Incremental diagnostic benefit of resolution recovery software in patients with equivocal myocardial perfusion single-photon emission computed tomography (SPECT)

被引:8
作者
Qutub, Mohammed A. [1 ,3 ]
Dowsley, Taylor [1 ]
Ali, Iftikhar [1 ]
Wells, R. Glenn [1 ]
Chen, Li [2 ]
Ruddy, Terrence D. [1 ,4 ]
Chow, Benjamin J. W. [1 ,4 ]
机构
[1] Univ Ottawa, Inst Heart, Dept Med Cardiol, Ottawa, ON K1Y 4W7, Canada
[2] Univ Ottawa, Inst Heart, Cardiovasc Methods Ctr, Ottawa, ON K1Y 4W7, Canada
[3] King Abdulaziz Univ, Dept Med, Jeddah 21413, Saudi Arabia
[4] Univ Ottawa, Dept Radiol, Ottawa, ON K1Y 4W7, Canada
关键词
SPECT; PET; resolution recovery; attenuation correction; COMPENSATION; TIME; ATTENUATION; AGREEMENT;
D O I
10.1007/s12350-013-9732-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Though myocardial perfusion imaging (MPI) with single-photon emission computed tomography (SPECT) is an established diagnostic method, equivocal studies are commonly encountered. New software has been introduced that incorporates resolution recovery (RR) and noise regulation into the reconstruction algorithm and has been used to facilitate "half-dose" and "half-time" studies. Its utility with "full-time, full-dose" acquisition has not been well studied. We sought to understand the potential benefit of incorporating RR software in equivocal SPECT studies. Patients with full-time, full-dose SPECT MPI were reviewed and those with equivocal results, who subsequently underwent cardiac Rb-82 positron emission tomography (PET) scan were identified. Image reconstruction was performed with iterative reconstruction (IR), attenuation correction (IR + AC), and RR software (IR + AC + RR). Images were anonymized and read blindly by consensus of two experienced readers. All images were qualitatively assessed and semi-quantitatively graded using summed stress and summed rest scores. 45 patients were included (28 males, age = 59.6 +/- A 9.9 years) and the diagnostic accuracy of each of the reconstruction algorithms (IR, IR + AC, IR + AC + RR) was compared to Rb-82 PET. Agreement of clinical diagnosis of each SPECT reconstruction with Rb-PET showed incremental improvement. The agreement with PET for IR + AC + RR (kappa = 0.66, CI 0.454-0.875) is significantly better than for IR (kappa = 0.22, CI 0.0-0.450, P = .005) and for IR + AC (kappa = 0.32, CI 0.077-0.563, P = .03). Also, IR + AC + RR improved the clinical diagnosis in 14 cases and with overall improvement of reclassification proportion of 23.5% compared to IR (P = .01). Using PET as a reference standard, ROC curves were created for IR + AC + RR, IR + AC, and IR which showed incremental value of the area under the curve of IR + AC + RR (AUC: 0.87; CI 0.76-0.98) over IR + AC (AUC: 0.75; CI 0.61-0.89, P = .078), and over IR (AUC 0.68; CI 0.52-0.84, P = .025). The addition of RR may help in the diagnosis of patients with equivocal SPECT MPI without the need for additional testing. Further prospective studies are needed to define the role of this new software.
引用
收藏
页码:545 / 552
页数:8
相关论文
共 18 条
[1]   Half-Time SPECT Myocardial Perfusion Imaging with Attenuation Correction [J].
Ali, Iftikhar ;
Ruddy, Terrence D. ;
Almgrahi, Abdulaziz ;
Anstett, Frank G. ;
Wells, R. Glenn .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (04) :554-562
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies [J].
Borges-Neto, Salvador ;
Pagnanelli, Robert A. ;
Shaw, Linda K. ;
Honeycutt, Emily ;
Shwartz, Shuli C. ;
Adams, George L. ;
Coleman, Ralph Edward .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (04) :555-565
[4]   COMPARING THE AREAS UNDER 2 OR MORE CORRELATED RECEIVER OPERATING CHARACTERISTIC CURVES - A NONPARAMETRIC APPROACH [J].
DELONG, ER ;
DELONG, DM ;
CLARKEPEARSON, DI .
BIOMETRICS, 1988, 44 (03) :837-845
[5]   Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" filtered backprojection [J].
DePuey, E. Gordon ;
Gadiraju, Ramesh ;
Clark, John ;
Thompson, Linda ;
Anstett, Frank ;
Shwartz, Shuli C. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2008, 15 (04) :547-563
[6]   ANALYSIS OF PROBABILITY AS AN AID IN THE CLINICAL-DIAGNOSIS OF CORONARY-ARTERY DISEASE [J].
DIAMOND, GA ;
FORRESTER, JS .
NEW ENGLAND JOURNAL OF MEDICINE, 1979, 300 (24) :1350-1358
[7]  
Dilsizian V., 2009, J. Nucl. Cardio, V16, P651, DOI [10.1007/s12350-009-9094-9, DOI 10.1007/S12350-009-9094-9]
[8]  
FICARO EP, 2008, J NUCL CARDIOL, V15, pS20
[9]   Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT [J].
Frey, EC ;
Gilland, KL ;
Tsui, BMW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (09) :1040-1050
[10]   LROC analysis of detector-response compensation in SPECT [J].
Gifford, HC ;
King, MA ;
Wells, RG ;
Hawkins, WG ;
Narayanan, MV ;
Pretorius, PH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (05) :463-473