3D interfractional patient position verification using 2D-3D registration of orthogonal images

被引:26
作者
Jans, H. -S.
Syme, A. M.
Rathee, S.
Fallone, B. G.
机构
[1] Univ Alberta, Dept Med Phys, Cross Canc Inst, Dept Oncol & Phys, Edmonton, AB T6G 1Z2, Canada
[2] Univ Alberta, Dept Med Phys, Cross Canc Inst, Dept Oncol, Edmonton, AB T6G 1Z2, Canada
关键词
2D-3D rigid registration; orthogonal epid images; image processing; mutual information; patient setup error; 6; DOF; fractionated radiation therapy; TCP; NTCP;
D O I
10.1118/1.2192907
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reproducible positioning of the patient during fractionated external beam radiation therapy is imperative to ensure that the delivered dose distribution matches the planned one. In this paper, we expand on a 2D-3D image registration method to verify a patient's setup in three dimensions (rotations and translations) using orthogonal portal images and megavoltage digitally reconstructed radiographs (MDRRs) derived from CT data. The accuracy of 2D-3D registration was improved by employing additional image preprocessing steps and a parabolic fit to interpolate the parameter space of the cost function utilized for registration. Using a humanoid phantom, precision for registration of three-dimensional translations was found to be better than 0.5 mm (1 s.d.)for any axis when no rotations were present. Three-dimensional rotations about any axis were registered with a precision of better than 0.2 degrees (1 s.d.) when no translations were present. Combined rotations and translations of up to 4 degrees and 15 mm were registered with 0.4 degrees and 0.7 mm accuracy for each axis. The influence of setup translations on registration of rotations and vice versa was also investigated and mostly agrees with a simple geometric model. Additionally, the dependence of registration accuracy on three cost functions, angular spacing between MDRRs, pixel size, and field-of-view, was examined. Best results were achieved by mutual information using 0.5 degrees angular spacing and a 10 x 10 CM2 field-of-view with 140 x 140 pixels. Approximating patient motion as rigid transformation, the registration method is applied to two treatment plans and the patients' setup errors are determined. Their magnitude was found to be <= 6.1 mm and <= 2.7 degrees for any axis in all of the six fractions measured for each treatment plan. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1420 / 1439
页数:20
相关论文
共 49 条
[1]   Entropy-based dual-portal-to-3-DCT registration incorporating pixel correlation [J].
Bansal, R ;
Staib, LH ;
Chen, Z ;
Rangarajan, A ;
Knisely, J ;
Nath, R ;
Duncan, JS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (01) :29-49
[2]   A faster method for 3D/2D medical image registration - a simulation study [J].
Birkfellner, W ;
Wirth, J ;
Burgstaller, W ;
Baumann, B ;
Staedele, H ;
Hammer, B ;
Gellrich, NC ;
Jacob, AL ;
Regazzoni, P ;
Messmer, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (16) :2665-2679
[3]   On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinoma [J].
Chung, PWM ;
Haycocks, T ;
Brown, T ;
Cambridge, Z ;
Kelly, V ;
Alasti, H ;
Jaffray, DA ;
Catton, CN .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :329-334
[4]   Patient setup error measurement using 3D intensity-based image registration techniques [J].
Clippe, S ;
Sarrut, D ;
Malet, C ;
Miguet, S ;
Ginestet, C ;
Carrie, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :259-265
[5]   Evaluation of mechanical precision and alignment uncertainties for an integrated CT/LINAC system [J].
Court, L ;
Rosen, I ;
Mohan, R ;
Dong, L .
MEDICAL PHYSICS, 2003, 30 (06) :1198-1210
[6]   Electronic portal image assisted reduction of systematic set-up errors in head and neck irradiation [J].
de Boer, HCJ ;
de Koste, JRV ;
Creutzberg, CL ;
Visser, AG ;
Levendag, PC ;
Heijmen, BJM .
RADIOTHERAPY AND ONCOLOGY, 2001, 61 (03) :299-308
[7]   Evaluation of cost functions for gray value matching of two-dimensional images in radiotherapy [J].
Dekker, N ;
Ploeger, LS ;
van Herk, M .
MEDICAL PHYSICS, 2003, 30 (05) :778-784
[8]   An image correlation procedure for digitally reconstructed radiographs and electronic portal images [J].
Dong, L ;
Boyer, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1053-1060
[9]   Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer [J].
Erridge, SC ;
Seppenwoolde, Y ;
Muller, SH ;
van Herk, M ;
De Jaeger, K ;
Belderbos, JSA ;
Boersma, LJ ;
Lebesque, JV .
RADIOTHERAPY AND ONCOLOGY, 2003, 66 (01) :75-85
[10]   Verification of patient position and delivery of IMRT by electronic portal imaging [J].
Fielding, AL ;
Evans, PM ;
Clark, CH .
RADIOTHERAPY AND ONCOLOGY, 2004, 73 (03) :339-347