A continuous loading apparatus for measuring three-dimensional stiffness of ankle-foot orthoses

被引:18
作者
Cappa, P
Patanè, F
Di Rosa, G
机构
[1] Univ Roma La Sapienza, Dept Mech & Aeronaut, Rome, Italy
[2] Childrens Hosp Bambino Gesu, Paediat Neurorehabil Div, Rome, Italy
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2005年 / 127卷 / 06期
关键词
D O I
10.1115/1.2049313
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper describes a novel device to evaluate the mechanical properties of ankle foot orthoses (AFOs). The apparatus permits the application to AFOs of continuous three-dimensional (3D) movements between specified and settable endpoints. Using all x-v robot with a rotary stage and a six-component load cell, characteristic displacement versus reaction force curves can be generoted and consequently the ankle moments can be determined as a function of dorsi/plantar flexion, inv/eversion and int/external rotation. Representative curves for two polypropylene lateral leaf AFOs, different in shape but produced for the same leg by a skilled orthotist, are presented to illustrate the capabilities of the novel testing system. The metrological investigation showed that the apparatus creates a highly repeatable data set (uncertainty <= 1% FSO).
引用
收藏
页码:1025 / 1029
页数:5
相关论文
共 22 条
[21]   ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion - part 1: ankle, hip, and spine [J].
Wu, G ;
Siegler, S ;
Allard, P ;
Kirtley, C ;
Leardini, A ;
Rosenbaum, D ;
Whittle, M ;
D'Lima, DD ;
Cristofolini, L ;
Witte, H ;
Schmid, O ;
Stokes, H .
JOURNAL OF BIOMECHANICS, 2002, 35 (04) :543-548
[22]  
YAMAMOTO S, 1993, ORTHOT INT, V5, P259