Repeatability of tear meniscus evaluation using spectral-domain Cirrus® HD-OCT and time-domain Visante® OCT

被引:27
作者
Chan, Hiok Hong [1 ]
Zhao, Yang [2 ]
Tun, Tin A. [3 ]
Tong, Louis [4 ,5 ,6 ,7 ]
机构
[1] Singapore Natl Eye Ctr, Singapore 168751, Singapore
[2] Natl Univ Singapore, Yong Loo Lin Sch Med, Singapore 119228, Singapore
[3] Singapore Eye Res Inst, Singapore 168751, Singapore
[4] Singapore Eye Res Inst, Ocular Surface Res Grp, Singapore 168751, Singapore
[5] Singapore Natl Eye Ctr, Dept Cornea & External Eye Dis, Singapore 168751, Singapore
[6] Duke NUS Grad Med Sch, Off Clin Acad & Fac Affairs, Singapore 169857, Singapore
[7] Natl Univ Singapore, Yong Loo Lin Sch Med, Dept Ophthalmol, Singapore 119228, Singapore
基金
英国医学研究理事会;
关键词
Tear meniscus; Tear prism; OCT; Dry eye; Imaging; OPTICAL COHERENCE TOMOGRAPHY; HEIGHT MEASUREMENT; DIAGNOSIS; REPRODUCIBILITY; AGREEMENT;
D O I
10.1016/j.clae.2015.04.002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate the intra-rater, inter-rater and inter-device repeatability of a spectral-domain OCT (Cirrus) and a time-domain OCT (Visante) for tear meniscus height (TMH) and area (TMA) measurements. Methods: 20 participants with no known eye disease were recruited. Both eyes of participants were imaged with both OCTs under the similar conditions. The inferior tear meniscus was imaged at 6 o'clock position whereas the superior meniscus was imaged at 12 o'clock position. Data from the right eyes was analyzed. Two raters independently measured TMH and TMA using the OCT images, and one rater repeated the measurements. Intra-rater, inter-rater and inter-device repeatability of measurements were assessed using Bland-Altman plots and pooled standard deviation. Results: For intra-rater repeatability, TMH and TMA measurements were more repeatable in Cirrus than Visante (95% limits of agreement (LOA): TMH (mu m), -22 to 66 (Cirrus), -125 to 45 (Visante); TMA ( mu m(2)), 1632 to 5331 (Cirrus), 38,050 to 21,874 (Visante)). For inter-rater agreement, TMH and TMA were also more repeatable in Cirrus than Visante (95%LOA: TMH (mu m), -29 to 107 (Cirrus), -215 to 252 (Visante); TMA (mu m(2)), -6650 to 9567 (Cirrus), -33,119 to 39,272 (Visante)). Inter-device agreement was poor (95%LOA: TMH (mu m), -158 to 150; TMA (mu m(2)), -32,903 to 14,076). There was no significant difference in inferior TMH between Cirrus and Visante (p >0.05). Inferior TMA was significantly lower in Cirrus by a mean difference of 10,223 mu m(2) (95% confidence interval, 5479, 14,966) (p = 0.0002). Conclusion: Spectral-domain OCT is superior to time-domain OCT for intra-rater and inter-rater repeatability of TMH and TMA measurements. (C) 2015 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:368 / 372
页数:5
相关论文
共 26 条
[1]   Variability of the analysis of the tear meniscus height by optical coherence tomography [J].
Bitton, Etty ;
Keech, Adam ;
Simpson, Trefford ;
Jones, Lyndon .
OPTOMETRY AND VISION SCIENCE, 2007, 84 (09) :903-908
[2]   Interexaminer Reproducibility of Optical Coherence Tomography for Measuring the Tear Film Meniscus [J].
Canan, Handan ;
Altan-Yaycioglu, Rana ;
Ulas, Burak ;
Sizmaz, Selcuk ;
Coban-Karatas, Muge .
CURRENT EYE RESEARCH, 2014, 39 (12) :1145-1150
[3]  
Carl Zeiss Meditec A.G., 2009, CARL ZEISS MED AG PU
[4]  
Carl Zeiss Meditec I., 2009, CIRR HD OCT US MAN E
[5]   Tear Meniscus Measurement by Spectral Optical Coherence Tomography [J].
Czajkowski, Grzegorz ;
Kaluzny, Bartlomiej J. ;
Laudencka, Adriana ;
Malukiewicz, Grazyna ;
Kaluzny, Jakub J. .
OPTOMETRY AND VISION SCIENCE, 2012, 89 (03) :336-342
[6]   Optical coherence tomography: An emerging technology for biomedical imaging and optical biopsy [J].
Fujimoto, JG ;
Pitris, C ;
Boppart, SA ;
Brezinski, ME .
NEOPLASIA, 2000, 2 (1-2) :9-25
[7]   Tear Meniscus Evaluation by Anterior Segment Swept-Source Optical Coherence Tomography [J].
Fukuda, Reina ;
Usui, Tomohiko ;
Miyai, Takashi ;
Yamagami, Satoru ;
Amano, Shiro .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 155 (04) :620-624
[8]  
HOLLY FJ, 1985, T OPHTHAL SOC UK, V104, P374
[9]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[10]  
Huang Zhen Ping, 2012, Eye Sci, V27, P217, DOI 10.3969/j.issn.1000-4432.2012.04.014