Speckle-based x-ray phase-contrast imaging with a laboratory source and the scanning technique

被引:42
作者
Zhou, Tunhe [1 ]
Zanette, Irene [2 ,3 ,4 ]
Zdora, Marie-Christine [2 ,3 ]
Lundstrom, Ulf [1 ,5 ]
Larsson, Daniel H. [1 ]
Hertz, Hans M. [1 ]
Pfeiffer, Franz [2 ,3 ,6 ]
Burvall, Anna [1 ]
机构
[1] KTH Royal Inst Technol, Biomed & Xray Phys, SE-10691 Stockholm, Sweden
[2] Tech Univ Munich, Dept Phys, D-85748 Garching, Germany
[3] Tech Univ Munich, Inst Med Tech, D-85748 Garching, Germany
[4] Diamond Light Source, Didcot OX11 0DE, Oxon, England
[5] Stanford Univ, Dept Biol Struct, Stanford, CA 94305 USA
[6] Tech Univ Munich, Inst Diagnost & Intervent Radiol, D-81675 Munich, Germany
基金
瑞典研究理事会;
关键词
RETRIEVAL;
D O I
10.1364/OL.40.002822
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The speckle-based scanning method for x-ray phase-contrast imaging is implemented with a liquid-metal-jet source. Using the two-dimensional scanning technique, the phase shift introduced by the object is retrieved in both transverse orientations, and the limitations on spatial resolution inherent to the speckle-tracking technique are avoided. This method opens up possibilities of new high-resolution multimodal applications for lab-based phasecontrast x-ray imaging. (C) 2015 Optical Society of America
引用
收藏
页码:2822 / 2825
页数:4
相关论文
共 19 条
[1]   At-wavelength metrology of hard X-ray mirror using near field speckle [J].
Berujon, Sebastien ;
Wang, Hongchang ;
Alcock, Simon ;
Sawhney, Kawal .
OPTICS EXPRESS, 2014, 22 (06) :6438-6446
[2]   X-ray multimodal imaging using a random-phase object [J].
Berujon, Sebastien ;
Wang, Hongchang ;
Sawhney, Kawal .
PHYSICAL REVIEW A, 2012, 86 (06)
[3]   Two-Dimensional X-Ray Beam Phase Sensing [J].
Berujon, Sebastien ;
Ziegler, Eric ;
Cerbino, Roberto ;
Peverini, Luca .
PHYSICAL REVIEW LETTERS, 2012, 108 (15)
[4]   X-ray-scattering information obtained from near-field speckle [J].
Cerbino, R. ;
Peverini, L. ;
Potenza, M. A. C. ;
Robert, A. ;
Bosecke, P. ;
Giglio, M. .
NATURE PHYSICS, 2008, 4 (03) :238-243
[5]   In Situ Measurements of Strains in Composite Battery Electrodes during Electrochemical Cycling [J].
Jones, E. M. C. ;
Silberstein, M. N. ;
White, S. R. ;
Sottos, N. R. .
EXPERIMENTAL MECHANICS, 2014, 54 (06) :971-985
[6]   A two-directional approach for grating based differential phase contrast imaging using hard x-rays [J].
Kottler, C. ;
David, C. ;
Pfeiffer, F. ;
Bunk, O. .
OPTICS EXPRESS, 2007, 15 (03) :1175-1181
[7]   First application of liquid-metal-jet sources for small-animal imaging: High-resolution CT and phase-contrast tumor demarcation [J].
Larsson, Daniel H. ;
Lundstrom, Ulf ;
Westermark, Ulrica K. ;
Henriksson, Marie Arsenian ;
Burvall, Anna ;
Hertz, Hans M. .
MEDICAL PHYSICS, 2013, 40 (02)
[8]   X-ray phase contrast for CO2 microangiography [J].
Lundstrom, U. ;
Larsson, D. H. ;
Burvall, A. ;
Takman, P. A. C. ;
Scott, L. ;
Brismar, H. ;
Hertz, H. M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09) :2603-2617
[9]   X-ray phase imaging with a paper analyzer [J].
Morgan, Kaye S. ;
Paganin, David M. ;
Siu, Karen K. W. .
APPLIED PHYSICS LETTERS, 2012, 100 (12)
[10]   Two-dimensional digital image correlation for in-plane displacement and strain measurement: a review [J].
Pan, Bing ;
Qian, Kemao ;
Xie, Huimin ;
Asundi, Anand .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (06)