Distal Locking of Femoral Nails: Evaluation of a New Radiation-Independent Targeting System

被引:11
作者
Arlettaz, Yvan [1 ]
Dominguez, Alexander [2 ]
Farron, Alain [3 ]
Ehlinger, Matthieu [4 ]
Moor, Beat K. [1 ]
机构
[1] Hosp Wallis, Orthopaed & Trauma Dept, CH-1951 Sion, Switzerland
[2] Univ Appl Sci, Sch Hlth Sci, Lausanne, Switzerland
[3] Univ Lausanne, Serv Orthopaed & Traumatol, CH-1015 Lausanne, Switzerland
[4] Univ Strasbourg, Hosp Hautepierre, Serv Orthopaed & Traumatol, Strasbourg, France
关键词
femur; intramedullary nailing; locking screws; targeting device; nail deformation; AIMING-DEVICE; COMMINUTED FRACTURES; SCREWS; INSERTION; EXPOSURE; DEFORMATION; TIBIA;
D O I
10.1097/BOT.0b013e318242d8b7
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: The purpose of this study was to assess the effectiveness of a novel radiation-independent aiming device for distal locking of intramedullary nails in a human cadaver model. Methods: A new targeting system was used in 25 intact human cadaver femora for the distal locking procedure after insertion of an intramedullary nail. The number of successful screw placements and the time needed for this locking procedure were recorded. The accuracy of the aiming process was evaluated by computed tomography. Results: The duration of the distal locking process was 8.0 +/- 6 1.8 minutes (mean +/- SD; range, 4-11 minutes). None of the screw placements required fluoroscopic guidance. Computed tomography revealed high accuracy of the locking process. The incidence angle (a) of the locking screws through the distal locking holes of the nail was 86.8 degrees +/- 5.0 degrees (mean +/- SD; range, 80 degrees-96 degrees). Targeting failed in 1 static locking screw because of a material defect in the drilling sleeve. Conclusions: This cadaver study indicated that an aiming armbased targeting device is highly reliable and accurate. The promising results suggest that it will help to decrease radiation exposure compared with the traditional "free-hand technique."
引用
收藏
页码:633 / 637
页数:5
相关论文
共 36 条
  • [1] Experimental model for a new distal locking aiming device for solid intramedullary tibia nails
    Abdlslam, KM
    Bonnaire, F
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 (05): : 363 - 366
  • [2] Evaluation of the Stryker S2™ IM Nail Distal Targeting Device for reduction of radiation exposure: A case series study
    Anastopoulos, George
    Ntagiopoulos, Panagiotis G.
    Chissas, Dionisios
    Loupasis, George
    Asimakopoulos, Antonios
    Athanaselis, Eustratios
    Megas, Panagiotis
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (10): : 1210 - 1215
  • [3] Barrick E F, 1993, J Orthop Trauma, V7, P248, DOI 10.1097/00005131-199306000-00009
  • [4] Bonshahi A Y, 2003, J Postgrad Med, V49, P99
  • [5] Brumback RJ, 1996, CLIN ORTHOP RELAT R, P292
  • [6] Dittmann V., 2008, VERWENDUNG LEICHEN L
  • [7] A SIMPLE JIG TO EASE THE INSERTION OF DISTAL SCREWS IN INTRAMEDULLARY LOCKING NAILS
    HASHEMINEJAD, A
    GARLICK, N
    GODDARD, NJ
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1994, 25 (06): : 407 - 408
  • [8] LOCKING NAILING - AN AID TO DISTAL TARGETTING
    HUDSON, I
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1989, 20 (03): : 129 - 130
  • [9] Johnson K D, 1987, Instr Course Lect, V36, P437
  • [10] JOHNSON KD, 1987, ORTHOP CLIN N AM, V18, P133