摘要
Objective: The description and evaluation of a novel patient-independent seizure detection for the EEG of the newborn term infant. Methods: We identified characteristics of neonatal seizures by which a human observer is able to detect them. Neonatal seizures were divided into two types. For each type, a fully automated detection algorithm was developed based on the identified human observer characteristics. The first algorithm analyzes the correlation between high-energetic segments of the EEG. The second detects increases in low-frequency activity (<8 Hz) with high autocorrelation. Results: The complete algorithm was tested on multi-channel EEG recordings of 21 patients with and 5 patients without electrographic seizures, totaling 217 h of EEG. Sensitivity of the combined algorithms was found to be 88%, Positive Predictive Value (PPV) 75% and the false positive rate 0.66 per hour. Conclusions: Our approach to separate neonatal seizures into two types yields a high sensitivity combined with a good PPV and much lower false positive rate than previously published algorithms. Significance: The proposed algorithm significantly improves neonatal seizure detection and monitoring. (C) 2008 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:2447 / 2454
页数:8
相关论文
共 29 条
[11]   Combination of EEG and ECG for improved automatic neonatal seizure detection [J].
Greene, Barry R. ;
Boylan, Geraldine B. ;
Reilly, Richard B. ;
de Chazal, Philip ;
Connolly, Sean .
CLINICAL NEUROPHYSIOLOGY, 2007, 118 (06) :1348-1359
[12]   SILENT SEIZURES IN SICK INFANTS IN EARLY LIFE - DIAGNOSIS BY CONTINUOUS CEREBRAL FUNCTION MONITORING [J].
HELLSTROMWESTAS, L ;
ROSEN, I ;
SWENNINGSEN, NW .
ACTA PAEDIATRICA SCANDINAVICA, 1985, 74 (05) :741-748
[13]   Consequences of neonatal seizures in the rat: Morphological and behavioral effects [J].
Holmes, GL ;
Gairsa, JL ;
Chevassus-Au-Louis, N ;
Ben-Ari, Y .
ANNALS OF NEUROLOGY, 1998, 44 (06) :845-857
[14]  
KAISER JF, 1990, INT CONF ACOUST SPEE, P381, DOI 10.1109/ICASSP.1990.115702
[15]   Wavelet analysis for neonatal electroencephalographic seizures [J].
Kitayama, M ;
Otsubo, H ;
Parvez, S ;
Lodha, A ;
Ying, E ;
Parvez, B ;
Ishii, R ;
Mizuno-Matsumoto, Y ;
Zoroofi, RA ;
Snead, OC .
PEDIATRIC NEUROLOGY, 2003, 29 (04) :326-333
[16]   Early-life seizures in rats increase susceptibility to seizure-induced brain injury in adulthood [J].
Koh, S ;
Storey, TW ;
Santos, TC ;
Mian, AY ;
Cole, AJ .
NEUROLOGY, 1999, 53 (05) :915-921
[17]   Teager-Kaiser energy operation of surface EMG improves muscle activity onset detection [J].
Li, Xiaoyan ;
Zhou, Ping ;
Aruin, Alexander S. .
ANNALS OF BIOMEDICAL ENGINEERING, 2007, 35 (09) :1532-1538
[18]   Consequences of recurrent seizures during early brain development [J].
Liu, Z ;
Yang, Y ;
Silveira, DC ;
Sarkisian, MR ;
Tandon, P ;
Huang, LT ;
Stafstrom, CE ;
Holmes, GL .
NEUROSCIENCE, 1999, 92 (04) :1443-1454
[19]   Neonatal seizures: A clinician's overview [J].
Lombroso, CT .
BRAIN & DEVELOPMENT, 1996, 18 (01) :1-28
[20]   Electrographic seizures in neonates correlate with poor neurodevelopmental outcome [J].
McBride, MC ;
Laroia, N ;
Guillet, R .
NEUROLOGY, 2000, 55 (04) :506-513