Flow optimization in one-dimensional and comprehensive two-dimensional gas chromatography

被引:12
作者
Blumberg, Leonid M. [1 ]
机构
[1] Advachrom, POB 1243, Wilmington, DE 19801 USA
关键词
Efficiency-optimized flow rate (EOF); Speed-optimized flow rate (SOF); Default flow rate; Flow reconciliation in GC x GC; PLATE HEIGHT; SEPARATION PERFORMANCE; METHOD-TRANSLATION; MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; COLUMN PERFORMANCE; CONSTANT-PRESSURE; KINETIC PLOTS; CAPILLARY GC; HEATING RATE;
D O I
10.1016/j.chroma.2017.08.040
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Theoretical considerations of optimal flow in a GC column under different conditions are outlined from a single perspective, and numerical recommendations are provided. The optimal flow rate in a temperature programmed column is about 30% lower than its isothermal counterpart in the same column with the same carrier gas. A simplified calculation of recommended optimal or near optimal flow rate in a single column under different conditions is summarized in a single equation. In GC x GC, the primary column operates under temperature-programmed conditions while each secondary run is essentially isothermal. As a result, their flow optimization requirements are different. The complementary IDs (internal diameters) corresponding to equal optimal flow rates in both GC x GC columns should relate as (2)d = 0.7(.1)d. Typical complementary ID pairs are tabulated. If choosing the complementary IDs is not an option, several ways to reconcile the difference in their optimal flow rates are considered. The most typical cases of the column mismatch are identified and their boundary conditions are formulated. The effect of the flow reconciliation in these column pairs on the performance of GC x GC analysis is evaluated. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:27 / 38
页数:12
相关论文
共 62 条
[31]   NOMENCLATURE FOR CHROMATOGRAPHY [J].
ETTRE, LS .
PURE AND APPLIED CHEMISTRY, 1993, 65 (04) :819-872
[32]   A NEW METHOD FOR PREDICTION OF BINARY GAS-PHASE DIFFUSION COEFFECIENTS [J].
FULLER, EN ;
SCHETTLE, PD ;
GIDDINGS, JC .
INDUSTRIAL AND ENGINEERING CHEMISTRY, 1966, 58 (05) :19-+
[33]   PLATE HEIGHT IN GAS CHROMATOGRAPHY [J].
GIDDINGS, JC ;
SEAGER, SL ;
STUCKI, LR ;
STEWART, GH .
ANALYTICAL CHEMISTRY, 1960, 32 (07) :867-870
[36]   THEORY OF CHROMATOGRAPHY .9. THE THEORETICAL PLATE CONCEPT IN COLUMN SEPARATIONS [J].
GLUECKAUF, E .
TRANSACTIONS OF THE FARADAY SOCIETY, 1955, 51 (01) :34-44
[37]  
Golay M.J. E., 1958, GAS CHROMATOGR INT S, P36
[38]  
Halasz I., 1965, GAS CHROMATOGRAPHY 1, V1964, P38
[39]  
Hinshaw JV, 2001, LC GC N AM, V19, P1056
[40]   Kinetic plots for programmed temperature gas chromatography [J].
Jespers, Sander ;
Roeleveld, Kevin ;
Lynen, Frederic ;
Broeckhoven, Ken ;
Desmet, Gert .