The placement of lumbar pedicle screws using computerised stereotactic guidance

被引:70
作者
Girardi, FP [1 ]
Cammisa, FP [1 ]
Sandhu, HS [1 ]
Alvarez, L [1 ]
机构
[1] Cornell Univ, Weill Med Coll, Hosp Special Surg, Spine Serv, New York, NY 10021 USA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1999年 / 81B卷 / 05期
关键词
D O I
10.1302/0301-620X.81B5.9244
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Computer-assisted frameless stereotactic image guidance allows precise preoperative planning and intraoperative localisation of the image, It has been developed and tested in the laboratory. We evaluated the efficacy, clinical results and complications of placement of a pedicle screw in the lumbar spine using this technique, A total of 62 patients (28 men, 34 women) had lumbar decompression and spinal fusion with segmental pedicle screws, Postoperative CT scans were taken of 35 patients to investigate the placement of 330 screws. None showed penetration of the medial or inferior wall of a pedicle, Registration was carried out 66 times. The number of fiducial points used on each registration averaged 5.8 (4 to 7) The mean registration error was 0.75 mm (0.32 to 1.72). This technique provides a safe and reliable guide for placement of transpedicular screws in the lumbar spine.
引用
收藏
页码:825 / 829
页数:5
相关论文
共 23 条
[11]   AN ACCURATE AND ERGONOMIC METHOD OF REGISTRATION FOR IMAGE-GUIDED NEUROSURGERY [J].
HENDERSON, JM ;
SMITH, KR ;
BUCHOLZ, RD .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (04) :273-277
[12]   APPLICATION OF FRAMELESS STEREOTAXY TO PEDICLE SCREW FIXATION OF THE SPINE [J].
KALFAS, IH ;
KORMOS, DW ;
MURPHY, MA ;
MCKENZIE, RL ;
BARNETT, GH ;
BELL, GR ;
STEINER, CP ;
TRIMBLE, MB ;
WEISENBERGER, JP .
JOURNAL OF NEUROSURGERY, 1995, 83 (04) :641-647
[13]  
Merloz P, 1997, CLIN ORTHOP RELAT R, P86
[14]   IMAGE-GUIDED INSERTION OF TRANSPEDICULAR SCREWS - A LABORATORY SET-UP [J].
NOLTE, LP ;
ZAMORANO, LJ ;
JIANG, ZW ;
WANG, QH ;
LANGLOTZ, F ;
BERLEMANN, U .
SPINE, 1995, 20 (04) :497-500
[15]   A FRAMELESS STEREOTAXIC INTEGRATION OF COMPUTERIZED TOMOGRAPHIC IMAGING AND THE OPERATING MICROSCOPE [J].
ROBERTS, DW ;
STROHBEHN, JW ;
HATCH, JF ;
MURRAY, W ;
KETTENBERGER, H .
JOURNAL OF NEUROSURGERY, 1986, 65 (04) :545-549
[16]  
ROYCAMILLE R, 1988, CLIN ORTHOP RELAT R, V203, P7
[17]   Accuracy of computer-assisted pedicle screw placement - An in vivo computed tomography analysis [J].
Schwarzenbach, O ;
Berlemann, U ;
Jost, B ;
Visarius, H ;
Arm, E ;
Langlotz, F ;
Nolte, LP ;
Ozdoba, C .
SPINE, 1997, 22 (04) :452-458
[18]  
*SCOL RES SOC, 1987, MORB MORT COMM REP
[19]   THE NEUROSTATION(TM) - A HIGHLY ACCURATE, MINIMALLY INVASIVE SOLUTION TO FRAMELESS STEREOTAXIC NEUROSURGERY [J].
SMITH, KR ;
FRANK, KJ ;
BUCHOLZ, RD .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (04) :247-256
[20]  
VACCARO AR, 1995, SPINE, V20, P157