction of the 12-Lead Electrocardiogram
被引:0
作者:
Finlay, D. D.
[1
]
Nugent, C. D.
[1
]
Donnelly, M. P.
[1
]
Nelwan, S. P.
[2
]
机构:
[1] Univ Ulster, Shore Rd, Belfast BT37 0QB, Antrim, North Ireland
[2] Erasmus MC, Rotterdam, Netherlands
来源:
COMPUTERS IN CARDIOLOGY 2008, VOLS 1 AND 2
|
2008年
关键词:
D O I:
10.1109/CIC.2008.4749053
中图分类号:
TP [自动化技术、计算机技术];
学科分类号:
0812 ;
摘要:
In this study we have developed a limited lead system to allow reconstruction of the 12-lead ECG. This has been based upon the analysis of eigen-vectors calculated from body surface potential map (BSPM) data. Eigen-vectors were calculated from a set of 117 lead BSPMs (normal = 172, LVH = 178, MI = 209). The extrema of the first three eigen-vectors were used to determine the positions of recording sites for three bipolar leads. These three bipolar leads were then used to reconstruct 12-lead ECGs recorded from an unseen group of subjects (normal = 5 7, LVH = 59, MI = 69). It was found that the suggested leads could reconstruct the 12-lead ECG with a median RMS error of 57.0 mu V and a correlation coefficient of 0.968. This was in comparison to the EASI leads, which, when tested on the same data, produced a median RMS error of 60.1 mu V and a correlation coefficient of 0.955. Following comparison on a lead-by-lead basis the eigen-vector based leads were found to perform particularly well in the reconstruction of the precordial leads V1-V5. Further work is required to refine and to determine the practical utility of the proposed leads in practice.
引用
收藏
页码:361 / +
页数:2
相关论文