Influential Factors in the Preparation of 68Ga-DOTATATE

被引:2
作者
Fang, Ping [1 ]
Jacobson, Mark S. [1 ]
Hung, Joseph C. [1 ]
机构
[1] Mayo Clin, Div Nucl Med, 200 First St SW, Rochester, MN 55905 USA
关键词
Ga-68-DOTATATE; Ge-68; breakthrough; NETSPOT; neuroendocrine tumors; PET; preparation;
D O I
10.2967/jnmt.119.241224
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Acceptable and reproducible radiochemical purity (RP) for Ga-68-DOTATATE was difficult to obtain with the NETSPOT kit because the manufacturer instructions lacked details on the heater or needles used. Methods: The drug was prepared in an International Organization for Standardization (ISO) 5 environment. Multiple dry baths and needle types were used to investigate the effects of reaction temperature and metal contamination, respectively. Temperature curves were obtained with a calibrated thermocouple. The influence of the accuracy of the NETSPOT reagent volume and its effect on outcome were investigated. Results: The AccuBlock dry bath required recalibration for the ISO 5 environment; after calibration, the temperature was stable (only +/- 0.1 degrees C from the set point). When we followed package insert recommendations (dry bath temperature set to 98 degrees C, reaction time of 8 min), the reaction temperature was 90.6 degrees C. When Becton Dickinson needles were used for reconstitution, 15 of 18 runs (83%) did not meet the RP specification. However, B. Braun Medical needles achieved satisfactory and stable RP. When the Ga-68 generator was eluted with 5.0 mL of 0.1 M hydrochloric acid (HCI), only 3.8-3.9 mL of eluate reached the reaction vial; this volume did not impact labeling (final pH was 3.8). The labeling success rate increased markedly if the Ga-68 eluate was passed through a conditioned silica gel cartridge or if no cartridge was used; then, RP was more than 99%. HCI contact with the septum of the labeling vial reduced RP. Conclusion: The needle type and the temperature setting of the dry bath have critical roles in Ga-68-DOTATATE preparation. The AccuBlock dry bath has excellent stability and accuracy and can be used for reliable preparation. By using a conditioned silica gel cartridge or by eliminating the cartridge altogether, the RP is reliably high and stable.
引用
收藏
页码:263 / 268
页数:6
相关论文
共 6 条
[1]  
[Anonymous], 2016, EUR PHARM, P1148
[2]  
[Anonymous], 2017, NETSPOT PACK INS
[3]  
[Anonymous], 2019, NETSPOT PACK INS
[4]   68Ga-DOTATATE Compared with 111In-DTPA-Octreotide and Conventional Imaging for Pulmonary and Gastroenteropancreatic Neuroendocrine Tumors: A Systematic Review and Meta-Analysis [J].
Deppen, Stephen A. ;
Blume, Jeffrey ;
Bobbey, Adam J. ;
Shah, Chirayu ;
Graham, Michael M. ;
Lee, Patricia ;
Delbeke, Dominique ;
Walker, Ronald C. .
JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (06) :872-878
[5]   Neuroendocrine Tumor Recurrence: Diagnosis with 68Ga-DOTATATE PET/CT [J].
Haug, Alexander R. ;
Cindea-Drimus, Ramona ;
Auernhammer, Christoph J. ;
Reincke, Martin ;
Beuschlein, Felix ;
Waengler, Bjoern ;
Uebleis, Christopher ;
Schmidt, Gerwin P. ;
Spitzweg, Christine ;
Bartenstein, Peter ;
Hacker, Marcus .
RADIOLOGY, 2014, 270 (02) :517-525
[6]   Influence of metal ions on the 68Ga-labeling of DOTATATE [J].
Oehlke, Elisabeth ;
Le, Van So ;
Lengkeek, Nigel ;
Pellegrini, Paul ;
Jackson, Tim ;
Greguric, Ivan ;
Weiner, Ron .
APPLIED RADIATION AND ISOTOPES, 2013, 82 :232-238