Piston-Based vs Peristaltic Pump-Based CT Injector Systems

被引:0
作者
Chaya, Amy [1 ]
Jost, Gregor [2 ]
Endrikat, Jan [3 ,4 ]
机构
[1] Bayer US LLC, Radiol Syst & Adv Syst Integrat, Indianola, PA 15051 USA
[2] Bayer AG, Magnet Resonance & Computed Tomog Contrast Media, Berlin, Germany
[3] Bayer AG, Radiol Res & Dev, Berlin, Germany
[4] Univ Med Sch Saarland, Dept Gynecol Obstet & Reprod Med, Homburg, Germany
关键词
iodinated contrast media; CT injector technology; fluid delivery parameters; piston-based; peristaltic pump; contrast media delivery; power injection; COMPUTED TOMOGRAPHIC ANGIOGRAPHY; HIGH-FLOW RATES; IODINE CONCENTRATION; CONTRAST MATERIAL; PEAK ENHANCEMENT; IN-VITRO; VISCOSITY; IMPACT;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To compare the fluid delivery performance of 2 different technologies for administration of iodinated contrast media in computed tomography (CT). Methods The maximum achievable flow rate and the steady-state flow rate variance of a piston-based contrast media injector system (Stellant MP, MEDRAD) was compared with peristaltic pump-based injector systems (CT motion, Ulrich Medical; CT Expres, Bracco). The contrast media iopromide (Ultravist) and iopamidol (Isovue) were used at 2 concentrations each (300 and 370 mg I/mL) and 3 catheter sizes (18, 20, and 22 G) to test the injector performance. Results Average maximum achievable flow rates for room temperature iopromide (370 mg I/mL) using a 20 G catheter were 7.6, 7.1, and 4.8 mL per second for the Stellant MP injector, CT motion injector, and CT Expres injector, respectively. The Stellant MP injector achieved significantly higher flow rates compared to the CT Expres injector for all catheter sizes tested (P < .001). Higher flow rates also were observed for the Stellant MP injector compared to the CT motion injector, with 20 G and 22 G catheters (P < .001). The Stellant MP injector featured a constant steady-state flow rate (variance, 0.04 mL/s), whereas the other systems injected in a pulsatile fashion, with significantly greater variance (P < .001). Discussion To the authors' knowledge, this is the first reported laboratory study providing preliminary evidence of differences between the fluid delivery performance of CT injection systems. Additional investigations using a dedicated flow phantom simulating human physiological flow parameters should be conducted, and depending on the results, a clinical study could assess the effect on image quality. Conclusion The piston-based injector demonstrated higher maximum achievable flow rates and more consistent steady-state flow when compared to peristaltic pump-based injectors.
引用
收藏
页码:344 / 352
页数:9
相关论文
共 22 条
  • [1] [Anonymous], 2018, ACR MAN CONTR MED VE
  • [2] Effect of contrast material injection duration and rate on aortic peak time and peak enhancement at dynamic CT involving injection protocol with dose tailored to patient weight
    Awai, K
    Hiraishi, K
    Hori, S
    [J]. RADIOLOGY, 2004, 230 (01) : 142 - 150
  • [3] Aortic and hepatic peak enhancement at CT: Effect of contrast medium injection rate - Pharmacokinetic analysis and experimental porcine model
    Bae, KT
    Heiken, JP
    Brink, JA
    [J]. RADIOLOGY, 1998, 206 (02) : 455 - 464
  • [4] Identification of the Iodine Concentration That Yields the Highest Intravascular Enhancement in MDCT Angiography
    Behrendt, Florian F.
    Pietsch, Hubertus
    Jost, Gregor
    Palmowski, Moritz
    Guenther, Rolf W.
    Mahnken, Andreas H.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 200 (05) : 1151 - 1156
  • [5] Intra-individual comparison of different contrast media concentrations (300 mg, 370 mg and 400 mg iodine) in MDCT
    Behrendt, Florian F.
    Pietsch, Hubertus
    Jost, Gregor
    Sieber, Martin A.
    Keil, Sebastian
    Plumhans, Cedric
    Seidensticker, Peter
    Guenther, Rolf W.
    Mahnken, Andreas H.
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (07) : 1644 - 1650
  • [6] Impact of Different Vein Catheter Sizes for Mechanical Power Injection in CT: In Vitro Evaluation with Use of a Circulation Phantom
    Behrendt, Florian F.
    Bruners, Philipp
    Keil, Sebastian
    Plumhans, Cedric
    Mahnken, Andreas H.
    Stanzel, Sven
    Das, Marco
    Gunther, Rolf W.
    Muehlenbruch, Georg
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (01) : 25 - 31
  • [7] BOLUS GEOMETRY AND DYNAMICS AFTER INTRAVENOUS CONTRAST-MEDIUM INJECTION
    CLAUSSEN, CD
    BANZER, D
    PFRETZSCHNER, C
    KALENDER, WA
    SCHORNER, W
    [J]. RADIOLOGY, 1984, 153 (02) : 365 - 368
  • [8] Iodine Concentration and Optimization in Computed Tomography Angiography Current Issues
    Faggioni, Lorenzo
    Gabelloni, Michela
    [J]. INVESTIGATIVE RADIOLOGY, 2016, 51 (12) : 816 - 822
  • [9] CT Angiography: Injection and Acquisition Technique
    Fleischmann, Dominik
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2010, 48 (02) : 237 - +
  • [10] Catheter Insertion for Intravenous (IV) Contrast Infusion in Multidetector-Row Computed Tomography (MDCT): Defining How Catheter Caliber Selection Affects Procedure of Catheter Insertion, IV Contrast Infusion Rate, Complication Rate, and MDCT Image Quality
    Johnson, Pamela T.
    Christensen, Gale
    Lai, Hong
    Eng, John
    Fishman, Elliot K.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2014, 38 (02) : 281 - 284