The isotopic composition of commercially available uranium chemical reagents

被引:12
作者
Richter, S [1 ]
Alonso, A [1 ]
Wellum, R [1 ]
Taylor, PDP [1 ]
机构
[1] Commiss European Communities, Inst Reference Mat & Measurements, JRC, B-2440 Geel, Belgium
关键词
D O I
10.1039/a900774a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The uranium isotopic compositions of 12 commercially available uranium chemicals, obtained from various suppliers, were measured by high accuracy isotope mass spectrometry. The goal was to establish the range of isotopic compositions in currently available commercial uranium chemicals, offered in different chemical forms and used for different purposes (e.g., atomic absorption spectrometry or inductively coupled plasma mass spectrometry). The relative isotope abundances of U-234, U-235, U-236 and U-238 were determined by thermal ionisation mass spectrometry with high abundance sensitivity (Finnigan MAT262-RPQ-Plus), using the n(235U)ln(238U) amount ratio measurements of a gravimetrically prepared Isotopic Reference Material (IRMM-183) for mass fractionation correction. All uncertainties were calculated according to the ISO Guide to the Expression of Uncertainty in Measurement (International Standards Organization, Geneva, Switzerland), using a coverage factor k = 2. For n(U-235)/n(U-238) relative uncertainties of 0.2-0.3% and for n(U-234)/n(U-238) and n(U-236)/n(U-238) of 0.4-3.0%, depending on the relative amounts of the U-234 and U-236 isotopes, were achieved. All samples were found to be depleted in U-235 by a factor of two compared with its natural molar abundance (0.72%). The non-natural isotope 236U was present in all samples analyzed. The range found for the molar abundances of U-234 is 0.0022-0.00081%, for U-235 0.35-0.21%, for U-236 0.021-0.000011% and for U-238 99.788-99.634%.
引用
收藏
页码:889 / 891
页数:3
相关论文
共 6 条
  • [1] [Anonymous], 1993 ISO
  • [2] De Bievre P., 1973, Z Anal Chem, V264, P365, DOI [10.1007/BF00423634, DOI 10.1007/BF00423634]
  • [3] Ovaskainen R., 1997, P 19 ANN S SAF NUCL
  • [4] RICHTER S, UNPUB
  • [5] Isotopic compositions of the elements 1997
    Rosman, KJR
    Taylor, PDP
    [J]. PURE AND APPLIED CHEMISTRY, 1998, 70 (01) : 217 - 235
  • [6] PERFORMANCE OF A THERMAL IONIZATION MASS-SPECTROMETER WITH A DECELERATION LENS SYSTEM AND POST-DECELERATION DETECTOR SELECTION
    VANCALSTEREN, P
    SCHWIETERS, JB
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1995, 146 : 119 - 129