Digital flat-panel X-ray technology based on cesium iodide and amorphous silicon:: experimental studies and preliminary clinical results.

被引:0
作者
Strotzer, M [1 ]
Gmeinwieser, J [1 ]
Völk, M [1 ]
Fründ, R [1 ]
Feuerbach, S [1 ]
机构
[1] Klinikum Univ Regensburg, Inst Rontgendiagnostik, D-93042 Regensburg, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1999年 / 170卷 / 01期
关键词
radiography; digital; technology; amorphous silicon;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Experimental and clinical evaluation of a digital flat-panel X-ray system based on cesium iodide (CsI) and amorphous silicon (a-Si). Methods: Performance of a prototype detector was compared with conventional screen-film radiography (SFR) using several phantom studies. Foreign bodies, fractures, osteolyses, and pulmonary lesions were analyzed. Additionally, 120 patients were studied prospectively, resulting in 400 comparative X-ray studies. The fiat-panel detector was exposed with standard dose and with a dose reduction of up to 75%. Detector size was 15 x 15 cm, pixel matrix was 1 x Ik with a pixel size of 143 mu m. Modulation-transfer function was determined to be 18% at the maximum spatial resolution of 3.5 Ip/mm. Results: The diagnostic results achieved with the digital detector were similar to those of conventional SFR; even at reduced radiation exposure. A potential for dose reduction was observed: 50% with respect to osteoarthrosis and fractures, end 75% for determining bony alignment. Discussion: This new technology can be used in thoracic and skeletal radiography. A significant dose reduction is possible, depending on the suspected disease.
引用
收藏
页码:66 / 72
页数:7
相关论文
共 30 条
  • [21] Amorphous silicon, flat-panel, X-ray detector versus storage phosphor-based computed radiography:: Contrast-detail phantom study at different tube voltages and detector entrance doses
    Hamer, OW
    Völk, M
    Zorger, N
    Feuerbach, S
    Strotzer, M
    INVESTIGATIVE RADIOLOGY, 2003, 38 (04) : 212 - 220
  • [22] Should 3K zoom function be used for detection of pneumothorax in cesium iodide/amorphous silicon flat-panel detector radiographs presented on 1K-matrix soft copies?
    Karin A. Herrmann
    H. M. Bonél
    A. Stäbler
    M. Voelk
    M. Strotzer
    C. J. Zech
    M. F. Reiser
    European Radiology, 2006, 16
  • [23] Should 3K zoom function be used for detection of pneumothorax in cesium iodide/amorphous silicon flat-panel detector radiographs presented on 1K-matrix soft copies?
    Herrmann, Karin A.
    Bonel, H. M.
    Staebler, A.
    Voelk, M.
    Strotzer, M.
    Zech, C. J.
    Reiser, M. F.
    EUROPEAN RADIOLOGY, 2006, 16 (12) : 2768 - 2774
  • [24] Digital chest radiography with a solid-state flat-panel X-ray detector Contrast-detail evaluation with processed images printed on film hard copy
    Chotas, HG
    Ravin, CE
    RADIOLOGY, 2001, 218 (03) : 679 - 682
  • [25] Technical and clinical results of an experimental Flat Dynamic (digital) X-ray image Detector (FDXD) system with real-time corrections
    Bruijns, TJC
    Alving, PL
    Baker, EL
    Bury, R
    Cowen, AR
    Jung, N
    Luijendijk, HA
    Meulenbrugge, HJ
    Stouten, HJ
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 33 - 44
  • [26] Performance of a 41X41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications
    Granfors, PR
    Aufrichtig, R
    Possin, GE
    Giambattista, BW
    Huang, ZS
    Liu, JQ
    Ma, B
    MEDICAL PHYSICS, 2003, 30 (10) : 2715 - 2726
  • [27] Optimization of key building blocks for a large area radiographic and fluoroscopic dynamic digital X-ray detector based on a-Si:H/CsI:Tl flat panel technology
    Ducourant, T
    Michel, M
    Vieux, G
    Peppler, T
    Trochet, JC
    Schulz, RF
    Bastiaens, R
    Busse, F
    MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 14 - 25
  • [28] Technical assessment of 2D and 3D imaging performance of an IGZO-based flat-panel X-ray detector
    Sheth, Niral Milan
    Uneri, Ali
    Helm, Patrick A.
    Zbijewski, Wojciech
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2022, 49 (05) : 3053 - 3066
  • [29] Dual energy chest x-ray for improved COVID-19 detection using a dual-layer flat-panel detector: Simulation and phantom studies
    Shi, Linxi
    Bennett, N. Robert
    Lu, Minghui
    Sun, Mingshan
    Zhang, Jin
    Star-Lack, Josh
    Tsai, Emily B.
    Guo, H. Henry
    Wang, Adam S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [30] Dual energy x-ray imaging and scoring of coronary calcium: physics-based digital phantom and clinical studies
    Zhou, Bo
    Wen, Di
    Nye, Katelyn
    Gilkeson, Robert C.
    Wilson, David L.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783