Comparison of automated measurements of electrocardiographic intervals and durations by computer-based algorithms of digital electrocardiographs

被引:48
作者
Kligfield, Paul [1 ]
Badilini, Fabio [2 ]
Rowlandson, Ian [3 ]
Xue, Joel [3 ]
Clark, Elaine [4 ]
Devine, Brian [4 ]
Macfarlane, Peter [4 ]
de Bie, Johan [5 ]
Mortara, David [5 ]
Babaeizadeh, Saeed [6 ]
Gregg, Richard [6 ]
Helfenbein, Eric D. [6 ]
Green, Cynthia L. [7 ]
机构
[1] Weill Cornell Med Coll, Div Cardiol, New York, NY 10065 USA
[2] AMPS LLC, New York, NY USA
[3] GE Healthcare, Milwaukee, WI USA
[4] Univ Glasgow, Glasgow Program, Glasgow, Lanark, Scotland
[5] Mortara Instrument, Milwaukee, WI USA
[6] Philips Healthcare, Andover, MA USA
[7] Duke Univ, Med Ctr, Duke Clin Res Inst, Durham, NC USA
关键词
SUDDEN CARDIAC DEATH; QT-INTERVAL; QRS DURATION; THOROUGH QT; CARDIOVASCULAR-DISEASE; HYPERTENSIVE PATIENTS; SCIENTIFIC STATEMENT; CLINICAL CARDIOLOGY; HEART-FAILURE; MORTALITY;
D O I
10.1016/j.ahj.2013.10.004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Purpose Automated measurements of electrocardiographic (ECG) intervals are widely used by clinicians for individual patient diagnosis and by investigators in population studies. We examined whether clinically significant systematic differences exist in ECG intervals measured by current generation digital electrocardiographs from different manufacturers and whether differences, if present, are dependent on the degree of abnormality of the selected ECGs. Methods Measurements of RR interval, PR interval, QRS duration, and QT interval were made blindly by 4 major manufacturers of digital electrocardiographs used in the United States from 600 XML files of ECG tracings stored in the US FDA ECG warehouse and released for the purpose of this study by the Cardiac Safety Research Consortium. Included were 3 groups based on expected QT interval and degree of repolarization abnormality, comprising 200 ECGs each from (1) placebo or baseline study period in normal subjects during thorough QT studies, (2) peak moxifloxacin effect in otherwise normal subjects during thorough QT studies, and (3) patients with genotyped variants of congenital long QT syndrome (LQTS). Results Differences of means between manufacturers were generally small in the normal and moxifloxacin subjects, but in the LQTS patients, differences of means ranged from 2.0 to 14.0 ms for QRS duration and from 0.8 to 18.1 ms for the QT interval. Mean absolute differences between algorithms were similar for QRS duration and QT intervals in the normal and in the moxifloxacin subjects (mean <= 6 ms) but were significantly larger in patients with LQTS. Conclusions Small but statistically significant group differences in mean interval and duration measurements and means of individual absolute differences exist among automated algorithms of widely used, current generation digital electrocardiographs. Measurement differences, including QRS duration and the QT interval, are greatest for the most abnormal ECGs.
引用
收藏
页码:150 / +
页数:11
相关论文
共 38 条
[1]   Comparing methods of measurement for detecting drug-induced changes in the QT interval: Implications for thoroughly conducted ECG studies [J].
Azie, NE ;
Adams, G ;
Darpo, B ;
Francom, SF ;
Polasek, EC ;
Wisser, JM ;
Fleishaker, JC .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2004, 9 (02) :166-174
[2]  
Brenyo A, 2011, CARDIOL J, V18, P8
[3]   Highly Automated QT Measurement Techniques in 7 Thorough QT Studies Implemented under ICH E14 Guidelines [J].
Couderc, Jean-Philippe ;
Garnett, Christine ;
Li, Mike ;
Handzel, Robert ;
McNitt, Scott ;
Xia, Xiajuan ;
Polonsky, Slava ;
Zareba, Wojciech .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2011, 16 (01) :13-24
[4]   Man versus machine: Is there an optimal method for QT measurements in thorough QT studies? [J].
Darpo, Borje ;
Agin, Marilyn ;
Kazierad, Dawd J. ;
Layton, Gary ;
Muirhead, Gary ;
Gray, Peter ;
Jorkasky, Diane K. .
JOURNAL OF CLINICAL PHARMACOLOGY, 2006, 46 (06) :598-612
[5]   Prolonged QT interval predicts cardiac and all-cause mortality in the elderly - The Rotterdam study [J].
de Bruyne, MC ;
Hoes, AW ;
Kors, JA ;
Hofman, A ;
van Bemmel, JH ;
Grobbee, DE .
EUROPEAN HEART JOURNAL, 1999, 20 (04) :278-284
[6]   Heart rate-corrected QT interval prolongation predicts risk of coronary heart disease in black and white middle-aged men and women - The ARIC study [J].
Dekker, JM ;
Crow, RS ;
Hannan, PJ ;
Schouten, EG ;
Folsom, AR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (04) :565-571
[7]   Normal ECG in childhood and adolescence [J].
Dickinson, DF .
HEART, 2005, 91 (12) :1626-1630
[8]   Practice standards for electrocardiographic monitoring in hospital settings - An American Heart Association Scientific Statement from the councils on cardiovascular nursing, clinical cardiology, and cardiovascular disease in the young [J].
Drew, BJ ;
Califf, RM ;
Funk, M ;
Kaufman, ES ;
Krucoff, MW ;
Laks, MM ;
Macfarlane, PW ;
Sommargren, C ;
Swiryn, S ;
Van Hare, GF .
CIRCULATION, 2004, 110 (17) :2721-2746
[9]  
Elming Hanne, 2002, Card Electrophysiol Rev, V6, P289, DOI 10.1023/A:1016345412555
[10]   Prolonged QT Interval Diagnosis Suppression by a Widely Used Computerized ECG Analysis System [J].
Garg, Anubhav ;